510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

DX-D 100

K103597510(k) TraditionalCleared Jan 3, 201126-day reviewClass II21 CFR 892.1720Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Agfa Healthcare N.V. · IZL System, X-Ray, Mobile

Cleared on Jan 3, 2011 as substantially equivalent to K090322 (Sedecal., Sa., 2009) and 1 other predicate after a 26-day review. Its family tree reaches back 6 generations to 1998. 1 later device descends from it. It has been recalled 6 times.

Decided
Jan 3, 2011
Review
26 days
Class
Class II
Product code
IZL
Recalls
6
Predicates
2

Its longest line of descent runs back 6 generations to K981556 (1998). 30 ancestors in all, the oldest K803130 (1981). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2009: 2 other ancestors2003: 1 other ancestor2001: 2 other ancestors2008: 1 other ancestor2000: 1 other ancestor1999: 2 other ancestors1998: 2 other ancestors2007: 1 other ancestor2002: 1 other ancestor1997: 2 other ancestors1995: 2 other ancestors1994: 2 other ancestors2006: 1 other ancestor1992: 1 other ancestor1986: 1 other ancestor1988: 1 other ancestor1981: 1 other ancestor2019: 1 descendant clearedReference device: K051901 · NIO 3MP-21 · Barcoview · 2005K981556 · Canon X-Ray Digital Camera Cxdi-11 · Canon USA, Inc. · 1998K992547 · Canon X-Ray Digital Camera, Model Cxdi-22 · Canon USA, Inc. · 1999K023750 · Cannon Digital Radiography Model Cxdi-40G · Canon USA, Inc. · 2002K031447 · Canon, Digital Radiography, Model Cxdi-50G · Canon, Inc. · 2003K043002 · Sedecal Mobile X-Ray Units with Digital Detector · Sedecal., Sa. · 2004K090322 · Easy Moving Plus, SM-XXHF-YY · Sedecal., Sa. · 2009 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 5 other ancestors4 generations back: 9 other ancestors5 generations back: 4 other ancestors6 generations back: 2 other ancestors 1 generation forward: 1 descendant K090322 · Easy Moving Plus, SM-XXHF-YY · Sedecal., Sa. · 2009 · recalledK043002 · Sedecal Mobile X-Ray Units with Digital Detector · Sedecal., Sa. · 2004K031447 · Canon, Digital Radiography, Model Cxdi-50G · Canon, Inc. · 2003K023750 · Cannon Digital Radiography Model Cxdi-40G · Canon USA, Inc. · 2002K992547 · Canon X-Ray Digital Camera, Model Cxdi-22 · Canon USA, Inc. · 1999K981556 · Canon X-Ray Digital Camera Cxdi-11 · Canon USA, Inc. · 1998−50
  1. Root · 6 generations backK981556 Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc. · 1998
  2. 5 generations backK992547 Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc. · 1999
  3. 4 generations backK023750 Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc. · 2002
  4. 3 generations backK031447 Canon, Digital Radiography, Model Cxdi-50GCanon, Inc. · 2003
  5. 2 generations backK043002 Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa. · 2004
  6. 1 generation backK090322 Easy Moving Plus, SM-XXHF-YYSedecal., Sa. · 2009 · Recalled
  7. This deviceK103597 DX-D 100Agfa Healthcare N.V. · 2011
  8. Descendant · cites it directly · 0 of its ownK191884 DR 100sAgfa N.V. · 2019

Most influential descendants

Showing 1 of 1 · See all
  • K191884DR 100sAgfa N.V. · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K981556Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc.1998—
−5K992547Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc.1999—
−4K023750Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc.2002—
−3K031447Canon, Digital Radiography, Model Cxdi-50GCanon, Inc.2003—
−2K043002Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa.2004—
−1K090322Easy Moving Plus, SM-XXHF-YYSedecal., Sa.200931
0 (this device)K103597DX-D 100Agfa Healthcare N.V.2011—
+1K191884DR 100sAgfa N.V.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198610
198810
199210
199420
199520
199720
199820
199920
200010
200120
200210
200310
200610
200710
200810
200920

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
FIG. 1 — Lineage of K103597, 1998–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IZL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZL device

  1. #1 K242015 TOPAZ Mobile X-ray System (Models : TOPAZ-32D…score 19.9 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090322Easy Moving Plus, SM-XXHF-YY
Summary page 1
…al mobile x-ray system with digital image capture equipment. The new device is essentially a modification of the predicate mobile x-ray system, Sedecal Easy Moving Plus (K090322). In this case, the digital detector available with the predicate has been replaced wit Agfa's system, its previously cleared DX-…
Sedecal., Sa.200931
*K092669DX-D Imaging Package
Summary page 1
…Moving Plus (K090322). In this case, the digital detector available with the predicate has been replaced wit Agfa's system, its previously cleared DX-D Imaging Package (K092669). The DX-D 100 uses Agfa's familiar NX workstation with MTUSICA 2 TM image processing and flat panel detectors of the scintillato…
Agfa Healthcare N.V.20091
†K051901NIO 3MP-21Barcoview2005—

Also cleared against K090322: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191884DR 100sAgfa N.V.20190

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 23, 2024Class IIZ-1462-2024Component design/selectionSteel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work…
Jan 20, 2016Class IIZ-1698-2017OtherThe DX-D100 User Manual already contained information about to move a DX-D100 mobile unit manually by using an Allen wrench, but did not indicate where the Allen wrench should be stored.
May 29, 2015Class IIZ-1745-2015OtherWhen using the DX-D 100 unit, sporadic unintended movements caused by an electrostatic discharge of the unit to the ground may occur.
Jun 6, 2014Class IIZ-2081-2014Use errorWhen liquid comes in contact with the DX-D 100 touch screen, the device may incorrectly recognize this as user input altering device settings.
May 13, 2014Class IIIZ-2176-2014Labeling mix-upsUnit was mislabeled with a factory label showing 32 KW output power.
May 15, 2013Class IIZ-1487-2013Under Investigation by firmCustomers could potentially experience intermittent, unintended and illogical movement when using the product.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 1.8 years on median.
  • 15 of the 30 ancestors have a recall of any class.
  • Of its descendants, K191884 (Agfa N.V., 2019) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K090322.

How does its category compare?

All IZL devices

IZL System, X-Ray, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · IZL

11in 2025

2007: 32008: 42009: 32010: 32011: 62012: 32014: 72015: 42016: 62017: 72018: 82019: 92020: 92021: 132022: 162023: 32024: 92025: 112026 (partial year): 6
2007high 16 (2022) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZL

153in 2025

2007: 30 days median2008: 49 days median2009: 125 days median2010: 65 days median2011: 66 days median2012: 83 days median2014: 104 days median2015: 88 days median2016: 47 days median2017: 45 days median2018: 47 days median2019: 33 days median2020: 43 days median2021: 48 days median2022: 89 days median2023: 55 days median2024: 159 days median2025: 153 days median2026 (partial year): 157 days median
2007high 159 (2024) · low 30 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZL

47in 2025

2007: 72010: 42011: 62012: 52013: 102014: 42015: 42019: 22020: 12022: 12023: 62024: 52025: 472026 (partial year): 5
2007high 47 (2025) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZL

27in 2025

2007: 162008: 272009: 352010: 752011: 1182012: 242013: 172014: 512015: 302016: 332017: 352018: 272019: 402020: 482021: 472022: 322023: 772024: 392025: 27
2007high 118 (2011) · low 16 (2007)2025

MAUDE reports naming product code IZL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007330716
2008449027
20093125035
2010365475
20116666118
2012383524
20130—1017
20147104451
2015488430
2016647033
2017745035
2018847027
2019933240
2020943148
20211348047
20221689132
2023355677
20249159539
2025111534727
202661575—

What is it indicated for?

Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis.

Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton (including skull, spinal column and extremities) chest, abdomen and other body parts.

Agfa's DX-D 100 is not indicated for use in mammography.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K103597: DX-D 100.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K103597

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026