510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Medtronic Temporary External Pacemaker 53401

K162054510(k) TraditionalCleared Oct 18, 201685-day reviewClass II21 CFR 870.3600Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic, Inc. · DTE Pulse-Generator, Pacemaker, External

Cleared on Oct 18, 2016 as substantially equivalent to K132924 (Medtronic, Inc., 2013) after an 85-day review. Its family tree reaches back 1 generation to 2013. 1 later device descends from it. It has been recalled once.

Decided
Oct 18, 2016
Review
85 days
Class
Class II
Product code
DTE
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K132924 (2013). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2018: 1 descendant clearedK132924 · Medtronic External Pulse Generator (Epg) · Medtronic, Inc. · 2013 · recalled2010201520202025
Descendants, by generation 1 generation forward: 1 descendant K132924 · Medtronic External Pulse Generator (Epg) · Medtronic, Inc. · 2013 · recalled0
  1. Root · 1 generation backK132924 Medtronic External Pulse Generator (Epg)Medtronic, Inc. · 2013 · Recalled
  2. This deviceK162054 Medtronic Temporary External Pacemaker 53401Medtronic, Inc. · 2016
  3. Descendant · cites it directly · 0 of its ownK180873 Medtronic Model 53401 External Pulse Generator (EPG)Medtronic, Inc. · 2018

Most influential descendants

Showing 1 of 1 · See all
  • K180873Medtronic Model 53401 External Pulse Generator (EPG)Medtronic, Inc. · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K132924Medtronic External Pulse Generator (Epg)Medtronic, Inc.20131
0 (this device)K162054Medtronic Temporary External Pacemaker 53401Medtronic, Inc.2016—
+1K180873Medtronic Model 53401 External Pulse Generator (EPG)Medtronic, Inc.2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
FIG. 1 — Lineage of K162054, 2013–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTE device

  1. #1 K232721 Lifetech Cardio Temporary Pacemakerscore 7.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K132924Medtronic External Pulse Generator (Epg)
Summary page 4
…the following predicate device:  Medtronic Model 5392 Dual Chamber Temporary External Pacemaker, initially cleared via 510(k) application, reference number K132924 on October 31, 2013. Brief Device Description The Medtronic Temporary External Pacemaker Model 53401 (hereafter simply referred…
Medtronic, Inc.20131

Also cleared against K132924: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180873Medtronic Model 53401 External Pulse Generator (EPG)Medtronic, Inc.20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 6, 2018Class IIZ-2317-2018Device DesignA subset of Medtronic Model 53401 Single Chamber External Pulse Generators (EPGs) may revert from programmed settings to nominal settings during use with a patient due to an unexpected Power on Reset (POR).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K180873 (Medtronic, Inc., 2018) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K132924.

How does its category compare?

All DTE devices

DTE Pulse-Generator, Pacemaker, External: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · DTE

1in 2025

2013: 22015: 12016: 22018: 22019: 22020: 12024: 12025: 1
2007high 2 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTE

Too few DTE 510(k)s in any year since 2007 to give a median.

Recalls per year · DTE

2in 2010

2010: 2
2007high 2 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DTE

636in 2025

2007: 172008: 222009: 392010: 3352011: 1,8652012: 1,7862013: 982014: 672015: 3422016: 4962017: 6542018: 9562019: 1,0012020: 1,0922021: 8992022: 8332023: 6502024: 6202025: 636
2007high 1,865 (2011) · low 17 (2007)2025

MAUDE reports naming product code DTE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—017
20080—022
20090—039
20100—2335
20110—01,865
20120—01,786
20132—098
20140—067
20151—0342
20162—0496
20170—0654
20182—0956
20192—01,001
20201—01,092
20210—0899
20220—0833
20230—0650
20241—0620
20251—0636
20260—0—

What is it indicated for?

The Model 53401 SC EPG is intended to be used in conjunction with a cardiac pacing lead system for temporary atrial or ventricular pacing in a clinical environment. The device can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylatic, or diagnostic purposes. Specification indications for temporary cardiac pacing include, but are not limited to, the following indications: • Complete heart block •…

Read the full statement (143 words)

The Model 53401 SC EPG is intended to be used in conjunction with a cardiac pacing lead system for temporary atrial or ventricular pacing in a clinical environment. The device can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylatic, or diagnostic purposes.

Specification indications for temporary cardiac pacing include, but are not limited to, the following indications:

• Complete heart block

• Sinus bradycardia

• Sick Sinus Syndrome

• Bradycardia with congestive heart failure

• Atrial and/or ventricular arrhythmias

• Cardiac arrest

• Temporary support, management, and evaluation of a patient prior to permanent pacemaker implantation

• Support during permanent pacemaker replacement

• Cardiac complications during invasive or surgical procedures

• Temporary support of a patient following cardiac surgery

• Acute myocardial infarction complicated by heart block

• High-rate burst pacing for the treatment of supraventricular tachyarrhythmias.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K162054: Medtronic Temporary External Pacemaker 53401.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162054

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026