510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

IntelliVue Patient Monitor

K150310510(k) SpecialCleared May 13, 201593-day reviewClass II21 CFR 870.1025Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medizin Systeme Boeblingen GmbH · DSI Detector And Alarm, Arrhythmia

Cleared on May 13, 2015 after a 93-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 32 later devices descend from it. It has been recalled 11 times.

Decided
May 13, 2015
Review
93 days
Class
Class II
Product code
DSI
Recalls
11
Predicates
0

32 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2016: 1 descendant cleared2018: 4 descendants cleared2017: 2 descendants cleared2019: 4 descendants cleared2020: 1 descendant cleared2021: 1 descendant cleared2022: 3 descendants cleared2023: 5 descendants cleared2026: 4 descendants cleared2025: 6 descendants cleared2024: 1 descendant cleared201620182020202220242026
  1. This deviceK150310 IntelliVue Patient MonitorPhilips Medizin Systeme Boeblingen GmbH · 2015
  2. Descendant · cites it directly · 31 of its ownK161531 IntelliVue Patient Monitors MP2, MP5, MP5T, MP5SC, and…Philips Medizin Systeme Boeblingen GmbH · 2016 · Recalled
  3. Descendant · 2 generations on · 8 of its ownK171801 IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3…Philips Medizin Systeme Boeblingen GmbH · 2017 · Recalled
  4. Descendant · 2 generations on · 6 of its ownK221704 HemoSphere Advanced Monitor, Right Ventricular Pressure (RVP)…Edwards Lifesciences, LLC · 2022
  5. Descendant · 2 generations on · 6 of its ownK210906 IntelliVue Patient Monitor MX750, IntelliVue Patient Monitor MX850…Philips Medizin Systeme Boeblingen GmbH · 2021 · Recalled
  6. Descendant · 2 generations on · 6 of its ownK181314 IntelliVue Patient Monitor MX100, IntelliVue Multi-Measurement…Philips Medizin Systeme Boeblingen GmbH · 2018

Showing 5 of 32 descendants. See all descendants

Descendants cleared per year

2016: 12017: 22018: 42019: 42020: 12021: 12022: 32023: 52024: 12025: 62026: 4
2016high 6 (2025) · low 1 (2016)2026

Most influential descendants

Showing 5 of 32 · See all
  • K161531IntelliVue Patient Monitors MP2, MP5, MP5T, MP5SC, and Multi-Measurement Module X2…Philips Medizin Systeme Boeblingen GmbH · 2016 · cites it directly31descendants
  • K171801IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3; intelliVue Hemodynamic…Philips Medizin Systeme Boeblingen GmbH · 2017 · 2 generations on8descendants
  • K221704HemoSphere Advanced Monitor, Right Ventricular Pressure (RVP), Global Hypoperfusion…Edwards Lifesciences, LLC · 2022 · 2 generations on6descendants
  • K210906IntelliVue Patient Monitor MX750, IntelliVue Patient Monitor MX850, IntelliVue 4-Slot…Philips Medizin Systeme Boeblingen GmbH · 2021 · 2 generations on6descendants
  • K181314IntelliVue Patient Monitor MX100, IntelliVue Multi-Measurement Module X3, IntelliVue…Philips Medizin Systeme Boeblingen GmbH · 2018 · 2 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K150310IntelliVue Patient MonitorPhilips Medizin Systeme Boeblingen GmbH2015—
+1K161531IntelliVue Patient Monitors MP2, MP5, MP5T, MP5SC, and Multi-Measurement…Philips Medizin Systeme Boeblingen GmbH20161
+2K171801IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3; intelliVue…Philips Medizin Systeme Boeblingen GmbH20174
+2K221704HemoSphere Advanced Monitor, Right Ventricular Pressure (RVP), Global…Edwards Lifesciences, LLC2022—
+2K210906IntelliVue Patient Monitor MX750, IntelliVue Patient Monitor MX850, IntelliVue…Philips Medizin Systeme Boeblingen GmbH20212
+2K181314IntelliVue Patient Monitor MX100, IntelliVue Multi-Measurement Module X3…Philips Medizin Systeme Boeblingen GmbH2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201720
201840
201940
202010
202110
202230
202350
202410
202560
202640
FIG. 1 — Lineage of K150310, 2015–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DSI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSI device

  1. #1 K243349 BodyGuardian Remote Monitoring System (BGRMS…score 13.1 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (32)

See all 32

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K161531IntelliVue Patient Monitors MP2, MP5, MP5T, MP5SC, and Multi-Measurement Module X2…Philips Medizin Systeme Boeblingen GmbH201631

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 31, 2025Class IIZ-0849-2026Use errorPotential issue where the IntelliVue monitors did not alarm.
Oct 31, 2025Class IIZ-0850-2026Use errorPotential issue where the IntelliVue monitors did not alarm.
Oct 31, 2025Class IIZ-0867-2026Use errorPotential issue where the IntelliVue monitors did not alarm.
Oct 31, 2025Class IIZ-0869-2026Use errorPotential issue where the IntelliVue monitors did not alarm.
Oct 31, 2025Class IIZ-0866-2026Use errorPotential issue where the IntelliVue monitors did not alarm.
Oct 31, 2025Class IIZ-0859-2026Use errorPotential issue where the IntelliVue monitors did not alarm.
Jun 30, 2020Class IIZ-2696-2020Process controlSome Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the…
Jun 30, 2020Class IIZ-2697-2020Process controlSome Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the…
Aug 23, 2017Class IIZ-0292-2018Software designDistorted ST-Segment when using 12-Lead ECG Monitoring with Philips IntelliVue Patient Monitors
Jan 4, 2016Class IIZ-0854-2016Software Design ChangeThe ST elevation alarm on the Patient Monitor or standalone X2 Measurement Module will not sound when indicated for all chest leads derived using Hexad 12-Lead ECG Monitoring in the Host Monitor
Jan 4, 2016Class IIZ-0853-2016Software Design ChangeThe ST elevation alarm on the Patient Monitor or standalone X2 Measurement Module will not sound when indicated for all chest leads derived using Hexad 12-Lead ECG Monitoring in the Host Monitor

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K161531 (Philips Medizin Systeme Boeblingen GmbH, 2016) has gone furthest, with 31 descendants of its own.

How does its category compare?

All DSI devices

DSI Detector And Alarm, Arrhythmia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · DSI

1in 2025

2007: 72008: 52009: 62010: 62011: 62012: 32013: 12014: 42015: 92016: 62018: 22019: 32020: 22023: 22024: 12025: 12026 (partial year): 1
2007high 9 (2015) · low 0 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSI

243in 2019

2007: 151 days median2008: 70 days median2009: 73 days median2010: 111 days median2011: 114 days median2012: 113 days median2014: 200 days median2015: 127 days median2016: 224 days median2019: 243 days median
2007high 243 (2019) · low 70 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSI

18in 2025

2007: 42008: 82009: 32010: 52011: 32012: 12018: 12019: 42020: 22022: 12023: 22025: 182026 (partial year): 1
2007high 18 (2025) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DSI

712in 2025

2007: 642008: 1562009: 2792010: 8232011: 7932012: 7662013: 7952014: 1,0002015: 1,6452016: 4,3522017: 3,5942018: 3,6482019: 3,4852020: 3,0842021: 3,0922022: 1,6382023: 1,6162024: 1,3202025: 712
2007high 4,352 (2016) · low 64 (2007)2025

MAUDE reports naming product code DSI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077151464
20085708156
20096733279
201061115823
201161143793
201231131766
20131—0795
2014420001,000
2015912701,645
2016622404,352
20170—03,594
20182—13,648
2019324343,485
20202—23,084
20210—03,092
20220—11,638
20232—21,616
20241—01,320
20251—18712
20261—1—
Cite this page

510(k) Report. “K150310: IntelliVue Patient Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K150310

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026