510(k) Report

510(k) · Cleared

Cleared

Nova Allegro HbA1c Assay, Nova Allegro Analyzer

K221326510(k) Dual TrackCleared Nov 22, 2024931-day reviewClass II21 CFR 864.7470

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nova Biomedical Corporation · LCP Assay, Glycosylated Hemoglobin

Cleared on Nov 22, 2024 as substantially equivalent to K171650 (Abbott Rapid Diagnostics, 2017) after a 931-day review. Its family tree reaches back 5 generations to 1995.

Decided
Nov 22, 2024
Review
931 days
Class
Class II
Product code
LCP
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K951361 (1995). 7 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 5 generations backK951361 Dca 2000 System for Hemoglobin A1CHeraeus Kulzer, Inc. · 1995 · Recalled
  2. 4 generations backK050574 Afinion HbalcAxis-Shield Poc AS · 2005
  3. 3 generations backK110056 Afinion AnalyzerAxis-Shield Poc AS · 2011
  4. 2 generations backK151809 Alere Afinion HbA1c, Alere Afinion AS100 AnalyzerAlere Technologies AS · 2015
  5. 1 generation backK171650 Alere Afinion HbA1c and ACR on Afinion 2 analyzerAbbott Rapid Diagnostics · 2017 · Recalled
  6. This deviceK221326 Nova Allegro HbA1c Assay, Nova Allegro AnalyzerNova Biomedical Corporation · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K951361Dca 2000 System for Hemoglobin A1CHeraeus Kulzer, Inc.19951
−4K050574Afinion HbalcAxis-Shield Poc AS2005—
−3K110056Afinion AnalyzerAxis-Shield Poc AS2011—
−2K151809Alere Afinion HbA1c, Alere Afinion AS100 AnalyzerAlere Technologies AS2015—
−1K171650Alere Afinion HbA1c and ACR on Afinion 2 analyzerAbbott Rapid Diagnostics20172
0 (this device)K221326Nova Allegro HbA1c Assay, Nova Allegro AnalyzerNova Biomedical Corporation2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200810
FIG. 1 — Lineage of K221326, 1995–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LCP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LCP device

  1. #1 K231465 Q-Pad Test Systemscore 7.7 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K171650Alere Afinion HbA1c and ACR on Afinion 2 analyzer
Summary page 5
…Code Glycosylated Hemoglobin Assay 864.7470 II LCP Analyzer 862.2400 I JQT Predicate Device: K171650 – Alere Afinion HbA1c Assay on the Afinion 2 Analyzer Device Description: Nova Allegro Analyzer The Nova Allegro Analyzer is a…
Abbott Rapid Diagnostics20172

Also cleared against K171650: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 29 years and passes through 5 companies; a new generation was cleared every 5.6 years on median.
  • 2 of the 7 ancestors have a recall of any class.
  • 1 other device also cites its predicate K171650.

How does its category compare?

All LCP devices

LCP Assay, Glycosylated Hemoglobin: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · LCP

1in 2024

2007: 72008: 52009: 62010: 22011: 102012: 92013: 22014: 82015: 32016: 22017: 52018: 22019: 42020: 12023: 22024: 1
2007high 10 (2011) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LCP

269in 2019

2007: 133 days median2008: 206 days median2009: 158 days median2011: 176 days median2012: 143 days median2014: 127 days median2015: 239 days median2017: 218 days median2019: 269 days median
2007high 269 (2019) · low 127 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LCP

1in 2025

2007: 12008: 32009: 62010: 52011: 512013: 42014: 22015: 22016: 62017: 12018: 12022: 12024: 12025: 1
2007high 51 (2011) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LCP

22in 2025

2007: 52008: 112009: 152010: 42011: 272012: 512013: 1932014: 1822015: 642016: 292017: 1492018: 1,9112019: 1142020: 132021: 72022: 82023: 272024: 392025: 22
2007high 1,911 (2018) · low 4 (2010)2025

MAUDE reports naming product code LCP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007713315
20085206311
20096158615
20102—54
2011101765127
20129143051
20132—4193
201481272182
20153239264
20162—629
201752181149
20182—11,911
201942690114
20201—013
20210—07
20220—18
20232—027
20241—139
20250—122
20260—0—

What is it indicated for?

The Nova Allegro HbA1c Assay is intended for in vitro diagnostic use on the Nova Allegro Analyzer for the quantitative determination of the glycated Hemoglobin (% hemoglobin A1c) in capillary whole blood obtained from the fingertip. The results from this assay are intended to be used for the monitoring of long-term blood glucose/metabolic control in individuals with diabetes mellitus.

The Nova Allegro Analyzer is intended for in vitro diagnostic use in clinical laboratory and near-patient testing (point-of-care) settings for the quantitative determination of Nova Allegro Assays using Nova Allegro Test Cartridges.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K221326: Nova Allegro HbA1c Assay, Nova Allegro Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K221326

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026