510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

IntelliCuff

K150893510(k) TraditionalCleared Jan 29, 2016302-day reviewClass II21 CFR 868.5750Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hamilton Medical AG · BSK Cuff, Tracheal Tube, Inflatable

Cleared on Jan 29, 2016 as substantially equivalent to K092733 (Arm Medical Devices, Inc., 2010) after a 302-day review. Its family tree reaches back 3 generations to 1991. 5 later devices descend from it. It has been recalled 2 times.

Decided
Jan 29, 2016
Review
302 days
Class
Class II
Product code
BSK
Recalls
2
Predicates
1

Its longest line of descent runs back 3 generations to K912723 (1991). 6 ancestors in all, the oldest K791045 (1979). 5 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor1990: 1 other ancestor1979: 1 other ancestor2025: 1 descendant cleared2023: 1 descendant cleared2022: 1 descendant cleared2021: 1 descendant cleared2026: 1 descendant clearedReference device: K103803 · Hamilton-G5 · Hamilton Medical AG · 2011 · Class I recallReference device: K131774 · Hamilton-G5 · Hamilton Medical AG · 2013 · Class I recallK912723 · Posey Cufflator · Posey Co. · 1991K081805 · Cuffalert · Sunmed · 2008K092733 · PYTON · Arm Medical Devices, Inc. · 201019801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 3 other ancestors 1 generation forward: 4 descendants2 generations forward: 1 descendant K092733 · PYTON · Arm Medical Devices, Inc. · 2010K081805 · Cuffalert · Sunmed · 2008K912723 · Posey Cufflator · Posey Co. · 19910
  1. Root · 3 generations backK912723 Posey CufflatorPosey Co. · 1991
  2. 2 generations backK081805 CuffalertSunmed · 2008
  3. 1 generation backK092733 PYTONArm Medical Devices, Inc. · 2010
  4. This deviceK150893 IntelliCuffHamilton Medical AG · 2016
  5. Descendant · cites it directly · 1 of its ownK220854 TrachCuff Cuff ControllerAw Technologies Aps · 2023
  6. Descendant · 2 generations on · 0 of its ownK261499 TrachFlushAw Technologies Aps · 2026
  7. Descendant · cites it directly · 0 of its ownK243562 CuffGuardIdmed · 2025
  8. Descendant · cites it directly · 0 of its ownK221477 AG100sHospitech Respiration, Ltd. · 2022
  9. Descendant · cites it directly · 0 of its ownK202874 Puritan Bennett Cuff Pressure ManagerCovidien, LLC · 2021 · Recalled

Most influential descendants

Showing 5 of 5 · See all
  • K220854TrachCuff Cuff ControllerAw Technologies Aps · 2023 · cites it directly1descendant
  • K261499TrachFlushAw Technologies Aps · 2026 · 2 generations on0descendants
  • K243562CuffGuardIdmed · 2025 · cites it directly0descendants
  • K221477AG100sHospitech Respiration, Ltd. · 2022 · cites it directly0descendants
  • K202874Puritan Bennett Cuff Pressure ManagerCovidien, LLC · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K912723Posey CufflatorPosey Co.1991—
−2K081805CuffalertSunmed2008—
−1K092733PYTONArm Medical Devices, Inc.2010—
0 (this device)K150893IntelliCuffHamilton Medical AG2016—
+1K220854TrachCuff Cuff ControllerAw Technologies Aps2023—
+2K261499TrachFlushAw Technologies Aps2026—
+1K243562CuffGuardIdmed2025—
+1K221477AG100sHospitech Respiration, Ltd.2022—
+1K202874Puritan Bennett Cuff Pressure ManagerCovidien, LLC20211

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
199010
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202210
202310
202510
202610
FIG. 1 — Lineage of K150893, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BSK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSK device

  1. #1 K261499 TrachFlushscore 13.0 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K092733PYTON
Summary page 5
…II Product Code: BSK III. Predicate Device(s) Primary Predicate Device ARM MEDICAL DEVICES, INC PYTON Cuff Regulator; K092733 This predicate has not been subject to a design-related recall Reference Device Hamilton Medical AG…
Arm Medical Devices, Inc.2010—
†K103803Hamilton-G5Hamilton Medical AG2011Class I
†K131774Hamilton-G5Hamilton Medical AG2013Class I

Also cleared against K092733: 4 other devices →

What descends from it? (5)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K220854TrachCuff Cuff ControllerAw Technologies Aps20231
K243562CuffGuardIdmed20250
K221477AG100sHospitech Respiration, Ltd.20220
K202874Puritan Bennett Cuff Pressure ManagerCovidien, LLC20210

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 22, 2026Class IIZ-2625-2026Process designDue to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
Jun 20, 2017Class IIZ-0335-2018Component design/selectionIssue related to the performance of the motor in recalled product. During use, motor may cease to function.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 5.9 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K220854 (Aw Technologies Aps, 2023) has gone furthest, with 1 descendant of its own.
  • 4 other devices also cite its predicate K092733.

How does its category compare?

All BSK devices

BSK Cuff, Tracheal Tube, Inflatable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · BSK

2in 2025

2008: 12010: 12011: 32012: 22013: 12015: 12016: 42018: 22019: 12020: 22021: 12022: 12023: 12025: 22026 (partial year): 1
2007high 4 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSK

376in 2016

2011: 171 days median2016: 376 days median
2007high 376 (2016) · low 171 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSK

1in 2017

2013: 1, 1 Class I2017: 12026 (partial year): 1
2007high 1 (2013) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · BSK

46in 2025

2008: 22009: 32010: 52011: 152012: 622013: 162014: 52015: 12016: 12017: 32018: 42019: 22020: 12021: 92022: 42023: 512024: 1332025: 46
2007high 133 (2024) · low 0 (2007)2025

MAUDE reports naming product code BSK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—02
20090—03
20101—05
20113171015
20122—062
20131—1 (1)16
20140—05
20151—01
2016437601
20170—13
20182—04
20191—02
20202—01
20211—09
20221—04
20231—051
20240—0133
20252—046
20261—1—

What is it indicated for?

The IntelliCuff device is intended to continuously measure and automatically maintain the user-set cuff pressure of an endotracheal tube (ETT) or tracheostomy tube (TT) during mechanical ventilation. The device can be used with any mechanical ventilator, as follows: When used with a non-Hamilton Medical ventilator, IntelliCuff adjusts the cuff pressure to values set on the device. When used with a Hamilton Medical ventilator, IntelliCuff adjusts the cuff…

Read the full statement (147 words)

The IntelliCuff device is intended to continuously measure and automatically maintain the user-set cuff pressure of an endotracheal tube (ETT) or tracheostomy tube (TT) during mechanical ventilation.

The device can be used with any mechanical ventilator, as follows:

When used with a non-Hamilton Medical ventilator, IntelliCuff adjusts the cuff pressure to values set on the device.

When used with a Hamilton Medical ventilator, IntelliCuff adjusts the cuff pressure to values set either on the device or on the ventilator, depending on configuration.

The device is to be used during ventilation of adults, pediatrics, and neonates, who are intubated with ETT or TT, in the following areas:

In the intensive care ward or in the recovery room

In the operation room during intubation narcosis

For emergency medical care or primary care

During transport within and outside of the hospital

During transfer by rescue vehicles, ship, jet, or helicopter

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K150893: IntelliCuff.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K150893

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026