510(k) Report

510(k) · Cleared

Cleared

Echo-Screen Iii Pro Hearing Screener with Barcode Scanner, Echo-Screen Iii Pro Hearing Screener Without Barcode Scanner,

K141446510(k) SpecialCleared Aug 22, 201481-day reviewClass II21 CFR 882.1900

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Natus Medical Incorporated · GWJ Stimulator, Auditory, Evoked Response

Cleared on Aug 22, 2014 as substantially equivalent to K013977 (Fischer-Zoth Audiologic Systems, Inc., 2002) after an 81-day review. Its family tree reaches back 4 generations to 1990.

Decided
Aug 22, 2014
Review
81 days
Class
Class II
Product code
GWJ
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K902540 (1990). 11 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1999: 1 other ancestor1997: 1 other ancestor1995: 2 other ancestors1993: 1 other ancestor1994: 1 other ancestor1991: 1 other ancestorK902540 · Gsi 17 Screening Audiometer · Grason-Stadler, Inc. · 1990K974237 · Gsi 70 Audiopath Screener · Grason-Stadler, Inc. · 1998K982642 · Echo-Screen · Fischer-Zoth Audiologic Systems, Inc. · 1998K013977 · Echo-Screen T, TA, TD, Tda, TC · Fischer-Zoth Audiologic Systems, Inc. · 2002199019952000200520102015
  1. Root · 4 generations backK902540 Gsi 17 Screening AudiometerGrason-Stadler, Inc. · 1990
  2. 3 generations backK974237 Gsi 70 Audiopath ScreenerGrason-Stadler, Inc. · 1998
  3. 2 generations backK982642 Echo-ScreenFischer-Zoth Audiologic Systems, Inc. · 1998
  4. 1 generation backK013977 Echo-Screen T, TA, TD, Tda, TCFischer-Zoth Audiologic Systems, Inc. · 2002
  5. This deviceK141446 Echo-Screen Iii Pro Hearing Screener with Barcode Scanner…Natus Medical Incorporated · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K902540Gsi 17 Screening AudiometerGrason-Stadler, Inc.1990—
−3K974237Gsi 70 Audiopath ScreenerGrason-Stadler, Inc.1998—
−2K982642Echo-ScreenFischer-Zoth Audiologic Systems, Inc.1998—
−1K013977Echo-Screen T, TA, TD, Tda, TCFischer-Zoth Audiologic Systems, Inc.2002—
0 (this device)K141446Echo-Screen Iii Pro Hearing Screener with Barcode Scanner, Echo-Screen Iii Pro…Natus Medical Incorporated2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199310
199410
199520
199710
199910
FIG. 1 — Lineage of K141446, 1990–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GWJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GWJ device

  1. #1 K242954 Integrity V500 (Integrity, Integrity with VEMP)score 10.0 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K013977Echo-Screen T, TA, TD, Tda, TC
Summary page 4
…Evoked Response Auditory Stimulator 21 CFR 882.1900 Product Code: GWJ Predicate Device: K013977 Echo-Screen T, TA, TD, TDA, TC Device Description: The Echo-Screen III hearing screener is a portable, handheld, battery-operat…
Fischer-Zoth Audiologic Systems, Inc.2002—

Also cleared against K013977: 5 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 3 companies; a new generation was cleared every 7.6 years on median.
  • 2 of the 11 ancestors have a recall of any class.
  • 5 other devices also cite its predicate K013977.

How does its category compare?

All GWJ devices

GWJ Stimulator, Auditory, Evoked Response: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · GWJ

2in 2024

2007: 22008: 42009: 42010: 12011: 22012: 12013: 22014: 22015: 12017: 22020: 12021: 12022: 12024: 2
2007high 4 (2008) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GWJ

111in 2009

2008: 64 days median2009: 111 days median
2007high 111 (2009) · low 64 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GWJ

1in 2025

2008: 12010: 12019: 12025: 1
2007high 1 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GWJ

1in 2025

2010: 22011: 52012: 12013: 12015: 32016: 12017: 432018: 242019: 52020: 32021: 32022: 82023: 62024: 72025: 1
2007high 43 (2017) · low 0 (2007)2025

MAUDE reports naming product code GWJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
200846410
2009411100
20101—12
20112—05
20121—01
20132—01
20142—00
20151—03
20160—01
20172—043
20180—024
20190—15
20201—03
20211—03
20221—08
20230—06
20242—07
20250—11
20260—0—

What is it indicated for?

The Echo-Screen III hearing screener models are based upon otoacoustic emission (OAE) and auditory brainstem response (ABR) technology. The device is intended to screen hearing for newborns through adults, including geriatric patients. The device does not measure hearing per se, but helps to determine whether or not a hearing loss may be present. The Echo-Screen III product family consists of handheld, automated OAE and ABR based hearing screening…

Read the full statement (125 words)

The Echo-Screen III hearing screener models are based upon otoacoustic emission (OAE) and auditory brainstem response (ABR) technology.

The device is intended to screen hearing for newborns through adults, including geriatric patients. The device does not measure hearing per se, but helps to determine whether or not a hearing loss may be present.

The Echo-Screen III product family consists of handheld, automated OAE and ABR based hearing screening systems which are easy to use. The measurement flow is menu guided and the evaluation is based upon signal statistics. The Echo-Screen III devices are intended to be used by trained personnel in a medical or school environment. The Echo-Screen III models are not intended for fitting assistive listening devices such as hearing aids or cochlear implants.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141446: Echo-Screen Iii Pro Hearing Screener with Barcode Scanner, Echo-Screen Iii Pro Hearing Screener Without Barcode Scanner,.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141446

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026