510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Verifynow-P2Y12 Assay

K051231510(k) TraditionalCleared Aug 5, 200584-day reviewClass II21 CFR 864.5700Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Accumetrics, Inc. · JOZ System, Automated Platelet Aggregation

Cleared on Aug 5, 2005 as substantially equivalent to K830749 (Chrono-Log Corp., 1983) after an 84-day review. Its family tree reaches back 1 generation to 1983. 2 later devices descend from it. It has been recalled 3 times.

Decided
Aug 5, 2005
Review
84 days
Class
Class II
Product code
JOZ
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K830749 (1983). 2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2015: 1 descendant cleared2019: 1 descendant clearedK830749 · Platelet Aggregometer · Chrono-Log Corp. · 198319801990200020102020
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant K830749 · Platelet Aggregometer · Chrono-Log Corp. · 19830
  1. Root · 1 generation backK830749 Platelet AggregometerChrono-Log Corp. · 1983
  2. This deviceK051231 Verifynow-P2Y12 AssayAccumetrics, Inc. · 2005
  3. Descendant · cites it directly · 1 of its ownK141427 Accumetrics,Inc. Verifynow Prutest, Accumetrics, Inc. Verifynow…Accumetrics, Inc. · 2015 · Recalled
  4. Descendant · 2 generations on · 0 of its ownK181777 AggreGuide A-100 ADPAggredyne, Inc. · 2019 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K141427Accumetrics,Inc. Verifynow Prutest, Accumetrics, Inc. Verifynow PrutestAccumetrics, Inc. · 2015 · cites it directly1descendant
  • K181777AggreGuide A-100 ADPAggredyne, Inc. · 2019 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K830749Platelet AggregometerChrono-Log Corp.1983—
0 (this device)K051231Verifynow-P2Y12 AssayAccumetrics, Inc.2005—
+1K141427Accumetrics,Inc. Verifynow Prutest, Accumetrics, Inc. Verifynow PrutestAccumetrics, Inc.20152
+2K181777AggreGuide A-100 ADPAggredyne, Inc.20191

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201910
FIG. 1 — Lineage of K051231, 1983–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOZ device

  1. #1 K191364 T-TAS 01 Total Thrombus-formation Analysis…score -0.6 · 3 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K830749Platelet Aggregometer
Summary page 1
…ces, specifically the CHRONO-LOG Whole Blood Aggregometer with CHRONO-PAR®ADP Reagent, as the predicate device. The CHRONO-LOG Whole Blood Aggregometer was cleared under k830749. Device Description: The VerifyNow System is a turbidimetric based optical detection system, which measures platelet induced aggr…
Chrono-Log Corp.1983—

Also cleared against K830749: 8 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K141427Accumetrics,Inc. Verifynow Prutest, Accumetrics, Inc. Verifynow PrutestAccumetrics, Inc.20151

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 6, 2012Class IIZ-1854-2012No Marketing ApplicationThe recall was initiated by Accumetrics because of a field correction to the VerifyNow System; specifically the results reported when running the VerifyNow P2Y12 Test. Currently the VerifyNow P2Y12 Test reports three values: PRU, BASE and…
Feb 6, 2012Class IIZ-1853-2012No Marketing ApplicationThe recall was initiated by Accumetrics because of a field correction to the VerifyNow System; specifically the results reported when running the VerifyNow P2Y12 Test. Currently the VerifyNow P2Y12 Test reports three values: PRU, BASE and…
Sep 5, 2006Not classifiedZ-0228-2007OtherThe VerifyNow P2Y12 assay can report an erroneous result instead of an error message when a sample with low hematocrit is run.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K141427 (Accumetrics, Inc., 2015) has gone furthest, with 1 descendant of its own.
  • 8 other devices also cite its predicate K830749.

How does its category compare?

All JOZ devices

JOZ System, Automated Platelet Aggregation: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JOZ

1in 2020

2012: 12013: 12015: 22016: 12017: 12019: 12020: 1
2007high 2 (2015) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOZ

Too few JOZ 510(k)s in any year since 2007 to give a median.

Recalls per year · JOZ

1in 2025

2013: 12024: 12025: 1
2007high 1 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOZ

3in 2025

2010: 12012: 12013: 12014: 12017: 12018: 252019: 92020: 22021: 22022: 162024: 12025: 3
2007high 25 (2018) · low 0 (2007)2025

MAUDE reports naming product code JOZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—01
20110—00
20121—01
20131—11
20140—01
20152—00
20161—00
20171—01
20180—025
20191—09
20201—02
20210—02
20220—016
20230—00
20240—11
20250—13
20260—0—

What is it indicated for?

The VerifyNow-P2Y1 2 Assay is a whole blood assay used in the laboratory or

point of care setting to measure the level of platelet P2Y1 2 receptor blockade.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051231: Verifynow-P2Y12 Assay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051231

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026