Recalled · Class II · 2017Stat Profile Prime CCS Analyzer System, Stat Profile Prime Auto QC Cartridge CCS, Stat Profile Prime Ampuled Control Abg/CCS, Stat Profile Prime Calibrtor Cartridge CCS/CCS Comp, Nova Linearity Standard Set a
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Nova Biomedical Corporation · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Cleared on May 29, 2015 as substantially equivalent to K131703 (Nova Biomedical Corporation, 2014) after a 318-day review. Its family tree reaches back 3 generations to 2006. It has been recalled once.
- Decided
- May 29, 2015
- Review
- 318 days
- Class
- Class II
- Product code
- CHL
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 3 generations to K061830 (2006). No later device cites it as a predicate yet.
- Root · 3 generations backK061830 Stat Profile Critical Care Xpress (CCX), Model 1+Nova Biomedical Corp. · 2006
- 2 generations backK110648 Nova Stat Profile Phox Ultra Analyzer SystemNova Biomedical Corporation · 2011
- 1 generation backK131703 Nova Stat Profile Prime CCS Analyzer SystemNova Biomedical Corporation · 2014 · Recalled
- This deviceK141907 Stat Profile Prime CCS Analyzer System, Stat Profile Prime Auto QC…Nova Biomedical Corporation · 2015
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K061830 | Stat Profile Critical Care Xpress (CCX), Model 1+ | Nova Biomedical Corp. | 2006 | — |
| −2 | K110648 | Nova Stat Profile Phox Ultra Analyzer System | Nova Biomedical Corporation | 2011 | — |
| −1 | K131703 | Nova Stat Profile Prime CCS Analyzer System | Nova Biomedical Corporation | 2014 | 1 |
| 0 (this device) | K141907 | Stat Profile Prime CCS Analyzer System, Stat Profile Prime Auto QC Cartridge… | Nova Biomedical Corporation | 2015 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new CHL device
- #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.1 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K131703 | Nova Stat Profile Prime CCS Analyzer SystemSummary page 5…862.1150 II Quality Control Materials 75JJS 862.1660 I Product Codes: CHL, JGS, CEM, JFP, CGZ, CGA, GKF, JIX, JJS, KHP Predicate Device: K131703 - Nova Stat Profile Prime CCS Analyzer System (including controls, calibrators and linearity standards) Device… | 2014 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 19, 2017 | Class II | Z-2732-2017 | Software design | Software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test results. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 9 years and passes through 2 companies; a new generation was cleared every 2.6 years on median.
- 1 of the 3 ancestors has a recall of any class.
- 1 other device also cites its predicate K131703.
How does its category compare?
All CHL devicesCHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.
FDA decisions per year · CHL
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · CHL
267in 2023
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · CHL
13so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · CHL
210in 2025
MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 6 | 16 |
| 2008 | 1 | — | 5 | 40 |
| 2009 | 0 | — | 6 | 74 |
| 2010 | 1 | — | 7 | 20 |
| 2011 | 4 | 108 | 1 | 50 |
| 2012 | 4 | 65 | 9 | 64 |
| 2013 | 5 | 152 | 1 | 62 |
| 2014 | 3 | 68 | 1 | 93 |
| 2015 | 4 | 155 | 3 | 55 |
| 2016 | 3 | 306 | 1 | 118 |
| 2017 | 1 | — | 1 | 206 |
| 2018 | 1 | — | 3 | 174 |
| 2019 | 0 | — | 0 | 141 |
| 2020 | 4 | 100 | 2 | 117 |
| 2021 | 2 | — | 0 | 147 |
| 2022 | 0 | — | 0 | 138 |
| 2023 | 5 | 267 | 4 | 177 |
| 2024 | 1 | — | 9 | 196 |
| 2025 | 1 | — | 0 | 210 |
| 2026 | 1 | — | 13 | — |
What is it indicated for?
The Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of pH, PCO2, PO2, Hct, Na+. K+, Cl-, iCa, Glu (Glucose), and Lac (Lactate) in heparinized whole blood. PC02,P02,pH: Whole blood measurement of certain gases in whole blood,or pH of whole blood,is used in the diagnosis and treatment of life-threatening…
Read the full statement (377 words)
The Stat Profile Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of pH, PCO2, PO2, Hct, Na+. K+, Cl-, iCa, Glu (Glucose), and Lac (Lactate) in heparinized whole blood.
PC02,P02,pH: Whole blood measurement of certain gases in whole blood,or pH of whole blood,is used in the diagnosis and treatment of life-threatening acid-base disturbances.
Hct: Whole blood measurements of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states,such as anemia and erythrocytosis (an increase in the number of red cells).
Na+: Sodium measurement is used in the diagnosis and treatment of aldosteronism,diabetes insipidus, adrenal hypertension, Addison's disease,dehydration,or diseases involving electrolyte imbalance.
K+: Potassium Measurement is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high potassium levels.
Cl-: Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
iCa: Calcium measurements are used in the diagnosis and treatment of parathyroid disease,a variety of bone diseases,chronic renal disease and tetany (intermittent muscular contractions or spasms).
Glu: Glucose measurement is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus,neonatal hypoglycemia,and idiopathic hypoglycemia,and of pancreatic islet cell carcinoma.
Lac: Lactate (lactic acid) measurement is used to evaluate the acid-base status of patients suspected of having lactic acidosis.
The Stat Profile Prime Auto QC Cartridge CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of the Stat Profile Prime CCS Analyzer.
The Stat Profile Prime Ampuled Control ABG/CCS is a quality control material intended for in vitro diagnostic use by healthcare professionals for monitoring the performance of Stat Profile Prime CCS Analyzer.
The Stat Profile Prime Calibrator Cartridge CCS is intended for the calibration of pH, PCO2, PO2, Hct, Na+, K+, Cl‐, iCa, Glucose and Lactate using the Stat Profile Prime CCS Analyzer.
Linearity Standard Set A is intended for in vitro diagnostic use with Stat Profile Prime CCS Analyzers to verify calibration, analytical linearity, estimate test imprecision, and detect systematic analytical deviations that may arise from calibrator cartridge or analytical instrument variation.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K141907 · printed Nov 6, 2026
Cite this page
510(k) Report. “K141907: Stat Profile Prime CCS Analyzer System, Stat Profile Prime Auto QC Cartridge CCS, Stat Profile Prime Ampuled Control Abg/CCS, Stat Profile Prime Calibrtor Cartridge CCS/CCS Comp, Nova Linearity Standard Set a.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141907
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026