510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Mobilediagnost WDR

K141895510(k) SpecialCleared Sep 18, 201466-day reviewClass II21 CFR 892.1720Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sedecal., Sa. · IZL System, X-Ray, Mobile

Cleared on Sep 18, 2014 as substantially equivalent to K111725 (Sedecal., Sa., 2011) after a 66-day review. Its family tree reaches back 7 generations to 1998. 3 later devices descend from it. It has been recalled 2 times.

Decided
Sep 18, 2014
Review
66 days
Class
Class II
Product code
IZL
Recalls
2
Predicates
1

Its longest line of descent runs back 7 generations to K981556 (1998). 15 ancestors in all, the oldest K931071 (1994). 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2001: 2 other ancestors1998: 1 other ancestor1997: 1 other ancestor1994: 2 other ancestors2002: 1 other ancestor2019: 1 descendant cleared2017: 1 descendant cleared2024: 1 descendant clearedK981556 · Canon X-Ray Digital Camera Cxdi-11 · Canon USA, Inc. · 1998K992547 · Canon X-Ray Digital Camera, Model Cxdi-22 · Canon USA, Inc. · 1999K023750 · Cannon Digital Radiography Model Cxdi-40G · Canon USA, Inc. · 2002K031447 · Canon, Digital Radiography, Model Cxdi-50G · Canon, Inc. · 2003K043002 · Sedecal Mobile X-Ray Units with Digital Detector · Sedecal., Sa. · 2004K090322 · Easy Moving Plus, SM-XXHF-YY · Sedecal., Sa. · 2009 · recalledK111725 · Mobilediagnost WDR · Sedecal., Sa. · 2011 · recalled1990200020102020
  1. Root · 7 generations backK981556 Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc. · 1998
  2. 6 generations backK992547 Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc. · 1999
  3. 5 generations backK023750 Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc. · 2002
  4. 4 generations backK031447 Canon, Digital Radiography, Model Cxdi-50GCanon, Inc. · 2003
  5. 3 generations backK043002 Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa. · 2004
  6. 2 generations backK090322 Easy Moving Plus, SM-XXHF-YYSedecal., Sa. · 2009 · Recalled
  7. 1 generation backK111725 Mobilediagnost WDRSedecal., Sa. · 2011 · Recalled
  8. This deviceK141895 Mobilediagnost WDRSedecal., Sa. · 2014
  9. Descendant · cites it directly · 1 of its ownK191813 MobileDiagnost wDR 2.2Sedecal., Sa. · 2019 · Recalled
  10. Descendant · 2 generations on · 0 of its ownK233678 Radiography 7000 MPhilips Medical Systems Dmc GmbH · 2024 · Recalled
  11. Descendant · cites it directly · 0 of its ownK173187 MobileDiagnost M50Philips India Limited · 2017

Most influential descendants

Showing 3 of 3 · See all
  • K191813MobileDiagnost wDR 2.2Sedecal., Sa. · 2019 · cites it directly1descendant
  • K233678Radiography 7000 MPhilips Medical Systems Dmc GmbH · 2024 · 2 generations on0descendants
  • K173187MobileDiagnost M50Philips India Limited · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K981556Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc.1998—
−6K992547Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc.1999—
−5K023750Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc.2002—
−4K031447Canon, Digital Radiography, Model Cxdi-50GCanon, Inc.2003—
−3K043002Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa.2004—
−2K090322Easy Moving Plus, SM-XXHF-YYSedecal., Sa.200931
−1K111725Mobilediagnost WDRSedecal., Sa.20118
0 (this device)K141895Mobilediagnost WDRSedecal., Sa.2014—
+1K191813MobileDiagnost wDR 2.2Sedecal., Sa.20194
+2K233678Radiography 7000 MPhilips Medical Systems Dmc GmbH20241
+1K173187MobileDiagnost M50Philips India Limited2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199710
199810
200120
200210
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
201910
202410
FIG. 1 — Lineage of K141895, 1998–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IZL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZL device

  1. #1 K242015 TOPAZ Mobile X-ray System (Models : TOPAZ-32D…score 19.9 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K111725Mobilediagnost WDR
Summary page 4
…MobileDiagnost wDR Manufacturer: Sedecal SA 510(k) Clearance: K111725 - July 19, 2011 (Primary) Product code IZL (Primary) Classific…
Sedecal., Sa.20118

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191813MobileDiagnost wDR 2.2Sedecal., Sa.20191
K173187MobileDiagnost M50Philips India Limited20170

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 10, 2025Class IIZ-0243-2026Device DesignSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper…
Oct 14, 2021Class IIZ-1083-2024Use errorWhile cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button.…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 2.3 years on median.
  • 7 of the 15 ancestors have a recall of any class.
  • Of its descendants, K191813 (Sedecal., Sa., 2019) has gone furthest, with 1 descendant of its own.

How does its category compare?

All IZL devices

IZL System, X-Ray, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · IZL

11in 2025

2007: 32008: 42009: 32010: 32011: 62012: 32014: 72015: 42016: 62017: 72018: 82019: 92020: 92021: 132022: 162023: 32024: 92025: 112026 (partial year): 6
2007high 16 (2022) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZL

153in 2025

2007: 30 days median2008: 49 days median2009: 125 days median2010: 65 days median2011: 66 days median2012: 83 days median2014: 104 days median2015: 88 days median2016: 47 days median2017: 45 days median2018: 47 days median2019: 33 days median2020: 43 days median2021: 48 days median2022: 89 days median2023: 55 days median2024: 159 days median2025: 153 days median2026 (partial year): 157 days median
2007high 159 (2024) · low 30 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZL

47in 2025

2007: 72010: 42011: 62012: 52013: 102014: 42015: 42019: 22020: 12022: 12023: 62024: 52025: 472026 (partial year): 5
2007high 47 (2025) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZL

27in 2025

2007: 162008: 272009: 352010: 752011: 1182012: 242013: 172014: 512015: 302016: 332017: 352018: 272019: 402020: 482021: 472022: 322023: 772024: 392025: 27
2007high 118 (2011) · low 16 (2007)2025

MAUDE reports naming product code IZL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007330716
2008449027
20093125035
2010365475
20116666118
2012383524
20130—1017
20147104451
2015488430
2016647033
2017745035
2018847027
2019933240
2020943148
20211348047
20221689132
2023355677
20249159539
2025111534727
202661575—

What is it indicated for?

Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not for mammography

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141895: Mobilediagnost WDR.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141895

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026