510(k) Report

510(k) · Cleared

Cleared

DRI Hydrocodone Assay, DRI Hydrocodone Calibrators, DRI Hydrocodone Controls

K150502510(k) TraditionalCleared Oct 9, 2015225-day reviewClass II21 CFR 862.3650

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Microgenics Corporation · DJG Enzyme Immunoassay, Opiates

Cleared on Oct 9, 2015 as substantially equivalent to K141055 (Lin-Zhi International, Inc., 2014) after a 225-day review. Its family tree reaches back 5 generations to 2002. 2 later devices descend from it.

Decided
Oct 9, 2015
Review
225 days
Class
Class II
Product code
DJG
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K014101 (2002). 2 later devices descend from it, 2 citing it directly.

  1. Root · 5 generations backK014101 Rapidone-Oxy TestAmerican Bio Medica Corp. · 2002
  2. 4 generations backK040411 Dri Oxycodone AssayMicrogenics Corp. · 2004
  3. 3 generations backK050733 Oxycodone Enzyme Immunoassay and Oxycodone Calibrators and ControlsLin-Zhi International, Inc. · 2005
  4. 2 generations backK120763 Lzi Oxycodone Homogeneous Enzyme Immunoassay, Lzi Oxycondone…Lin-Zhi International, Inc. · 2012
  5. 1 generation backK141055 Lzi Hydrocodone; Enzyme Immunoassay, Calibrators; (5ML), (15ML)…Lin-Zhi International, Inc. · 2014
  6. This deviceK150502 DRI Hydrocodone Assay, DRI Hydrocodone Calibrators, DRI Hydrocodone…Microgenics Corporation · 2015
  7. Descendant · cites it directly · 0 of its ownK231752 ARK Hydrocodone AssayArk Diagnostics, Inc. · 2023
  8. Descendant · cites it directly · 0 of its ownK173195 DRI Hydrocodone AssayMicrogenics Corporation · 2018

Most influential descendants

Showing 2 of 2 · See all
  • K231752ARK Hydrocodone AssayArk Diagnostics, Inc. · 2023 · cites it directly0descendants
  • K173195DRI Hydrocodone AssayMicrogenics Corporation · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K014101Rapidone-Oxy TestAmerican Bio Medica Corp.2002—
−4K040411Dri Oxycodone AssayMicrogenics Corp.2004—
−3K050733Oxycodone Enzyme Immunoassay and Oxycodone Calibrators and ControlsLin-Zhi International, Inc.2005—
−2K120763Lzi Oxycodone Homogeneous Enzyme Immunoassay, Lzi Oxycondone Calibrators, Lzi…Lin-Zhi International, Inc.2012—
−1K141055Lzi Hydrocodone; Enzyme Immunoassay, Calibrators; (5ML), (15ML); Controls…Lin-Zhi International, Inc.2014—
0 (this device)K150502DRI Hydrocodone Assay, DRI Hydrocodone Calibrators, DRI Hydrocodone ControlsMicrogenics Corporation2015—
+1K231752ARK Hydrocodone AssayArk Diagnostics, Inc.2023—
+1K173195DRI Hydrocodone AssayMicrogenics Corporation2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202310
FIG. 1 — Lineage of K150502, 2002–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DJG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DJG device

  1. #1 K232659 Dochek® Multi-Drug Urine Test Cup Rx, Dochek®…score 17.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141055Lzi Hydrocodone; Enzyme Immunoassay, Calibrators; (5ML), (15ML); Controls; (5ML), (15ML)
Summary page 5
…lly Marketed Predicate Device(s) The DRI® Hydrocodone Assay is substantially equivalent to the predicate Lin-Zhi International, Inc., LZI Hydrocodone Enzyme Immunoassay (K141055). The DRI® Hydrocodone Assay, calibrators, and controls are identical or similar to the predicate in terms of intended use, metho…
Lin-Zhi International, Inc.2014—

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231752ARK Hydrocodone AssayArk Diagnostics, Inc.20230
K173195DRI Hydrocodone AssayMicrogenics Corporation20180

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.0 years on median.
  • None of the 5 ancestors has a recall.
  • Of its descendants, K231752 (Ark Diagnostics, Inc., 2023) has gone furthest, with 0 descendants of its own.

How does its category compare?

All DJG devices

DJG Enzyme Immunoassay, Opiates: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · DJG

3in 2025

2007: 22008: 22009: 22010: 62011: 62012: 52013: 52014: 92015: 122016: 22017: 42018: 52019: 32020: 52021: 32022: 22023: 102024: 52025: 32026 (partial year): 3
2007high 12 (2015) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DJG

31in 2025

2010: 187 days median2011: 233 days median2012: 95 days median2013: 60 days median2014: 62 days median2015: 141 days median2017: 101 days median2018: 134 days median2019: 235 days median2020: 30 days median2021: 422 days median2023: 140 days median2024: 150 days median2025: 31 days median2026 (partial year): 157 days median
2007high 422 (2021) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DJG

2in 2025

2011: 12019: 12025: 2
2007high 2 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DJG

1in 2025

2011: 42012: 12013: 42015: 32016: 22017: 12018: 12021: 22022: 12024: 32025: 1
2007high 4 (2011) · low 0 (2007)2025

MAUDE reports naming product code DJG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—00
20092—00
2010618700
2011623314
201259501
201356004
201496200
20151214103
20162—02
2017410101
2018513401
2019323510
202053000
2021342202
20222—01
20231014000
2024515003
202533121
202631570—

What is it indicated for?

DRI® Hydrocodone Assay The DRI® Hydrocodone Assay is intended for the qualitative and semi-quantitative detection and estimation of Hydrocodone and its metabolites in human urine at a cutoff of 300 ng/mL. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of specimen for confirmation by a confirmatory method such as LC-MS/MS or GC-MS and permitting laboratories to establish quality control measures.…

Read the full statement (191 words)

DRI® Hydrocodone Assay

The DRI® Hydrocodone Assay is intended for the qualitative and semi-quantitative detection and estimation of Hydrocodone and its metabolites in human urine at a cutoff of 300 ng/mL. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of specimen for confirmation by a confirmatory method such as LC-MS/MS or GC-MS and permitting laboratories to establish quality control measures.

This assay provides a preliminary analytical test result. A more specific alternative chemical method must be used in order to confirm an analytical result. Gas chromatography/mass spectrometry (GC/MS) and Liquid Chromatography/tandem mass spectrometry (LC-MS/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

DRI® Hydrocodone Calibrators

The DRI® Hydrocodone Assay Calibrators are intended for the calibration of the DRI® Hydrocodone Assay. For In Vitro Diagnostic Use Only.

DRI® Hydrocodone Controls

The DRI® Hydrocodone Controls are unassayed quality control material intended for use in the DRI Hydrocodone Assay to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. For In Vitro Diagnostics Use Only

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K150502: DRI Hydrocodone Assay, DRI Hydrocodone Calibrators, DRI Hydrocodone Controls.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K150502

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026