510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Accumetrics,Inc. Verifynow Prutest, Accumetrics, Inc. Verifynow Prutest

K141427510(k) TraditionalCleared Jul 24, 2015420-day reviewClass II21 CFR 864.5700Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Accumetrics, Inc. · JOZ System, Automated Platelet Aggregation

Cleared on Jul 24, 2015 as substantially equivalent to K051231 (Accumetrics, Inc., 2005) after a 420-day review. Its family tree reaches back 2 generations to 1983. 1 later device descends from it. It has been recalled 2 times.

Decided
Jul 24, 2015
Review
420 days
Class
Class II
Product code
JOZ
Recalls
2
Predicates
1

Its longest line of descent runs back 2 generations to K830749 (1983). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2019: 1 descendant clearedK830749 · Platelet Aggregometer · Chrono-Log Corp. · 1983K051231 · Verifynow-P2Y12 Assay · Accumetrics, Inc. · 2005 · recalled19801990200020102020
Descendants, by generation 1 generation forward: 1 descendant K051231 · Verifynow-P2Y12 Assay · Accumetrics, Inc. · 2005 · recalledK830749 · Platelet Aggregometer · Chrono-Log Corp. · 19830
  1. Root · 2 generations backK830749 Platelet AggregometerChrono-Log Corp. · 1983
  2. 1 generation backK051231 Verifynow-P2Y12 AssayAccumetrics, Inc. · 2005 · Recalled
  3. This deviceK141427 Accumetrics,Inc. Verifynow Prutest, Accumetrics, Inc. Verifynow…Accumetrics, Inc. · 2015
  4. Descendant · cites it directly · 0 of its ownK181777 AggreGuide A-100 ADPAggredyne, Inc. · 2019 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K181777AggreGuide A-100 ADPAggredyne, Inc. · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K830749Platelet AggregometerChrono-Log Corp.1983—
−1K051231Verifynow-P2Y12 AssayAccumetrics, Inc.20053
0 (this device)K141427Accumetrics,Inc. Verifynow Prutest, Accumetrics, Inc. Verifynow PrutestAccumetrics, Inc.2015—
+1K181777AggreGuide A-100 ADPAggredyne, Inc.20191

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
FIG. 1 — Lineage of K141427, 1983–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOZ device

  1. #1 K191364 T-TAS 01 Total Thrombus-formation Analysis…score -0.6 · 3 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051231Verifynow-P2Y12 Assay
Summary page 4
…n name: Automated platelet aggregation system (21 CFR §864.5700; Product Code: JOZ) Predicate device: VerifyNow P2Y12 Assay, Accumetrics, Inc., k051231 Device description: The VerifyNow System is a turbidimetric-based optical detection system that measures platelet- induced aggr…
Accumetrics, Inc.20053

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181777AggreGuide A-100 ADPAggredyne, Inc.20190

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 21, 2025Class IIIZ-0500-2026Under Investigation by firmDue to an device without a premarket clearance being incorrectly package and distributed.
Mar 30, 2022Class IIZ-1121-2022Employee errorTwo whole blood Platelet Reactivity tests exist that share the same reagent formulation and are used to obtain P2Y12 Reactivity Unit results. The test without US-FDA market clearance was distributed and it displays BASE results not…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 32 years and passes through 2 companies; a new generation was cleared every 22.2 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K181777 (Aggredyne, Inc., 2019) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JOZ devices

JOZ System, Automated Platelet Aggregation: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · JOZ

1in 2020

2012: 12013: 12015: 22016: 12017: 12019: 12020: 1
2007high 2 (2015) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOZ

Too few JOZ 510(k)s in any year since 2007 to give a median.

Recalls per year · JOZ

1in 2025

2013: 12024: 12025: 1
2007high 1 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOZ

3in 2025

2010: 12012: 12013: 12014: 12017: 12018: 252019: 92020: 22021: 22022: 162024: 12025: 3
2007high 25 (2018) · low 0 (2007)2025

MAUDE reports naming product code JOZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—01
20110—00
20121—01
20131—11
20140—01
20152—00
20161—00
20171—01
20180—025
20191—09
20201—02
20210—02
20220—016
20230—00
20240—11
20250—13
20260—0—

What is it indicated for?

The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. For in vitro diagnostic use. For professional use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141427: Accumetrics,Inc. Verifynow Prutest, Accumetrics, Inc. Verifynow Prutest.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141427

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026