510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

GM60A-32S, GM60A-40S

K142492510(k) TraditionalCleared Dec 17, 2014104-day reviewClass II21 CFR 892.1720Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Samsung Electronics Co., Ltd. · IZL System, X-Ray, Mobile

Cleared on Dec 17, 2014 as substantially equivalent to K090322 (Sedecal., Sa., 2009) and 1 other predicate after a 104-day review. Its family tree reaches back 6 generations to 1998. 25 later devices descend from it. It has been recalled once.

Decided
Dec 17, 2014
Review
104 days
Class
Class II
Product code
IZL
Recalls
1
Predicates
2

Its longest line of descent runs back 6 generations to K981556 (1998). 16 ancestors in all, the oldest K862120 (1986). 25 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2014: 1 other ancestor2013: 1 other ancestor2001: 3 other ancestors2002: 1 other ancestor1999: 1 other ancestor1998: 1 other ancestor1988: 1 other ancestor1986: 1 other ancestor2015: 1 descendant cleared2016: 1 descendant cleared2017: 1 descendant cleared2018: 8 descendants cleared2024: 4 descendants cleared2026: 2 descendants cleared2021: 2 descendants cleared2019: 1 descendant cleared2022: 2 descendants cleared2020: 1 descendant cleared2023: 1 descendant cleared2025: 1 descendant clearedK981556 · Canon X-Ray Digital Camera Cxdi-11 · Canon USA, Inc. · 1998K992547 · Canon X-Ray Digital Camera, Model Cxdi-22 · Canon USA, Inc. · 1999K023750 · Cannon Digital Radiography Model Cxdi-40G · Canon USA, Inc. · 2002K031447 · Canon, Digital Radiography, Model Cxdi-50G · Canon, Inc. · 2003K043002 · Sedecal Mobile X-Ray Units with Digital Detector · Sedecal., Sa. · 2004K090322 · Easy Moving Plus, SM-XXHF-YY · Sedecal., Sa. · 2009 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors3 generations back: 2 other ancestors4 generations back: 2 other ancestors5 generations back: 2 other ancestors6 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 4 descendants5 generations forward: 7 descendants6 generations forward: 3 descendants7 generations forward: 3 descendants8 generations forward: 3 descendants9 generations forward: 2 descendants K090322 · Easy Moving Plus, SM-XXHF-YY · Sedecal., Sa. · 2009 · recalledK043002 · Sedecal Mobile X-Ray Units with Digital Detector · Sedecal., Sa. · 2004K031447 · Canon, Digital Radiography, Model Cxdi-50G · Canon, Inc. · 2003K023750 · Cannon Digital Radiography Model Cxdi-40G · Canon USA, Inc. · 2002K992547 · Canon X-Ray Digital Camera, Model Cxdi-22 · Canon USA, Inc. · 1999K981556 · Canon X-Ray Digital Camera Cxdi-11 · Canon USA, Inc. · 1998−50+5
  1. Root · 6 generations backK981556 Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc. · 1998
  2. 5 generations backK992547 Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc. · 1999
  3. 4 generations backK023750 Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc. · 2002
  4. 3 generations backK031447 Canon, Digital Radiography, Model Cxdi-50GCanon, Inc. · 2003
  5. 2 generations backK043002 Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa. · 2004
  6. 1 generation backK090322 Easy Moving Plus, SM-XXHF-YYSedecal., Sa. · 2009 · Recalled
  7. This deviceK142492 GM60A-32S, GM60A-40SSamsung Electronics Co., Ltd. · 2014
  8. Descendant · cites it directly · 24 of its ownK150097 GM60A-32S & GM60A-40SSamsung Electronics Co., Ltd. · 2015 · Recalled
  9. Descendant · 2 generations on · 23 of its ownK162278 Digital Diagnostic Mobile X-ray SystemSamsung Electronics Co., Ltd. · 2016
  10. Descendant · 3 generations on · 22 of its ownK171119 GM85Samsung Electronics Co., Ltd. · 2017 · Recalled
  11. Descendant · 4 generations on · 19 of its ownK173828 GU60A & GU60A-65Samsung Electronics Co., Ltd. · 2018
  12. Descendant · 4 generations on · 18 of its ownK180543 GC70, GU60A&GU60A-65, GF50, GF50A, GR40CW, and GM85Samsung Electronics Co., Ltd. · 2018

Showing 5 of 25 descendants. See all descendants

Descendants cleared per year

2015: 12016: 12017: 12018: 82019: 12020: 12021: 22022: 22023: 12024: 42025: 12026: 2
2015high 8 (2018) · low 1 (2015)2026

Most influential descendants

Showing 5 of 25 · See all
  • K150097GM60A-32S & GM60A-40SSamsung Electronics Co., Ltd. · 2015 · cites it directly24descendants
  • K162278Digital Diagnostic Mobile X-ray SystemSamsung Electronics Co., Ltd. · 2016 · 2 generations on23descendants
  • K171119GM85Samsung Electronics Co., Ltd. · 2017 · 3 generations on22descendants
  • K173828GU60A & GU60A-65Samsung Electronics Co., Ltd. · 2018 · 4 generations on19descendants
  • K180543GC70, GU60A&GU60A-65, GF50, GF50A, GR40CW, and GM85Samsung Electronics Co., Ltd. · 2018 · 4 generations on18descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K981556Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc.1998—
−5K992547Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc.1999—
−4K023750Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc.2002—
−3K031447Canon, Digital Radiography, Model Cxdi-50GCanon, Inc.2003—
−2K043002Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa.2004—
−1K090322Easy Moving Plus, SM-XXHF-YYSedecal., Sa.200931
0 (this device)K142492GM60A-32S, GM60A-40SSamsung Electronics Co., Ltd.2014—
+1K150097GM60A-32S & GM60A-40SSamsung Electronics Co., Ltd.20152
+2K162278Digital Diagnostic Mobile X-ray SystemSamsung Electronics Co., Ltd.2016—
+3K171119GM85Samsung Electronics Co., Ltd.20171
+4K173828GU60A & GU60A-65Samsung Electronics Co., Ltd.2018—
+4K180543GC70, GU60A&GU60A-65, GF50, GF50A, GR40CW, and GM85Samsung Electronics Co., Ltd.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198810
199810
199910
200130
200210
201310
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201610
201710
201880
201910
202010
202120
202220
202310
202440
202510
202620
FIG. 1 — Lineage of K142492, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IZL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZL device

  1. #1 K242015 TOPAZ Mobile X-ray System (Models : TOPAZ-32D…score 19.9 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090322Easy Moving Plus, SM-XXHF-YY
Summary page 5
…Classification Name: Mobile X-ray system, Solid State X-Ray Imager(Flat Panel/Digital Imager) D. Product Code: IZL E. 510(k) Number: K090322 F. 510(k) Decision Date: March 17, 2009 6.2 Predicate Device II G. Manufacturer: SAMSUNG ELECTRONI…
Sedecal., Sa.200931
*K140334Xgeo GC80
Summary page 5
…S Co., Ltd. H. Trade Name: XGEO GC80 I. Classification Name: System, X-ray, Stationary J. Product Code: KPR K. 510(k) Number: K140334 L. 510(k) Decision Date: May 28, 2014 7. Device Description The GM60A Digital Mobile X-ray imaging syste…
Samsung Electronics Co., Ltd.2014—

Also cleared against K090322: 3 other devices →

What descends from it? (25)

See all 25

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K150097GM60A-32S & GM60A-40SSamsung Electronics Co., Ltd.201524

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 14, 2015Class IIZ-0263-2016Device DesignWhen the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 3.1 years on median.
  • 5 of the 16 ancestors have a recall of any class.
  • Of its descendants, K150097 (Samsung Electronics Co., Ltd., 2015) has gone furthest, with 24 descendants of its own.
  • 3 other devices also cite its predicate K090322.

How does its category compare?

All IZL devices

IZL System, X-Ray, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · IZL

11in 2025

2007: 32008: 42009: 32010: 32011: 62012: 32014: 72015: 42016: 62017: 72018: 82019: 92020: 92021: 132022: 162023: 32024: 92025: 112026 (partial year): 6
2007high 16 (2022) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZL

153in 2025

2007: 30 days median2008: 49 days median2009: 125 days median2010: 65 days median2011: 66 days median2012: 83 days median2014: 104 days median2015: 88 days median2016: 47 days median2017: 45 days median2018: 47 days median2019: 33 days median2020: 43 days median2021: 48 days median2022: 89 days median2023: 55 days median2024: 159 days median2025: 153 days median2026 (partial year): 157 days median
2007high 159 (2024) · low 30 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZL

47in 2025

2007: 72010: 42011: 62012: 52013: 102014: 42015: 42019: 22020: 12022: 12023: 62024: 52025: 472026 (partial year): 5
2007high 47 (2025) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZL

27in 2025

2007: 162008: 272009: 352010: 752011: 1182012: 242013: 172014: 512015: 302016: 332017: 352018: 272019: 402020: 482021: 472022: 322023: 772024: 392025: 27
2007high 118 (2011) · low 16 (2007)2025

MAUDE reports naming product code IZL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007330716
2008449027
20093125035
2010365475
20116666118
2012383524
20130—1017
20147104451
2015488430
2016647033
2017745035
2018847027
2019933240
2020943148
20211348047
20221689132
2023355677
20249159539
2025111534727
202661575—

What is it indicated for?

The GM60A Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142492: GM60A-32S, GM60A-40S.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142492

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026