510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Medtronic Model 5392 External Pulse Generator (EPG)

K150246510(k) SpecialCleared Feb 18, 201516-day reviewClass II21 CFR 870.3600Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic, Inc. · DTE Pulse-Generator, Pacemaker, External

Cleared on Feb 18, 2015 as substantially equivalent to K132924 (Medtronic, Inc., 2013) after a 16-day review. Its family tree reaches back 1 generation to 2013. 4 later devices descend from it. It has been recalled once.

Decided
Feb 18, 2015
Review
16 days
Class
Class II
Product code
DTE
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K132924 (2013). 4 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2016: 1 descendant cleared2018: 1 descendant cleared2019: 1 descendant cleared2020: 1 descendant clearedK132924 · Medtronic External Pulse Generator (Epg) · Medtronic, Inc. · 2013 · recalled2010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant K132924 · Medtronic External Pulse Generator (Epg) · Medtronic, Inc. · 2013 · recalled0
  1. Root · 1 generation backK132924 Medtronic External Pulse Generator (Epg)Medtronic, Inc. · 2013 · Recalled
  2. This deviceK150246 Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc. · 2015
  3. Descendant · cites it directly · 3 of its ownK162550 Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc. · 2016
  4. Descendant · 2 generations on · 2 of its ownK181973 Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc. · 2018
  5. Descendant · 3 generations on · 1 of its ownK190825 Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc. · 2019
  6. Descendant · 4 generations on · 0 of its ownK201011 Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc. · 2020

Most influential descendants

Showing 4 of 4 · See all
  • K162550Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc. · 2016 · cites it directly3descendants
  • K181973Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc. · 2018 · 2 generations on2descendants
  • K190825Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc. · 2019 · 3 generations on1descendant
  • K201011Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc. · 2020 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K132924Medtronic External Pulse Generator (Epg)Medtronic, Inc.20131
0 (this device)K150246Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.2015—
+1K162550Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.2016—
+2K181973Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.2018—
+3K190825Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.2019—
+4K201011Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.2020—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201810
201910
202010
FIG. 1 — Lineage of K150246, 2013–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTE device

  1. #1 K232721 Lifetech Cardio Temporary Pacemakerscore 7.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K132924Medtronic External Pulse Generator (Epg)
Summary page 4
…Classification: Class I1I Classification Panel: Cardiovascular Special Controls: Not applicable Predicate Devices: Medtronic Model 5392 External Pulse Generator (EPG) K132924 Update Battery Box Design for 5392 EPG Kxxxxxx Medtronic Confidential 1-22 Device Description The Model 5392 EPG…
Medtronic, Inc.20131

Also cleared against K132924: 1 other device →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162550Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.20163

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 17, 2015Class IIZ-1356-2015Device DesignPossible performance issue when used with specific AA-sized (LR6) batteries. The negative terminal of a commercially available AA (LR6) battery did not maintain a sufficient connection with the battery drawer electrical contact. This…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K162550 (Medtronic, Inc., 2016) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K132924.

How does its category compare?

All DTE devices

DTE Pulse-Generator, Pacemaker, External: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · DTE

1in 2025

2013: 22015: 12016: 22018: 22019: 22020: 12024: 12025: 1
2007high 2 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTE

Too few DTE 510(k)s in any year since 2007 to give a median.

Recalls per year · DTE

2in 2010

2010: 2
2007high 2 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DTE

636in 2025

2007: 172008: 222009: 392010: 3352011: 1,8652012: 1,7862013: 982014: 672015: 3422016: 4962017: 6542018: 9562019: 1,0012020: 1,0922021: 8992022: 8332023: 6502024: 6202025: 636
2007high 1,865 (2011) · low 17 (2007)2025

MAUDE reports naming product code DTE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—017
20080—022
20090—039
20100—2335
20110—01,865
20120—01,786
20132—098
20140—067
20151—0342
20162—0496
20170—0654
20182—0956
20192—01,001
20201—01,092
20210—0899
20220—0833
20230—0650
20241—0620
20251—0636
20260—0—

What is it indicated for?

The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating…

Read the full statement (152 words)

The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient.

Specific indications for temporary cardiac pacing include, but are not limited to, the following:

Complete heart block

Sinus bradycardia

Sick sinus syndrome

Bradycardia with congestive heart failure

Atrial and/or ventricular arrhythmias

Cardiac arrest

Support, management, and evaluation of a patient prior to permanent pacemaker Implantation

Support during permanent pacemaker replacement

Cardiac complications during invasive or surgical procedures

Support following cardiac surgery

Acute myocardial infarction complicated by heart block

Atrial tachyarrhythmias that require high-rate burst pacing for treatment

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K150246: Medtronic Model 5392 External Pulse Generator (EPG).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K150246

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026