Zevo (TM) Anterior Cervical Plate System

K141632510(k) TraditionalCleared Dec 4, 2014169-day reviewClass II21 CFR 888.3060Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Sofamor Danek USA, Inc. · KWQ Appliance, Fixation, Spinal Intervertebral Body

Cleared on Dec 4, 2014 as substantially equivalent to K030866 (Synthes (Usa), 2003) and 4 other predicates after a 169-day review. Its family tree reaches back 7 generations to 1995. 18 later devices descend from it. It has been recalled once.

Decided
Dec 4, 2014
Review
169 days
Class
Class II
Product code
KWQ
Recalls
1
Predicates
5

Its longest line of descent runs back 7 generations to K953730 (1995). 17 ancestors in all. 18 later devices descend from it, 7 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 2 other ancestors2000: 1 other ancestor2007: 2 other ancestors2002: 1 other ancestor1997: 2 other ancestors2006: 1 other ancestor2004: 1 other ancestor2023: 4 descendants cleared2019: 4 descendants cleared2015: 1 descendant cleared2020: 1 descendant cleared2024: 3 descendants cleared2021: 1 descendant cleared2026: 1 descendant cleared2025: 3 descendants clearedK953730 · Codman Anterior Cervical Plate System · Johnson & Johnson Professionals, Inc. · 1995K013877 · Codman Slim-Loc System · Codman & Shurtleff, Inc. · 2001K040655 · Swift Anterior Cervical Plate System · Depuy Spine, Inc. · 2004K072546 · Swift Plus Anterior Cervical Plate System · Depuy Spine, Inc. · 2007×2: 2 generations cleared close together, K080646 (2008), K081038 (2008) K130640 · Atlantis Anterior Cervical Plate System · Medtronic Sofamor Danek USA, Inc. · 20131995200020052010201520202025
  1. Root · 7 generations backK953730 Codman Anterior Cervical Plate SystemJohnson & Johnson Professionals, Inc. · 1995
  2. 6 generations backK013877 Codman Slim-Loc SystemCodman & Shurtleff, Inc. · 2001
  3. 5 generations backK040655 Swift Anterior Cervical Plate SystemDepuy Spine, Inc. · 2004
  4. 4 generations backK072546 Swift Plus Anterior Cervical Plate SystemDepuy Spine, Inc. · 2007
  5. 3 generations backK080646 C-Tek Maxan Anterior Cervical Plate SystemBiomet Spine · 2008 · Recalled
  6. 2 generations backK081038 Atlantis Anterior Cervical Plate SystemMedtronic Sofamor Danek · 2008 · Recalled
  7. 1 generation backK130640 Atlantis Anterior Cervical Plate SystemMedtronic Sofamor Danek USA, Inc. · 2013
  8. This deviceK141632 Zevo (TM) Anterior Cervical Plate SystemMedtronic Sofamor Danek USA, Inc. · 2014
  9. Descendant · cites it directly · 9 of its ownK152623 ATLANTIS Anterior Cervical Plate SystemMedtronic Sofamor Danek USA, Inc. · 2015
  10. Descendant · cites it directly · 8 of its ownK182885 Atlantis™ Anterior Cervical Plate, Divergence™ Anterior Cervical…Medtronic Sofamor Danek USA, Inc. · 2019
  11. Descendant · 2 generations on · 4 of its ownK192502 Anatomic PEEK™ Cervical Fusion System, Anatomic PEEK™ PTC Cervical…Medtronic Sofamor Danek USA, Inc. · 2020
  12. Descendant · cites it directly · 3 of its ownK192314 RESONATE Anterior Cervical Plate SystemGlobus Medical, Inc. · 2019
  13. Descendant · 2 generations on · 2 of its ownK231251 FlexWing Anterior Cervical Plate SystemJeil Medical Corporation · 2024

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2015: 12019: 42020: 12021: 12023: 42024: 32025: 32026: 1
2015high 4 (2019) · low 0 (2016)2026

Most influential descendants

Showing 5 of 18 · See all
  • K152623ATLANTIS Anterior Cervical Plate SystemMedtronic Sofamor Danek USA, Inc. · 2015 · cites it directly9descendants
  • K182885Atlantis™ Anterior Cervical Plate, Divergence™ Anterior Cervical Fusion System, Premier™…Medtronic Sofamor Danek USA, Inc. · 2019 · cites it directly8descendants
  • K192502Anatomic PEEK™ Cervical Fusion System, Anatomic PEEK™ PTC Cervical Fusion System…Medtronic Sofamor Danek USA, Inc. · 2020 · 2 generations on4descendants
  • K192314RESONATE Anterior Cervical Plate SystemGlobus Medical, Inc. · 2019 · cites it directly3descendants
  • K231251FlexWing Anterior Cervical Plate SystemJeil Medical Corporation · 2024 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K953730Codman Anterior Cervical Plate SystemJohnson & Johnson Professionals, Inc.1995—
−6K013877Codman Slim-Loc SystemCodman & Shurtleff, Inc.2001—
−5K040655Swift Anterior Cervical Plate SystemDepuy Spine, Inc.2004—
−4K072546Swift Plus Anterior Cervical Plate SystemDepuy Spine, Inc.2007—
−3K080646C-Tek Maxan Anterior Cervical Plate SystemBiomet Spine20082
−2K081038Atlantis Anterior Cervical Plate SystemMedtronic Sofamor Danek20083
−1K130640Atlantis Anterior Cervical Plate SystemMedtronic Sofamor Danek USA, Inc.2013—
0 (this device)K141632Zevo (TM) Anterior Cervical Plate SystemMedtronic Sofamor Danek USA, Inc.2014—
+1K152623ATLANTIS Anterior Cervical Plate SystemMedtronic Sofamor Danek USA, Inc.2015—
+1K182885Atlantis™ Anterior Cervical Plate, Divergence™ Anterior Cervical Fusion…Medtronic Sofamor Danek USA, Inc.2019—
+2K192502Anatomic PEEK™ Cervical Fusion System, Anatomic PEEK™ PTC Cervical Fusion…Medtronic Sofamor Danek USA, Inc.2020—
+1K192314RESONATE Anterior Cervical Plate SystemGlobus Medical, Inc.2019—
+2K231251FlexWing Anterior Cervical Plate SystemJeil Medical Corporation2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199720
200010
200210
200320
200410
200610
200720

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201940
202010
202110
202340
202430
202530
202610
FIG. 1 — Lineage of K141632, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWQ device

  1. #1 K233640 Segmental Plating System (SPS);IdentiTi SPS…score 21.9 · 31 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030866Synthes Anterior CSLP System
Summary page 5
…d bone screws are substantially equivalent to the predicates: Primary Predicate:  Anterior Cervical Spine Locking Plate (CSLP) System, (K030866, SE 04/18/2003) [SYNTHES® Spine]…
Synthes (Usa)2003—
*K130640Atlantis Anterior Cervical Plate System
Summary page 6
…Page 3 of 3 Additional predicates:  ATLANTIS® Anterior Cervical Plate System (K130640, SE 06/04/2013; K081038, SE 08/15/2008) [Medtronic]  ZEPHIR® Anterior Cervical System (K030327, SE 02/26…
Medtronic Sofamor Danek USA, Inc.2013—
*K081038Atlantis Anterior Cervical Plate System
Summary page 6
…Page 3 of 3 Additional predicates:  ATLANTIS® Anterior Cervical Plate System (K130640, SE 06/04/2013; K081038, SE 08/15/2008) [Medtronic]  ZEPHIR® Anterior Cervical System (K030327, SE 02/26/2003; K994239, SE 06/19…
Medtronic Sofamor Danek20083
*K030327Zephir Anterior Cervical System
Summary page 6
… ATLANTIS® Anterior Cervical Plate System (K130640, SE 06/04/2013; K081038, SE 08/15/2008) [Medtronic]  ZEPHIR® Anterior Cervical System (K030327, SE 02/26/2003; K994239, SE 06/19/2000) [Medtronic] To the best of Medtronic’s knowledge, none of the predic…
Medtronic Sofamor Danek2003—
*K994239Zephir Anterior Cervical Plate System
Summary page 6
…late System (K130640, SE 06/04/2013; K081038, SE 08/15/2008) [Medtronic]  ZEPHIR® Anterior Cervical System (K030327, SE 02/26/2003; K994239, SE 06/19/2000) [Medtronic] To the best of Medtronic’s knowledge, none of the predicates listed here have been subj…
Medtronic Sofamor Danek, Inc.2000—

Also cleared against K030866: 41 other devices →

What descends from it? (18)

See all 18

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K152623ATLANTIS Anterior Cervical Plate SystemMedtronic Sofamor Danek USA, Inc.20159
K182885Atlantis™ Anterior Cervical Plate, Divergence™ Anterior Cervical Fusion System, Premier™…Medtronic Sofamor Danek USA, Inc.20198
K192314RESONATE Anterior Cervical Plate SystemGlobus Medical, Inc.20193
K191584FORTICO Anterior Cervical Fixation SystemCamber Spine Technologies, LLC20192
K231090Zevo Anterior Cervical Plate SystemMedtronic Sofamor Danek USA, Inc.20231
K1824894CIS® Pinehurst Anterior Cervical Plate systemSolco Biomedical Co., Ltd.20191
K223753Cervical Plate SystemDouble Medical Technology, Inc.20230

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 23, 2018Class IIZ-1750-2018Process controlSlot screw had an overall length of 15mm, but per the specification, the screw should have an overall length of approximately 17mm.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 6 companies; a new generation was cleared every 2.5 years on median.
  • 4 of the 17 ancestors have a recall of any class.
  • Of its descendants, K152623 (Medtronic Sofamor Danek USA, Inc., 2015) has gone furthest, with 9 descendants of its own.
  • 41 other devices also cite its predicate K030866.

How does its category compare?

All KWQ devices

KWQ Appliance, Fixation, Spinal Intervertebral Body: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · KWQ

7in 2025

2007: 272008: 322009: 232010: 232011: 222012: 262013: 222014: 172015: 252016: 152017: 202018: 222019: 262020: 262021: 172022: 182023: 162024: 172025: 72026 (partial year): 10
2007high 32 (2008) · low 7 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWQ

64in 2025

2007: 101 days median2008: 70 days median2009: 90 days median2010: 104 days median2011: 95 days median2012: 90 days median2013: 88 days median2014: 124 days median2015: 102 days median2016: 121 days median2017: 67 days median2018: 95 days median2019: 81 days median2020: 55 days median2021: 59 days median2022: 85 days median2023: 59 days median2024: 64 days median2025: 64 days median2026 (partial year): 76 days median
2007high 124 (2014) · low 55 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWQ

1in 2018

2007: 22008: 332009: 352010: 22011: 52012: 72013: 62015: 22016: 92017: 82018: 1
2007high 35 (2009) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWQ

248in 2025

2007: 2882008: 3482009: 3552010: 6912011: 4152012: 5442013: 7962014: 5462015: 5622016: 5022017: 4922018: 3412019: 2982020: 4002021: 5912022: 3922023: 3202024: 4192025: 248
2007high 796 (2013) · low 248 (2025)2025

MAUDE reports naming product code KWQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007271012288
2008327033348
2009239035355
2010231042691
201122955415
201226907544
201322886796
2014171240546
2015251022562
2016151219502
201720678492
201822951341
201926810298
202026550400
202117590591
202218850392
202316590320
202417640419
20257640248
202610760—

What is it indicated for?

The ZEVO™ Anterior Cervical Plate System is intended for anterior interbody screw fixation from C2 to T1. The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with: 1) degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), 2) trauma (including fractures), 3) tumors, 4) deformity (defined as kyphosis, lordosis, or scoliosis, 5) pseudoathrosis, and/or 6) failed previous fusions.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K141632: Zevo (TM) Anterior Cervical Plate System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141632

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026