510(k) · Cleared
Recalled · Class II · 2024DX-D 100
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Agfa Healthcare N.V. · IZL System, X-Ray, Mobile
Cleared on Jan 3, 2011 as substantially equivalent to K090322 (Sedecal., Sa., 2009) and 1 other predicate after a 26-day review. Its family tree reaches back 6 generations to 1998. 1 later device descends from it. It has been recalled 6 times.
- Decided
- Jan 3, 2011
- Review
- 26 days
- Class
- Class II
- Product code
- IZL
- Recalls
- 6
- Predicates
- 2
Its longest line of descent runs back 6 generations to K981556 (1998). 30 ancestors in all, the oldest K803130 (1981). 1 later device descends from it, 1 citing it directly.
- Root · 6 generations backK981556 Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc. · 1998
- 5 generations backK992547 Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc. · 1999
- 4 generations backK023750 Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc. · 2002
- 3 generations backK031447 Canon, Digital Radiography, Model Cxdi-50GCanon, Inc. · 2003
- 2 generations backK043002 Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa. · 2004
- 1 generation backK090322 Easy Moving Plus, SM-XXHF-YYSedecal., Sa. · 2009 · Recalled
- This deviceK103597 DX-D 100Agfa Healthcare N.V. · 2011
- Descendant · cites it directly · 0 of its ownK191884 DR 100sAgfa N.V. · 2019
Most influential descendants
Showing 1 of 1 · See all- K191884DR 100sAgfa N.V. · 2019 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K981556 | Canon X-Ray Digital Camera Cxdi-11 | Canon USA, Inc. | 1998 | — |
| −5 | K992547 | Canon X-Ray Digital Camera, Model Cxdi-22 | Canon USA, Inc. | 1999 | — |
| −4 | K023750 | Cannon Digital Radiography Model Cxdi-40G | Canon USA, Inc. | 2002 | — |
| −3 | K031447 | Canon, Digital Radiography, Model Cxdi-50G | Canon, Inc. | 2003 | — |
| −2 | K043002 | Sedecal Mobile X-Ray Units with Digital Detector | Sedecal., Sa. | 2004 | — |
| −1 | K090322 | Easy Moving Plus, SM-XXHF-YY | Sedecal., Sa. | 2009 | 31 |
| 0 (this device) | K103597 | DX-D 100 | Agfa Healthcare N.V. | 2011 | — |
| +1 | K191884 | DR 100s | Agfa N.V. | 2019 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1981 | 1 | 0 |
| 1986 | 1 | 0 |
| 1988 | 1 | 0 |
| 1992 | 1 | 0 |
| 1994 | 2 | 0 |
| 1995 | 2 | 0 |
| 1997 | 2 | 0 |
| 1998 | 2 | 0 |
| 1999 | 2 | 0 |
| 2000 | 1 | 0 |
| 2001 | 2 | 0 |
| 2002 | 1 | 0 |
| 2003 | 1 | 0 |
| 2006 | 1 | 0 |
| 2007 | 1 | 0 |
| 2008 | 1 | 0 |
| 2009 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2019 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in IZL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IZL device
- #1 K242015 TOPAZ Mobile X-ray System (Models : TOPAZ-32D…score 19.9 · 19 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K090322 | Easy Moving Plus, SM-XXHF-YYSummary page 1…al mobile x-ray system with digital image capture equipment. The new device is essentially a modification of the predicate mobile x-ray system, Sedecal Easy Moving Plus (K090322). In this case, the digital detector available with the predicate has been replaced wit Agfa's system, its previously cleared DX-… | 2009 |
| * | K092669 | DX-D Imaging PackageSummary page 1…Moving Plus (K090322). In this case, the digital detector available with the predicate has been replaced wit Agfa's system, its previously cleared DX-D Imaging Package (K092669). The DX-D 100 uses Agfa's familiar NX workstation with MTUSICA 2 TM image processing and flat panel detectors of the scintillato… | 2009 |
| † | K051901 | NIO 3MP-21 | 2005 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K191884 | DR 100s | 2019 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 23, 2024 | Class II | Z-1462-2024 | Component design/selection | Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work… |
| Jan 20, 2016 | Class II | Z-1698-2017 | Other | The DX-D100 User Manual already contained information about to move a DX-D100 mobile unit manually by using an Allen wrench, but did not indicate where the Allen wrench should be stored. |
| May 29, 2015 | Class II | Z-1745-2015 | Other | When using the DX-D 100 unit, sporadic unintended movements caused by an electrostatic discharge of the unit to the ground may occur. |
| Jun 6, 2014 | Class II | Z-2081-2014 | Use error | When liquid comes in contact with the DX-D 100 touch screen, the device may incorrectly recognize this as user input altering device settings. |
| May 13, 2014 | Class III | Z-2176-2014 | Labeling mix-ups | Unit was mislabeled with a factory label showing 32 KW output power. |
| May 15, 2013 | Class II | Z-1487-2013 | Under Investigation by firm | Customers could potentially experience intermittent, unintended and illogical movement when using the product. |
The 5 latest of 6 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 1.8 years on median.
- 15 of the 30 ancestors have a recall of any class.
- Of its descendants, K191884 (Agfa N.V., 2019) has gone furthest, with 0 descendants of its own.
- 3 other devices also cite its predicate K090322.
How does its category compare?
All IZL devicesIZL System, X-Ray, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · IZL
11in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IZL
153in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IZL
47in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · IZL
27in 2025
MAUDE reports naming product code IZL, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 30 | 7 | 16 |
| 2008 | 4 | 49 | 0 | 27 |
| 2009 | 3 | 125 | 0 | 35 |
| 2010 | 3 | 65 | 4 | 75 |
| 2011 | 6 | 66 | 6 | 118 |
| 2012 | 3 | 83 | 5 | 24 |
| 2013 | 0 | — | 10 | 17 |
| 2014 | 7 | 104 | 4 | 51 |
| 2015 | 4 | 88 | 4 | 30 |
| 2016 | 6 | 47 | 0 | 33 |
| 2017 | 7 | 45 | 0 | 35 |
| 2018 | 8 | 47 | 0 | 27 |
| 2019 | 9 | 33 | 2 | 40 |
| 2020 | 9 | 43 | 1 | 48 |
| 2021 | 13 | 48 | 0 | 47 |
| 2022 | 16 | 89 | 1 | 32 |
| 2023 | 3 | 55 | 6 | 77 |
| 2024 | 9 | 159 | 5 | 39 |
| 2025 | 11 | 153 | 47 | 27 |
| 2026 | 6 | 157 | 5 | — |
What is it indicated for?
Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis.
Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton (including skull, spinal column and extremities) chest, abdomen and other body parts.
Agfa's DX-D 100 is not indicated for use in mammography.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K103597 · printed Nov 6, 2026
Cite this page
510(k) Report. “K103597: DX-D 100.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K103597
Report an extraction error
Are you the manufacturer?