510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Genesis Reusable Rigid Container System

K142529510(k) TraditionalCleared Jan 8, 2015121-day reviewClass II21 CFR 880.6850Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Carefusion 2200, Inc. · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Jan 8, 2015 as substantially equivalent to K140850 (Carefusion 2200, Inc., 2014) after a 121-day review. Its family tree reaches back 4 generations to 1984. 2 later devices descend from it. It has been recalled 2 times.

Decided
Jan 8, 2015
Review
121 days
Class
Class II
Product code
KCT
Recalls
2
Predicates
1

Its longest line of descent runs back 4 generations to K844652 (1984). 6 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor2000: 1 other ancestor2016: 2 descendants clearedK844652 · C.a.S.E. Container · General Medical Co. · 1984K983299 · Allegiance Genesis Container System · Allegiance Healthcare Corp. · 2000K112535 · Genesis Reusable Rigid Container System · Carefusion 2200, Inc. · 2012K140850 · Genesis Reusable Rigid Container System · Carefusion 2200, Inc. · 201419801990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor4 generations back: 1 other ancestor 1 generation forward: 2 descendants K140850 · Genesis Reusable Rigid Container System · Carefusion 2200, Inc. · 2014K112535 · Genesis Reusable Rigid Container System · Carefusion 2200, Inc. · 2012K983299 · Allegiance Genesis Container System · Allegiance Healthcare Corp. · 2000K844652 · C.a.S.E. Container · General Medical Co. · 19840
  1. Root · 4 generations backK844652 C.a.S.E. ContainerGeneral Medical Co. · 1984
  2. 3 generations backK983299 Allegiance Genesis Container SystemAllegiance Healthcare Corp. · 2000
  3. 2 generations backK112535 Genesis Reusable Rigid Container SystemCarefusion 2200, Inc. · 2012
  4. 1 generation backK140850 Genesis Reusable Rigid Container SystemCarefusion 2200, Inc. · 2014
  5. This deviceK142529 Genesis Reusable Rigid Container SystemCarefusion 2200, Inc. · 2015
  6. Descendant · cites it directly · 0 of its ownK162600 Jewel Precision Reusable Rigid Sterilization Container SystemJewel Precision Sheet Metal & Machine, Inc. · 2016 · Recalled
  7. Descendant · cites it directly · 0 of its ownK153554 Genesis Low Temperature Reusable Rigid Container SystemCarefusion 2200, Inc. · 2016

Most influential descendants

Showing 2 of 2 · See all
  • K162600Jewel Precision Reusable Rigid Sterilization Container SystemJewel Precision Sheet Metal & Machine, Inc. · 2016 · cites it directly0descendants
  • K153554Genesis Low Temperature Reusable Rigid Container SystemCarefusion 2200, Inc. · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K844652C.a.S.E. ContainerGeneral Medical Co.1984—
−3K983299Allegiance Genesis Container SystemAllegiance Healthcare Corp.2000—
−2K112535Genesis Reusable Rigid Container SystemCarefusion 2200, Inc.2012—
−1K140850Genesis Reusable Rigid Container SystemCarefusion 2200, Inc.2014—
0 (this device)K142529Genesis Reusable Rigid Container SystemCarefusion 2200, Inc.2015—
+1K162600Jewel Precision Reusable Rigid Sterilization Container SystemJewel Precision Sheet Metal & Machine, Inc.20161
+1K153554Genesis Low Temperature Reusable Rigid Container SystemCarefusion 2200, Inc.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201620
FIG. 1 — Lineage of K142529, 1984–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140850Genesis Reusable Rigid Container System
Summary page 10
…21CFR 880.6850, Product code KCT Product Code(s) KCT Legally marketed Genesis™ Reusable Sterilization Container System device(s) to which K140850 equivalence is claimed Reason for 510(k) Expand indications for use regarding the quantity and types of lumen devices s…
Carefusion 2200, Inc.2014—

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162600Jewel Precision Reusable Rigid Sterilization Container SystemJewel Precision Sheet Metal & Machine, Inc.20160
K153554Genesis Low Temperature Reusable Rigid Container SystemCarefusion 2200, Inc.20160

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 30, 2023Class IIZ-1236-2023Packaging change controlProduct did not meet shelf-life testing requirements resulting in a breach of the sterility of the Genesis container holding surgical equipment.
Sep 26, 2022Class IIZ-0203-2023Under Investigation by firmGenesis STERRAD Reusable Rigid Sterilization Containers have not consistently met the requirements for aerosol challenge testing.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 3 companies; a new generation was cleared every 11.7 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K162600 (Jewel Precision Sheet Metal & Machine, Inc., 2016) has gone furthest, with 0 descendants of its own.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The Genesis Reusable Rigid Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It allows sterilization of the enclosed medical device and maintains sterility of the enclosed medical device until used for a maximum of 180 days. Containers are suitable for dynamic air removal (pre-vacuum) steam sterilization, immediate use pre-vacuum steam sterilization and…

Read the full statement (189 words)

The Genesis Reusable Rigid Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It allows sterilization of the enclosed medical device and maintains sterility of the enclosed medical device until used for a maximum of 180 days.

Containers are suitable for dynamic air removal (pre-vacuum) steam sterilization, immediate use pre-vacuum steam sterilization and 100% ethylene oxide sterilization when used as described in the instructions for use.

Reusable baskets and accessory items (pins, dividers, mats, etc.) are intended to organize and secure enclosed medical devices during sterilization and storage of the container.

Data cards are used to record information regarding a specific sterilization process load. Filter media allows ingress and egress of sterilant while providing a microbial barrier. Tamper evident arrows provide a visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains a modality-specific external process indicator that serves as a visual indication that the system has been exposed to a specific sterilization cycle parameter. Data cards, filters and tamper evident arrows are single use only.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K142529: Genesis Reusable Rigid Container System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142529

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026