510(k) Report

510(k) · Cleared

Cleared

Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip Card

K150356510(k) SpecialCleared May 13, 201590-day reviewClass II21 CFR 862.3170

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Healgen Scientific,, LLC · JXM Enzyme Immunoassay, Benzodiazepine

Cleared on May 13, 2015 as substantially equivalent to K142280 (Healgen Scientific,, LLC, 2014) and 3 other predicates after a 90-day review. Its family tree reaches back 3 generations to 2005.

Decided
May 13, 2015
Review
90 days
Class
Class II
Product code
JXM
Recalls
None
Predicates
4

Its longest line of descent runs back 3 generations to K050186 (2005). 6 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK050186 Ameditech Immutest Multi-Drug Screen Panel IiiAmeditech, Inc. · 2005 · Recalled
  2. 2 generations backK052115 First Check Multi Drug Cup 12 for Marijuana (THC), Cocaine (Coc)…First Check Diagnostics, LLC · 2005 · Recalled
  3. 1 generation backK140546 Healgen THC One Step Marijuana Test , Healgen Mamp One Step…Healgen Scientific,, LLC · 2014 · Recalled
  4. This deviceK150356 Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip…Healgen Scientific,, LLC · 2015
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K050186Ameditech Immutest Multi-Drug Screen Panel IiiAmeditech, Inc.20053
−2K052115First Check Multi Drug Cup 12 for Marijuana (THC), Cocaine (Coc), Amphetamine…First Check Diagnostics, LLC200586
−1K140546Healgen THC One Step Marijuana Test , Healgen Mamp One Step Methamphetamine TestHealgen Scientific,, LLC20142
0 (this device)K150356Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip CardHealgen Scientific,, LLC2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
201420
201510
FIG. 1 — Lineage of K150356, 2005–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JXM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXM device

  1. #1 K243498 Alinity c Benzodiazepines Reagent Kitscore 8.8 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K142280Healgen Oxazepam Test (Strip, Cassette, Cup, Dip Card), Healgen Morphine Test (Strip…
Summary page 5
…colored line in the designated test region for the drug. To serve as a test control, a color line will always appear at the control region. 6. Predicate Devices: K142280; Healgen Oxazepam and Morphine Tests (Strip, Cassette, Dip Card, Cup) K143187; Healgen Amphetamine and Oxycodone Tests (Strip,…
Healgen Scientific,, LLC2014—
*K143187Healgen Amphetamine Test (Strip, Cassette, Cup, Dip Card), Healgen Oxycodone Test…
Summary page 5
…ontrol, a color line will always appear at the control region. 6. Predicate Devices: K142280; Healgen Oxazepam and Morphine Tests (Strip, Cassette, Dip Card, Cup) K143187; Healgen Amphetamine and Oxycodone Tests (Strip, Cassette, Dip Card, Cup) K141647; Healgen Cocaine and Morphine Tests (Strip,…
Healgen Scientific,, LLC2015—
*K141647Healgen Coc One Step Cocaine Test Strip:Morhpine Test Strip (Cassette, Cup, Dip Card)
Summary page 5
…K142280; Healgen Oxazepam and Morphine Tests (Strip, Cassette, Dip Card, Cup) K143187; Healgen Amphetamine and Oxycodone Tests (Strip, Cassette, Dip Card, Cup) K141647; Healgen Cocaine and Morphine Tests (Strip, Cassette, Dip Card, Cup) K140546; Healgen Marijuana and Methamphetamine Tests (Str…
Healgen Scientific,, LLC2014—
*K140546Healgen THC One Step Marijuana Test , Healgen Mamp One Step Methamphetamine Test
Summary page 5
…) K143187; Healgen Amphetamine and Oxycodone Tests (Strip, Cassette, Dip Card, Cup) K141647; Healgen Cocaine and Morphine Tests (Strip, Cassette, Dip Card, Cup) K140546; Healgen Marijuana and Methamphetamine Tests (Strip, Cassette, Dip Card, Cup) 7. Intended Use / Indications for Use Healgen M…
Healgen Scientific,, LLC20142

Also cleared against K142280: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 11 months on median.
  • 3 of the 6 ancestors have a recall of any class.
  • 3 other devices also cite its predicate K142280.

How does its category compare?

All JXM devices

JXM Enzyme Immunoassay, Benzodiazepine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · JXM

1in 2024

2010: 12011: 12012: 12015: 22016: 22017: 22018: 12019: 12022: 12024: 1
2007high 2 (2015) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXM

Too few JXM 510(k)s in any year since 2007 to give a median.

Recalls per year · JXM

79in 2015

2015: 79
2007high 79 (2015) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXM

6in 2025

2008: 32009: 12010: 32011: 32012: 12013: 32014: 32015: 52016: 22017: 42018: 42019: 32021: 12023: 52024: 62025: 6
2007high 6 (2024) · low 0 (2007)2025

MAUDE reports naming product code JXM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—03
20090—01
20101—03
20111—03
20121—01
20130—03
20140—03
20152—795
20162—02
20172—04
20181—04
20191—03
20200—00
20210—01
20221—00
20230—05
20241—06
20250—06
20260—0—

What is it indicated for?

Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test Dip Card are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine and Oxycodone in human urine at the cutoff concentrations of: Drug(Identifier) Cut-off level Amphetamine(AMP) 1000 ng/mL Oxazepam (OXA) 300 ng/mL Cocaine (COC) 300 ng/mL Cannabinoids (THC) 50…

Read the full statement (185 words)

Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test Dip Card are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine and Oxycodone in human urine at the cutoff concentrations of:

Drug(Identifier) Cut-off level

Amphetamine(AMP) 1000 ng/mL

Oxazepam (OXA) 300 ng/mL

Cocaine (COC) 300 ng/mL

Cannabinoids (THC) 50 ng/mL

Methamphetamine (MET) 1000 ng/mL

Morphine (MOR) 2000 ng/mL

Oxycodone (OXY) 100 ng/mL

Configuration of the Healgen Multi-Drug Urine Test Cup and Healgen Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs oxazepam and oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K150356: Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip Card.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K150356

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026