510(k) · Cleared
Recalled · Class II · 2022VS3-IR-MMS System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Visionsense, Ltd. · OWN Confocal Optical Imaging
Cleared on Jun 17, 2015 as substantially equivalent to K063345 (Novadaq Technologies, Inc., 2007) after a 163-day review. Its family tree reaches back 4 generations to 1994. 35 later devices descend from it. It has been recalled 3 times.
- Decided
- Jun 17, 2015
- Review
- 163 days
- Class
- Class II
- Product code
- OWN
- Recalls
- 3
- Predicates
- 1
Its longest line of descent runs back 4 generations to K944261 (1994). 6 ancestors in all. 35 later devices descend from it, 2 citing it directly.
- Root · 4 generations backK944261 Retina AngiographHeidelberg Engineering · 1994
- 3 generations backK042961 Spy Intra-Operative Imaging System: SP2000Novadaq Technologies, Inc. · 2005 · Recalled
- 2 generations backK060867 Spy Intraoperative Imaging SystemNovadaq Technologies, Inc. · 2006 · Recalled
- 1 generation backK063345 Spy Imaging SystemNovadaq Technologies, Inc. · 2007 · Recalled
- This deviceK150018 VS3-IR-MMS SystemVisionsense, Ltd. · 2015
- Descendant · cites it directly · 34 of its ownK152204 VS3-IR systemVisionsense, Ltd. · 2015
- Descendant · 2 generations on · 32 of its ownK162882 KARL STORZ ICG Imaging SystemKARL STORZ Endoscopy-America, Inc. · 2017
- Descendant · 3 generations on · 31 of its ownK171238 KARL STORZ ICG Imaging SystemKARL STORZ Endoscopy-America, Inc. · 2017
- Descendant · 3 generations on · 30 of its ownK180146 KARL STORZ ICG Imaging SystemKarl Storz Endoscopy America, Inc. · 2018
- Descendant · 4 generations on · 19 of its ownK211202 1688 4K Camera System, L11 LED Light Source with AIM, AIM SafeLight…Stryker · 2021
Showing 5 of 35 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 35 · See all- K152204VS3-IR systemVisionsense, Ltd. · 2015 · cites it directly34descendants
- K162882KARL STORZ ICG Imaging SystemKARL STORZ Endoscopy-America, Inc. · 2017 · 2 generations on32descendants
- K171238KARL STORZ ICG Imaging SystemKARL STORZ Endoscopy-America, Inc. · 2017 · 3 generations on31descendants
- K180146KARL STORZ ICG Imaging SystemKarl Storz Endoscopy America, Inc. · 2018 · 3 generations on30descendants
- K2112021688 4K Camera System, L11 LED Light Source with AIM, AIM SafeLight Cable, Precision S…Stryker · 2021 · 4 generations on19descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K944261 | Retina Angiograph | Heidelberg Engineering | 1994 | — |
| −3 | K042961 | Spy Intra-Operative Imaging System: SP2000 | Novadaq Technologies, Inc. | 2005 | 1 |
| −2 | K060867 | Spy Intraoperative Imaging System | Novadaq Technologies, Inc. | 2006 | 1 |
| −1 | K063345 | Spy Imaging System | Novadaq Technologies, Inc. | 2007 | 1 |
| 0 (this device) | K150018 | VS3-IR-MMS System | Visionsense, Ltd. | 2015 | — |
| +1 | K152204 | VS3-IR system | Visionsense, Ltd. | 2015 | — |
| +2 | K162882 | KARL STORZ ICG Imaging System | KARL STORZ Endoscopy-America, Inc. | 2017 | — |
| +3 | K171238 | KARL STORZ ICG Imaging System | KARL STORZ Endoscopy-America, Inc. | 2017 | — |
| +3 | K180146 | KARL STORZ ICG Imaging System | Karl Storz Endoscopy America, Inc. | 2018 | — |
| +4 | K211202 | 1688 4K Camera System, L11 LED Light Source with AIM, AIM SafeLight Cable… | Stryker | 2021 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1999 | 1 | 0 |
| 2006 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2015 | 1 | 0 |
| 2017 | 2 | 0 |
| 2018 | 1 | 0 |
| 2020 | 2 | 0 |
| 2021 | 3 | 0 |
| 2022 | 5 | 0 |
| 2023 | 6 | 0 |
| 2024 | 4 | 0 |
| 2025 | 3 | 0 |
| 2026 | 8 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in OWN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OWN device
- #1 K231854 1788 4K Camera System with Advanced Imaging…score 24.9 · 24 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K063345 | Spy Imaging SystemSummary page 4…sual Name / Classification Name Surgical Microscope / Angiographic X-ray System / Endoscopes and Accessories Predicate Device Novadaq Technologies SPY Imaging System (K063345) Intended Use / Indications for Use The Endoscope Module of the VS3-IR-MMS System is intended for viewing internal surgical sit… | 2007 |
What descends from it? (35)
See all 352 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 17, 2022 | Class II | Z-0280-2023 | Under Investigation by firm | Broken Optical fiber cable damage can occur through handling or bending and can lead to loss of functionality and heating/melting of the outer cable surface. |
| Oct 17, 2022 | Class II | Z-0282-2023 | Under Investigation by firm | Broken Optical fiber cable damage can occur through handling or bending and can lead to loss of functionality and heating/melting of the outer cable surface. |
| Oct 17, 2022 | Class II | Z-0281-2023 | Under Investigation by firm | XXX |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 8.4 years on median.
- 4 of the 6 ancestors have a recall of any class.
- Of its descendants, K152204 (Visionsense, Ltd., 2015) has gone furthest, with 34 descendants of its own.
- 7 other devices also cite its predicate K063345.
How does its category compare?
All OWN devicesOWN Confocal Optical Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.
FDA decisions per year · OWN
3in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · OWN
196in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OWN
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · OWN
197in 2025
MAUDE reports naming product code OWN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 1 | — | 0 | 0 |
| 2012 | 3 | 87 | 0 | 0 |
| 2013 | 2 | — | 0 | 0 |
| 2014 | 5 | 208 | 0 | 0 |
| 2015 | 5 | 163 | 0 | 1 |
| 2016 | 2 | — | 0 | 4 |
| 2017 | 5 | 87 | 0 | 12 |
| 2018 | 1 | — | 0 | 11 |
| 2019 | 7 | 81 | 0 | 59 |
| 2020 | 7 | 81 | 0 | 105 |
| 2021 | 4 | 104 | 0 | 312 |
| 2022 | 4 | 41 | 3 | 346 |
| 2023 | 6 | 69 | 0 | 164 |
| 2024 | 1 | — | 1 | 165 |
| 2025 | 3 | 196 | 0 | 197 |
| 2026 | 7 | 63 | 0 | — |
What is it indicated for?
The Endoscope Module of the VS3-IR-MMS System is intended for viewing internal surgical sites during general surgical procedures, for use in visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and knee arthroscopic procedures. The Iridium Module of the VS3-IR-MMS…
Read the full statement (104 words)
The Endoscope Module of the VS3-IR-MMS System is intended for viewing internal surgical sites during general surgical procedures, for use in visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and knee arthroscopic procedures. The Iridium Module of the VS3-IR-MMS System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K150018 · printed Nov 4, 2026
Cite this page
510(k) Report. “K150018: VS3-IR-MMS System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K150018