AirLife Infant Heated Wire Circuit

K151303510(k) TraditionalCleared Jan 21, 2016248-day reviewClass II21 CFR 868.5270Recalls 2 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Care Fusion · BZE Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer

Cleared on Jan 21, 2016 as substantially equivalent to K020332 (Fisher & Paykel Healthcare, Ltd., 2003) and 1 other predicate after a 248-day review. Its family tree reaches back 4 generations to 1992. 1 of the 9 devices in its family, counting itself, has a Class I recall. It has been recalled 2 times, including 2 Class I recalls.

Decided
Jan 21, 2016
Review
248 days
Class
Class II
Product code
BZE
Recalls
2 (2 Class I)
Predicates
2

Its longest line of descent runs back 4 generations to K913368 (1992). 8 ancestors in all, the oldest K902506 (1991). No later device cites it as a predicate yet. 1 of the 9 devices in this family, counting this one, has a Class I recall.

  1. Root · 4 generations backK913368 Models MR700/MR720/MR730 Dual Servo Resp Humid AccFisher & Paykel Electronics, Ltd. · 1992
  2. 3 generations backK953392 HC500 Servo-Controlled Heated Respiratory Humidifier (& Accessories)Fisher & Paykel Electronics, Ltd. · 1996
  3. 2 generations backK983112 Respiratory Humidifier, Model # MR850Fisher & Paykel Healthcare, Ltd. · 1998 · Recalled
  4. 1 generation backK020332 Respiratory Humidifier, Model MR850Fisher & Paykel Healthcare, Ltd. · 2003 · Recalled
  5. This deviceK151303 AirLife Infant Heated Wire CircuitCare Fusion · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K913368Models MR700/MR720/MR730 Dual Servo Resp Humid AccFisher & Paykel Electronics, Ltd.1992—
−3K953392HC500 Servo-Controlled Heated Respiratory Humidifier (& Accessories)Fisher & Paykel Electronics, Ltd.1996—
−2K983112Respiratory Humidifier, Model # MR850Fisher & Paykel Healthcare, Ltd.19985
−1K020332Respiratory Humidifier, Model MR850Fisher & Paykel Healthcare, Ltd.20031
0 (this device)K151303AirLife Infant Heated Wire CircuitCare Fusion2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199210
199410
200010
FIG. 1 — Lineage of K151303, 1992–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZE device

  1. #1 K252066 inspired™ Heated Breathing Circuit (510-008)…score 13.3 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020332Respiratory Humidifier, Model MR850
Summary page 4
…II Product Code: BZE 3. Predicate Device This submission demonstrates substantial equivalence to the Fisher and Paykel Heated Breathing Circuit K020332 cleared on July 7th, 2003 and AirLife Heated Ventilator and Anesthesia Breathing Circuits, K000697 cleared on March 30, 2000.…
Fisher & Paykel Healthcare, Ltd.20031
*K000697Modification to Airlife Heated Ventilator and Anesthesia Breathing Circuits
Summary page 4
…al equivalence to the Fisher and Paykel Heated Breathing Circuit K020332 cleared on July 7th, 2003 and AirLife Heated Ventilator and Anesthesia Breathing Circuits, K000697 cleared on March 30, 2000. The predicate devices have not been subject to a design-related recall. 4. Device Description T…
Allegiance Healthcare Corp.2000—

Also cleared against K020332: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 10, 2025Class IZ-1698-2025Component design/selectionAdapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.
Apr 10, 2025Class IZ-1697-2025Component design/selectionAdapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 3 companies; a new generation was cleared every 4.7 years on median.
  • 2 of the 8 ancestors have a recall of any class.
  • 4 other devices also cite its predicate K020332.

How does its category compare?

All BZE devices

BZE Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · BZE

1in 2025

2010: 22011: 12012: 22014: 12015: 12016: 52017: 12018: 12021: 12023: 12024: 22025: 12026 (partial year): 2
2007high 5 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZE

241in 2016

2016: 241 days median
2007high 241 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZE

No recall names BZE since 2007.

Adverse event reports per year · BZE

193in 2025

2007: 692008: 742009: 802010: 1262011: 1922012: 3342013: 3872014: 3382015: 1552016: 2142017: 2092018: 1032019: 1342020: 1232021: 1242022: 1522023: 1252024: 1722025: 193
2007high 387 (2013) · low 69 (2007)2025

MAUDE reports naming product code BZE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—069
20080—074
20090—080
20102—0126
20111—0192
20122—0334
20130—0387
20141—0338
20151—0155
201652410214
20171—0209
20181—0103
20190—0134
20200—0123
20211—0124
20220—0152
20231—0125
20242—0172
20251—0193
20262—0—

What is it indicated for?

The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient’s airway. The heated breathing circuit is used with Fisher and Paykel MR850 humidifier. The AirLife Infant Heated Wire Circuit is used with pediatric population, specifically the neonate (birth to 1 month of age) and infant (greater than 1 month to 2 years of age) pediatric subgroups. The product is single use device, non-sterile and used in professional healthcare environments and intra-hospital transport environments under a doctor’s supervision and by skilled clinicians. The AirLife Infant Heated Wire Circuit is used for flow rates greater than 4 LPM.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K151303: AirLife Infant Heated Wire Circuit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151303

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026