510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

MobileDiagnost wDR 2.2

K191813510(k) SpecialCleared Aug 2, 201928-day reviewClass II21 CFR 892.1720Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sedecal., Sa. · IZL System, X-Ray, Mobile

Cleared on Aug 2, 2019 as substantially equivalent to K141895 (Sedecal., Sa., 2014) after a 28-day review. Its family tree reaches back 8 generations to 1998. 1 later device descends from it. It has been recalled 4 times.

Decided
Aug 2, 2019
Review
28 days
Class
Class II
Product code
IZL
Recalls
4
Predicates
1

Its longest line of descent runs back 8 generations to K981556 (1998). 16 ancestors in all, the oldest K931071 (1994). 1 later device descends from it, 1 citing it directly.

  1. Root · 8 generations backK981556 Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc. · 1998
  2. 7 generations backK992547 Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc. · 1999
  3. 6 generations backK023750 Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc. · 2002
  4. 5 generations backK031447 Canon, Digital Radiography, Model Cxdi-50GCanon, Inc. · 2003
  5. 4 generations backK043002 Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa. · 2004
  6. 3 generations backK090322 Easy Moving Plus, SM-XXHF-YYSedecal., Sa. · 2009 · Recalled
  7. 2 generations backK111725 Mobilediagnost WDRSedecal., Sa. · 2011 · Recalled
  8. 1 generation backK141895 Mobilediagnost WDRSedecal., Sa. · 2014 · Recalled
  9. This deviceK191813 MobileDiagnost wDR 2.2Sedecal., Sa. · 2019
  10. Descendant · cites it directly · 0 of its ownK233678 Radiography 7000 MPhilips Medical Systems Dmc GmbH · 2024 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K233678Radiography 7000 MPhilips Medical Systems Dmc GmbH · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K981556Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc.1998—
−7K992547Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc.1999—
−6K023750Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc.2002—
−5K031447Canon, Digital Radiography, Model Cxdi-50GCanon, Inc.2003—
−4K043002Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa.2004—
−3K090322Easy Moving Plus, SM-XXHF-YYSedecal., Sa.200931
−2K111725Mobilediagnost WDRSedecal., Sa.20118
−1K141895Mobilediagnost WDRSedecal., Sa.20142
0 (this device)K191813MobileDiagnost wDR 2.2Sedecal., Sa.2019—
+1K233678Radiography 7000 MPhilips Medical Systems Dmc GmbH20241

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199710
199810
200120
200210
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
FIG. 1 — Lineage of K191813, 1998–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IZL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZL device

  1. #1 K242015 TOPAZ Mobile X-ray System (Models : TOPAZ-32D…score 19.9 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141895Mobilediagnost WDR
Summary page 4
…MobileDiagnost wDR 2.0 Manufacturer: Sedecal SA 510(k) Clearance: K141895-September 18, 2014 Classification Regulation: 21CFR 892.1720 Cl…
Sedecal., Sa.20142
†K141736Philips Prograde; Philips Eleva Workspot for Digital DiagnostPhilips Medical Systems Dmc GmbH20146
†K153318Philips Eleva Workspot with SkyFlowPhilips Medical Systems Dmc GmbH2015—
†K182973DigitalDiagnost C90Philips Medical Systems Dmc GmbH2019—

Also cleared against K141895: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K233678Radiography 7000 MPhilips Medical Systems Dmc GmbH20240

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 10, 2025Class IIZ-0276-2026Device DesignSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper…
Jun 10, 2025Class IIZ-0243-2026Device DesignSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper…
May 23, 2023Class IIZ-2045-2023Software change controlThere is a software login in issue that may prevent the user from logging in.
Oct 14, 2021Class IIZ-1083-2024Use errorWhile cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button.…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 3.1 years on median.
  • 8 of the 16 ancestors have a recall of any class.
  • Of its descendants, K233678 (Philips Medical Systems Dmc GmbH, 2024) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K141895.

How does its category compare?

All IZL devices

IZL System, X-Ray, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · IZL

11in 2025

2007: 32008: 42009: 32010: 32011: 62012: 32014: 72015: 42016: 62017: 72018: 82019: 92020: 92021: 132022: 162023: 32024: 92025: 112026 (partial year): 6
2007high 16 (2022) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZL

153in 2025

2007: 30 days median2008: 49 days median2009: 125 days median2010: 65 days median2011: 66 days median2012: 83 days median2014: 104 days median2015: 88 days median2016: 47 days median2017: 45 days median2018: 47 days median2019: 33 days median2020: 43 days median2021: 48 days median2022: 89 days median2023: 55 days median2024: 159 days median2025: 153 days median2026 (partial year): 157 days median
2007high 159 (2024) · low 30 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZL

47in 2025

2007: 72010: 42011: 62012: 52013: 102014: 42015: 42019: 22020: 12022: 12023: 62024: 52025: 472026 (partial year): 5
2007high 47 (2025) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZL

27in 2025

2007: 162008: 272009: 352010: 752011: 1182012: 242013: 172014: 512015: 302016: 332017: 352018: 272019: 402020: 482021: 472022: 322023: 772024: 392025: 27
2007high 118 (2011) · low 16 (2007)2025

MAUDE reports naming product code IZL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007330716
2008449027
20093125035
2010365475
20116666118
2012383524
20130—1017
20147104451
2015488430
2016647033
2017745035
2018847027
2019933240
2020943148
20211348047
20221689132
2023355677
20249159539
2025111534727
202661575—

What is it indicated for?

Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not for mammography.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K191813: MobileDiagnost wDR 2.2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K191813

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026