510(k) Report

510(k) · Cleared

Cleared

SmartCardia 7L Platform

K231276510(k) TraditionalCleared Aug 30, 2023120-day reviewClass II21 CFR 870.1025

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smartcardia SA · DSI Detector And Alarm, Arrhythmia

Cleared on Aug 30, 2023 as substantially equivalent to K192757 (Vitalconnect, Inc., 2020) and 1 other predicate after a 120-day review. Its family tree reaches back 13 generations to 1997.

Decided
Aug 30, 2023
Review
120 days
Class
Class II
Product code
DSI
Recalls
None
Predicates
2

Its longest line of descent runs back 13 generations to K962603 (1997). 69 ancestors in all, the oldest K843503 (1984). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2016: 1 other ancestor2010: 1 other ancestor2009: 4 other ancestors2008: 3 other ancestors2007: 2 other ancestors1999: 4 other ancestors2006: 2 other ancestors2005: 4 other ancestors2004: 4 other ancestors2002: 1 other ancestor1998: 3 other ancestors2001: 3 other ancestors2000: 4 other ancestors1996: 4 other ancestors1995: 2 other ancestors1992: 4 other ancestors1988: 2 other ancestors1984: 1 other ancestor1997: 2 other ancestors2003: 1 other ancestor1994: 1 other ancestor1993: 1 other ancestor1991: 1 other ancestor1989: 1 other ancestorK962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999K012451 · Propaq Encore Models 202, 204, 206; Propaq CS Models 242, 244, 246 · Welch Allyn Protocol, Inc. · 2001K061993 · Equivital, Model EQ-01 · Hidalgo Limited · 2006K113054 · Equivital · Hidalgo Limited · 2012K141167 · Vitalconnect Platform (Consisting of Vitalconnect Sensor, Relay Software Library and Secure Server Software Library) · Vitalconnect, Inc. · 2014K152139 · Vital Connect Platform, HealthPatch MD, VitalPatch · Vitalconnect, Inc. · 2015K163453 · VitalPatch® VitalConnect Platform · Vitalconnect, Inc. · 2017×2: 2 generations cleared close together, K183078 (2019), K190916 (2019) K192757 · VitalPatch 5D Biosensor - single, VitalPatch 7D Biosensor - single, VitalPatch 5D Biosensor – bag of 20, VitalPatch 7D Biosensor – bag of 20 · Vitalconnect, Inc. · 202019801990200020102020
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 3 other ancestors4 generations back: 6 other ancestors5 generations back: 9 other ancestors6 generations back: 16 other ancestors7 generations back: 2 other ancestors8 generations back: 6 other ancestors9 generations back: 8 other ancestors10 generations back: 4 other ancestors K192757 · VitalPatch 5D Biosensor - single, VitalPatch 7D Biosensor - single, VitalPatch 5D Biosensor – bag of 20, VitalPatch 7D Biosensor – bag of 20 · Vitalconnect, Inc. · 2020K190916 · VitalPatch 7D Biosensor - single, VitalPatch 7D Biosensor - bag of 20 · Vitalconnect, Inc. · 2019K183078 · VitalConnect Platform, VitalPatch Biosensor · Vitalconnect, Inc. · 2019K163453 · VitalPatch® VitalConnect Platform · Vitalconnect, Inc. · 2017K152139 · Vital Connect Platform, HealthPatch MD, VitalPatch · Vitalconnect, Inc. · 2015K141167 · Vitalconnect Platform (Consisting of Vitalconnect Sensor, Relay Software Library and Secure Server Software Library) · Vitalconnect, Inc. · 2014K113054 · Equivital · Hidalgo Limited · 2012K061993 · Equivital, Model EQ-01 · Hidalgo Limited · 2006K012451 · Propaq Encore Models 202, 204, 206; Propaq CS Models 242, 244, 246 · Welch Allyn Protocol, Inc. · 2001K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997−10−50
  1. Root · 13 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 9 generations between
  3. 3 generations backK183078 VitalConnect Platform, VitalPatch BiosensorVitalconnect, Inc. · 2019
  4. 2 generations backK190916 VitalPatch 7D Biosensor - single, VitalPatch 7D Biosensor - bag of 20Vitalconnect, Inc. · 2019
  5. 1 generation backK192757 VitalPatch 5D Biosensor - single, VitalPatch 7D Biosensor - single…Vitalconnect, Inc. · 2020
  6. This deviceK231276 SmartCardia 7L PlatformSmartcardia SA · 2023
Show all 13 generations
  1. Root · 13 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 12 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 11 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 10 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 9 generations backK012451 Propaq Encore Models 202, 204, 206; Propaq CS Models 242, 244, 246Welch Allyn Protocol, Inc. · 2001
  6. 8 generations backK061993 Equivital, Model EQ-01Hidalgo Limited · 2006
  7. 7 generations backK113054 EquivitalHidalgo Limited · 2012
  8. 6 generations backK141167 Vitalconnect Platform (Consisting of Vitalconnect Sensor, Relay…Vitalconnect, Inc. · 2014
  9. 5 generations backK152139 Vital Connect Platform, HealthPatch MD, VitalPatchVitalconnect, Inc. · 2015
  10. 4 generations backK163453 VitalPatch® VitalConnect PlatformVitalconnect, Inc. · 2017
  11. 3 generations backK183078 VitalConnect Platform, VitalPatch BiosensorVitalconnect, Inc. · 2019
  12. 2 generations backK190916 VitalPatch 7D Biosensor - single, VitalPatch 7D Biosensor - bag of 20Vitalconnect, Inc. · 2019
  13. 1 generation backK192757 VitalPatch 5D Biosensor - single, VitalPatch 7D Biosensor - single…Vitalconnect, Inc. · 2020
  14. This deviceK231276 SmartCardia 7L PlatformSmartcardia SA · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−13K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−12K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−11K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−10K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−9K012451Propaq Encore Models 202, 204, 206; Propaq CS Models 242, 244, 246Welch Allyn Protocol, Inc.2001—
−8K061993Equivital, Model EQ-01Hidalgo Limited2006—
−7K113054EquivitalHidalgo Limited2012—
−6K141167Vitalconnect Platform (Consisting of Vitalconnect Sensor, Relay Software…Vitalconnect, Inc.2014—
−5K152139Vital Connect Platform, HealthPatch MD, VitalPatchVitalconnect, Inc.2015—
−4K163453VitalPatch® VitalConnect PlatformVitalconnect, Inc.2017—
−3K183078VitalConnect Platform, VitalPatch BiosensorVitalconnect, Inc.2019—
−2K190916VitalPatch 7D Biosensor - single, VitalPatch 7D Biosensor - bag of 20Vitalconnect, Inc.2019—
−1K192757VitalPatch 5D Biosensor - single, VitalPatch 7D Biosensor - single, VitalPatch…Vitalconnect, Inc.2020—
0 (this device)K231276SmartCardia 7L PlatformSmartcardia SA2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
198820
198910
199110
199240
199310
199410
199520
199640
199720
199830
199940
200040
200130
200210
200310
200440
200540
200620
200720
200830
200940
201010
201610
FIG. 1 — Lineage of K231276, 1997–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DSI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSI device

  1. #1 K243349 BodyGuardian Remote Monitoring System (BGRMS…score 13.1 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K192757VitalPatch 5D Biosensor - single, VitalPatch 7D Biosensor - single, VitalPatch 5D…
Summary page 4
…870.2910 Regulation: Primary Product Code: DRG Secondary Codes: DSI Primary Predicate Device: VitalConnect, Inc.; VitalConnect Biosensor; K192757 K231276 Page 1 of 6 Seco…
Vitalconnect, Inc.2020—
*K160064MoMe Kardia Wireless Ambulatory ECG Monitoring and Detection System
Summary page 5
…Page 1 of 6 Secondary Predicate / InfoBionics; MoMe Kardia Wireless Ambulatory ECG Monitoring and Reference Device: Detection System; K160064 Device Description: The SmartCardia 7L Platform is a body worn Holter monitoring product…
Infobionic, Inc.2016—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 7 companies; a new generation was cleared every 1.4 years on median.
  • 6 of the 69 ancestors have a recall of any class.

How does its category compare?

All DSI devices

DSI Detector And Alarm, Arrhythmia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · DSI

1in 2025

2007: 72008: 52009: 62010: 62011: 62012: 32013: 12014: 42015: 92016: 62018: 22019: 32020: 22023: 22024: 12025: 12026 (partial year): 1
2007high 9 (2015) · low 0 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSI

243in 2019

2007: 151 days median2008: 70 days median2009: 73 days median2010: 111 days median2011: 114 days median2012: 113 days median2014: 200 days median2015: 127 days median2016: 224 days median2019: 243 days median
2007high 243 (2019) · low 70 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSI

18in 2025

2007: 42008: 82009: 32010: 52011: 32012: 12018: 12019: 42020: 22022: 12023: 22025: 182026 (partial year): 1
2007high 18 (2025) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DSI

712in 2025

2007: 642008: 1562009: 2792010: 8232011: 7932012: 7662013: 7952014: 1,0002015: 1,6452016: 4,3522017: 3,5942018: 3,6482019: 3,4852020: 3,0842021: 3,0922022: 1,6382023: 1,6162024: 1,3202025: 712
2007high 4,352 (2016) · low 64 (2007)2025

MAUDE reports naming product code DSI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077151464
20085708156
20096733279
201061115823
201161143793
201231131766
20131—0795
2014420001,000
2015912701,645
2016622404,352
20170—03,594
20182—13,648
2019324343,485
20202—23,084
20210—03,092
20220—11,638
20232—21,616
20241—01,320
20251—18712
20261—1—

What is it indicated for?

The SmartCardia 7L Platform is a prescribed device used for the purpose of identifying non-lethal arrhythmias. The SmartCardia 7L Platform is intended for: 1. Patients who experience transient symptoms that may suggest cardiac arrhythmia. 2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g., atrial fibrillation). 3. Patients with symptoms that may be due to cardiac arrhythmias. These may…

Read the full statement (241 words)

The SmartCardia 7L Platform is a prescribed device used for the purpose of identifying non-lethal arrhythmias. The SmartCardia 7L Platform is intended for:

1. Patients who experience transient symptoms that may suggest cardiac arrhythmia.

2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g., atrial fibrillation).

3. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath).

4. Patients recovering from cardiac surgery or interventional procedures who are indicated for outpatient arrhythmia monitoring.

5. ECG data recorded by the device can be analyzed by other processing systems to provide Holter style reports. Measurements include: electrocardiogram (ECG signal), R-R interval, Heart Rate. Notification alerts can be set for one or more of these measures.

The SmartCardia 7L Platform is indicated for use on patients who are 18 years of age or older to provide monitoring of physiological information. It is intended for use in a physician office, outpatient facility, or in the patient's home.

SmartCardia 7L Platform contraindications:

1. The SmartCardia 7L Platform is contraindicated for use in a critical care setting where an immediate response to life threatening conditions such as lethal arrhythmias is required.

2. The SmartCardia 7L Platform is contraindicated for use during external defibrillation.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K231276: SmartCardia 7L Platform.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K231276

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026