510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

IMMULITE 2000 TSI Assay, IMMULITE 2000 TSI Calibration Verification Material

K152061510(k) TraditionalCleared Mar 3, 2016223-day reviewClass II21 CFR 866.5870Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Healthcare Diagnostics, Inc. · JZO System, Test, Thyroid Autoantibody

Cleared on Mar 3, 2016 as substantially equivalent to K092229 (Diagnostic Hybrids, Inc., 2010) and 2 other predicates after a 223-day review. Its family tree reaches back 5 generations to 2000. It has been recalled once.

Decided
Mar 3, 2016
Review
223 days
Class
Class II
Product code
JZO
Recalls
1
Predicates
3

Its longest line of descent runs back 5 generations to K001015 (2000). 10 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2010: 1 other ancestor2008: 1 other ancestor2009: 1 other ancestor2004: 1 other ancestor2003: 1 other ancestorK001015 · Elecsys Thyroglobulin Calcheck · Roche Diagnostics Corp. · 2000K020883 · Elecsys Probnp Calcheck · Roche Diagnostics Corp. · 2002K031698 · Elecys SHBG Calcheck · Roche Diagnostics Corp. · 2003K040157 · Elecsys C-Peptide Calcheck · Roche Diagnostics Corp. · 2004K080643 · Elecsys Anti-TSHR Calcheck · Roche Diagnostics · 20082000200520102015
  1. Root · 5 generations backK001015 Elecsys Thyroglobulin CalcheckRoche Diagnostics Corp. · 2000
  2. 4 generations backK020883 Elecsys Probnp CalcheckRoche Diagnostics Corp. · 2002
  3. 3 generations backK031698 Elecys SHBG CalcheckRoche Diagnostics Corp. · 2003
  4. 2 generations backK040157 Elecsys C-Peptide CalcheckRoche Diagnostics Corp. · 2004
  5. 1 generation backK080643 Elecsys Anti-TSHR CalcheckRoche Diagnostics · 2008
  6. This deviceK152061 IMMULITE 2000 TSI Assay, IMMULITE 2000 TSI Calibration Verification…Siemens Healthcare Diagnostics, Inc. · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K001015Elecsys Thyroglobulin CalcheckRoche Diagnostics Corp.2000—
−4K020883Elecsys Probnp CalcheckRoche Diagnostics Corp.2002—
−3K031698Elecys SHBG CalcheckRoche Diagnostics Corp.2003—
−2K040157Elecsys C-Peptide CalcheckRoche Diagnostics Corp.2004—
−1K080643Elecsys Anti-TSHR CalcheckRoche Diagnostics2008—
0 (this device)K152061IMMULITE 2000 TSI Assay, IMMULITE 2000 TSI Calibration Verification MaterialSiemens Healthcare Diagnostics, Inc.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200310
200410
200810
200910
201010
FIG. 1 — Lineage of K152061, 2000–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JZO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JZO device

  1. #1 K250250 ADVIA Centaur Anti-Thyroid Peroxidase IIscore 10.5 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K092229Thyretain TM Tsi Reporter Bioassay
Summary page 5
…ting MAb (M22) • Multi-Diluent 2 J. Substantial Equivalence Information: 1. Predicate device names(s) and 510(k) number(s): Thyretain TSI Reporter BioAssay, k092229 Elecsys Anti-TSHR Immunoassay, k080092 Elecsys Anti-TSHR CalCheck, k080643 2. Comparison with predicate:…
Diagnostic Hybrids, Inc.2010—
*K080092Elecsys Anti-TSHR Immunoassay, Elecsys Precicontrol Thyroab
Summary page 5
…antial Equivalence Information: 1. Predicate device names(s) and 510(k) number(s): Thyretain TSI Reporter BioAssay, k092229 Elecsys Anti-TSHR Immunoassay, k080092 Elecsys Anti-TSHR CalCheck, k080643 2. Comparison with predicate: ASSAY SIMILARITIES – Pred…
Roche Diagnostics Corp.2008—
*K080643Elecsys Anti-TSHR Calcheck
Summary page 5
…icate device names(s) and 510(k) number(s): Thyretain TSI Reporter BioAssay, k092229 Elecsys Anti-TSHR Immunoassay, k080092 Elecsys Anti-TSHR CalCheck, k080643 2. Comparison with predicate: ASSAY SIMILARITIES – Predicate 1 Item…
Roche Diagnostics2008—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 11, 2022Class IIZ-1820-2022Under Investigation by firmAn average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for erroneously depressed patient results. This may lead to a…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 2.6 years on median.
  • None of the 10 ancestors has a recall.

How does its category compare?

All JZO devices

JZO System, Test, Thyroid Autoantibody: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · JZO

1in 2025

2008: 22009: 22010: 12016: 22020: 12023: 22024: 12025: 1
2007high 2 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JZO

Too few JZO 510(k)s in any year since 2007 to give a median.

Recalls per year · JZO

1in 2025

2025: 1
2007high 1 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JZO

44in 2025

2010: 12011: 52012: 42013: 42014: 22015: 42016: 82017: 122018: 352019: 392020: 172021: 192022: 252023: 332024: 292025: 44
2007high 44 (2025) · low 0 (2007)2025

MAUDE reports naming product code JZO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—00
20092—00
20101—01
20110—05
20120—04
20130—04
20140—02
20150—04
20162—08
20170—012
20180—035
20190—039
20201—017
20210—019
20220—025
20232—033
20241—029
20251—144
20260—0—

What is it indicated for?

The IMMULITE® 2000 TSI (thyroid-stimulating immunoglobulins) Assay is an in vitro diagnostic immunoassay for the semi-quantitative determination of thyroid stimulating autoantibodies specific to thyroid stimulating hormone receptors (TSHR) in human serum (including Serum Separator tubes) or plasma (K2-EDTA or lithium heparin). The IMMULITE® 2000 TSI Assay is for use on the IMMULITE® 2000 system. The measurement of thyroid stimulating autoantibodies, in…

Read the full statement (111 words)

The IMMULITE® 2000 TSI (thyroid-stimulating immunoglobulins) Assay is an in vitro diagnostic immunoassay for the semi-quantitative determination of thyroid stimulating autoantibodies specific to thyroid stimulating hormone receptors (TSHR) in human serum (including Serum Separator tubes) or plasma (K2-EDTA or lithium heparin). The IMMULITE® 2000 TSI Assay is for use on the IMMULITE® 2000 system. The measurement of thyroid stimulating autoantibodies, in conjunction with other clinical and laboratory findings, is used as an aid in the diagnosis of patients suspected of having Graves' disease.

The IMMULITE® TSI Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE® TSI Assay on the IMMULITE® 2000 Systems.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K152061: IMMULITE 2000 TSI Assay, IMMULITE 2000 TSI Calibration Verification Material.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152061

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026