510(k) Report

510(k) · Cleared

Cleared

StatStrip Glucose Hospital Meter

K150281510(k) TraditionalCleared May 6, 201590-day reviewClass II21 CFR 862.1345

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nova Biomedical Corporation · PZI Prescription Use Blood Glucose Meter For Near-Patient Testing

Cleared on May 6, 2015 as substantially equivalent to K060345 (Nova Biomedical Corp., 2006) and 2 other predicates after a 90-day review. Its family tree reaches back 2 generations to 2006. 1 of the 9 devices in its family, counting itself, has a Class I recall. 5 later devices descend from it.

Decided
May 6, 2015
Review
90 days
Class
Class II
Product code
PZI
Recalls
None
Predicates
3

Its longest line of descent runs back 2 generations to K060345 (2006). 3 ancestors in all. 5 later devices descend from it, 1 citing it directly. 1 of the 9 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2018: 3 descendants cleared2024: 1 descendant cleared, 1 with a Class I recall12026: 1 descendant clearedK060345 · Nova Statstrip Glucose Hospital Meter, Statstrip Control Solutions · Nova Biomedical Corp. · 2006 · recalledK132121 · Statstrip Glucose Hospital Meter System · Nova Biomedical Corporation · 201420052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 4 descendants, 1 with a Class I recall1 K132121 · Statstrip Glucose Hospital Meter System · Nova Biomedical Corporation · 2014K060345 · Nova Statstrip Glucose Hospital Meter, Statstrip Control Solutions · Nova Biomedical Corp. · 2006 · recalled0
  1. Root · 2 generations backK060345 Nova Statstrip Glucose Hospital Meter, Statstrip Control SolutionsNova Biomedical Corp. · 2006 · Recalled
  2. 1 generation backK132121 Statstrip Glucose Hospital Meter SystemNova Biomedical Corporation · 2014
  3. This deviceK150281 StatStrip Glucose Hospital MeterNova Biomedical Corporation · 2015
  4. Descendant · cites it directly · 4 of its ownK181043 StatStrip Glucose Hospital Meter SystemNova Biomedical Corporation · 2018
  5. Descendant · 2 generations on · 0 of its ownK252323 cobas pulse blood glucose monitoring systemRoche Diagnostics · 2026
  6. Descendant · 2 generations on · 0 of its ownK232075 StatStrip Glucose Hospital Meter SystemNova Biomedical Corporation · 2024 · Recalled (Class I)
  7. Descendant · 2 generations on · 0 of its ownK182552 StatStrip Xpress Glucose Hospital Meter SystemNova Biomedical Corporation · 2018
  8. Descendant · 2 generations on · 0 of its ownK182549 StatStrip Xpress 2 Glucose Hospital Meter SystemNova Biomedical Corporation · 2018

Class I recalls in this family

Most influential descendants

Showing 5 of 5 · See all
  • K181043StatStrip Glucose Hospital Meter SystemNova Biomedical Corporation · 2018 · cites it directly4descendants
  • K252323cobas pulse blood glucose monitoring systemRoche Diagnostics · 2026 · 2 generations on0descendants
  • K232075StatStrip Glucose Hospital Meter SystemNova Biomedical Corporation · 2024 · 2 generations on0descendants
  • K182552StatStrip Xpress Glucose Hospital Meter SystemNova Biomedical Corporation · 2018 · 2 generations on0descendants
  • K182549StatStrip Xpress 2 Glucose Hospital Meter SystemNova Biomedical Corporation · 2018 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K060345Nova Statstrip Glucose Hospital Meter, Statstrip Control SolutionsNova Biomedical Corp.20063
−1K132121Statstrip Glucose Hospital Meter SystemNova Biomedical Corporation2014—
0 (this device)K150281StatStrip Glucose Hospital MeterNova Biomedical Corporation2015—
+1K181043StatStrip Glucose Hospital Meter SystemNova Biomedical Corporation2018—
+2K252323cobas pulse blood glucose monitoring systemRoche Diagnostics2026—
+2K232075StatStrip Glucose Hospital Meter SystemNova Biomedical Corporation2024Class I
+2K182552StatStrip Xpress Glucose Hospital Meter SystemNova Biomedical Corporation2018—
+2K182549StatStrip Xpress 2 Glucose Hospital Meter SystemNova Biomedical Corporation2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201830
202411
202610
FIG. 1 — Lineage of K150281, 2006–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PZI device

  1. #1 K252323 cobas pulse blood glucose monitoring systemscore 12.4 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060345Nova Statstrip Glucose Hospital Meter, Statstrip Control Solutions
Summary page 4
…Glucose Test System 75CGA 862.1345 II Product Codes: CGA Predicate Device: StatStrip Glucose Hospital Meter System, K060345, K063821, and K132121. Device Description: The StatStrip Glucose Hospital Meter System previously cleared under K060345, K06382…
Nova Biomedical Corp.20063
*K063821Nova Statstrip Glucose Hospital Meter System
Summary page 4
…Test System 75CGA 862.1345 II Product Codes: CGA Predicate Device: StatStrip Glucose Hospital Meter System, K060345, K063821, and K132121. Device Description: The StatStrip Glucose Hospital Meter System previously cleared under K060345, K063821, and K1…
Nova Biomedical Corp.20074
*K132121Statstrip Glucose Hospital Meter System
Summary page 4
…75CGA 862.1345 II Product Codes: CGA Predicate Device: StatStrip Glucose Hospital Meter System, K060345, K063821, and K132121. Device Description: The StatStrip Glucose Hospital Meter System previously cleared under K060345, K063821, and K132121 is bein…
Nova Biomedical Corporation2014—

Also cleared against K060345: 2 other devices →

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181043StatStrip Glucose Hospital Meter SystemNova Biomedical Corporation20184

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K232075StatStrip Glucose Hospital Meter SystemNova Biomedical Corporation · 2024 · 2 generations on · ◆ in the 2024 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 9 years and passes through 2 companies; a new generation was cleared every 8.4 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K181043 (Nova Biomedical Corporation, 2018) has gone furthest, with 4 descendants of its own.
  • 2 other devices also cite its predicate K060345.

How does its category compare?

All PZI devices

PZI Prescription Use Blood Glucose Meter For Near-Patient Testing: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · PZI

3in 2024

2014: 12015: 22016: 22017: 12018: 32022: 12024: 32026 (partial year): 1
2007high 3 (2018) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PZI

816in 2024

2018: 87 days median2024: 816 days median
2007high 816 (2024) · low 87 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PZI

5so far in 2026, the most since 2007

2024: 1, 1 Class I2026 (partial year): 5, 5 Class I
2007high 5 (2026) · low 0 (2007)2026

Class I (6 since 2007) Class II/III not classified

Adverse event reports per year · PZI

35in 2025

2019: 192020: 82021: 182022: 72023: 22024: 52025: 35
2007high 35 (2025) · low 0 (2007)2025

MAUDE reports naming product code PZI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20141—00
20152—00
20162—00
20171—00
201838700
20190—019
20200—08
20210—018
20221—07
20230—02
202438161 (1)5
20250—035
20261—5 (5)—

What is it indicated for?

The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood and neonate heel stick specimens. The StatStrip Glucose Hospital Meter System is also intended for use in the quantitative determination of glucose in venous whole blood, arterial whole blood…

Read the full statement (131 words)

The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood and neonate heel stick specimens.

The StatStrip Glucose Hospital Meter System is also intended for use in the quantitative determination of glucose in venous whole blood, arterial whole blood, neonatal heel stick and neonatal arterial samples throughout all hospital and all professional healthcare settings.

The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.

It is not intended for use with neonate cord blood specimens.

It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K150281: StatStrip Glucose Hospital Meter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K150281

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026