510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Jewel Precision Reusable Rigid Sterilization Container System

K162600510(k) TraditionalCleared Dec 9, 201681-day reviewClass II21 CFR 880.6850Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jewel Precision Sheet Metal & Machine, Inc. · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Dec 9, 2016 as substantially equivalent to K142529 (Carefusion 2200, Inc., 2015) after an 81-day review. Its family tree reaches back 5 generations to 1984. It has been recalled once.

Decided
Dec 9, 2016
Review
81 days
Class
Class II
Product code
KCT
Recalls
1
Predicates
1

Its longest line of descent runs back 5 generations to K844652 (1984). 7 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 5 generations backK844652 C.a.S.E. ContainerGeneral Medical Co. · 1984
  2. 4 generations backK983299 Allegiance Genesis Container SystemAllegiance Healthcare Corp. · 2000
  3. 3 generations backK112535 Genesis Reusable Rigid Container SystemCarefusion 2200, Inc. · 2012
  4. 2 generations backK140850 Genesis Reusable Rigid Container SystemCarefusion 2200, Inc. · 2014
  5. 1 generation backK142529 Genesis Reusable Rigid Container SystemCarefusion 2200, Inc. · 2015 · Recalled
  6. This deviceK162600 Jewel Precision Reusable Rigid Sterilization Container SystemJewel Precision Sheet Metal & Machine, Inc. · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K844652C.a.S.E. ContainerGeneral Medical Co.1984—
−4K983299Allegiance Genesis Container SystemAllegiance Healthcare Corp.2000—
−3K112535Genesis Reusable Rigid Container SystemCarefusion 2200, Inc.2012—
−2K140850Genesis Reusable Rigid Container SystemCarefusion 2200, Inc.2014—
−1K142529Genesis Reusable Rigid Container SystemCarefusion 2200, Inc.20152
0 (this device)K162600Jewel Precision Reusable Rigid Sterilization Container SystemJewel Precision Sheet Metal & Machine, Inc.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200210
FIG. 1 — Lineage of K162600, 1984–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K142529Genesis Reusable Rigid Container System
Summary page 4
…21CFR 880.6850 Product Code(s) KCT Legally Marketed Device to Genesis Reusable Rigid Sterilization Container System which equivalence is claimed K142529 Reason for 51 O(k) submission To obtain clearance to market for a new medical device Device Description The Jewel…
Carefusion 2200, Inc.20152
†K800123Dextex II Sterilization WrapAhlstrom Dexter1980—

Also cleared against K142529: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 22, 2025Class IIZ-2026-2025No Marketing ApplicationThe Instructions for Use do not match the 510(k) FDA cleared indications for use or sterility shelf life.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 32 years and passes through 4 companies; a new generation was cleared every 2.3 years on median.
  • 1 of the 7 ancestors has a recall of any class.
  • 1 other device also cites its predicate K142529.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The Jewel Precision Reusable Rigid Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It allows sterilization of the enclosed medical device and maintains sterility of the enclosed device until used, for a maximum of 30 days. Jewel Precision Reusable Rigid Sterilization Containers are suitable for dynamic air removal (pre-vacuum) steam sterilization at 270 °F…

Read the full statement (141 words)

The Jewel Precision Reusable Rigid Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It allows sterilization of the enclosed medical device and maintains sterility of the enclosed device until used, for a maximum of 30 days.

Jewel Precision Reusable Rigid Sterilization Containers are suitable for dynamic air removal (pre-vacuum) steam sterilization at 270 °F (132 °C) for 4 minutes with a drying time of 30 minutes.

Reusable containers, covers, filter holders, insert trays, and accessory items such as brackets of various configuration, dividers, caddies, and silicone nipple mats are intended to organize and secure the enclosed medical devices during sterilization and storage of the container.

Filter media allows ingress and egress of sterilant gas while providing a microbial barrier. Filter media is single use only.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K162600: Jewel Precision Reusable Rigid Sterilization Container System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162600

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026