510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Elecsys Anti-TSHR

K193313510(k) TraditionalCleared Feb 27, 202090-day reviewClass II21 CFR 866.5870Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics · JZO System, Test, Thyroid Autoantibody

Cleared on Feb 27, 2020 as substantially equivalent to K080092 (Roche Diagnostics Corp., 2008) after a 90-day review. Its family tree reaches back 2 generations to 2004. It has been recalled once.

Decided
Feb 27, 2020
Review
90 days
Class
Class II
Product code
JZO
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K033454 (2004). No later device cites it as a predicate yet.

  1. Root · 2 generations backK033454 Brahms Diagnostica Lumitest Trak HumanBrahms Diagnostica, LLC · 2004
  2. 1 generation backK080092 Elecsys Anti-TSHR Immunoassay, Elecsys Precicontrol ThyroabRoche Diagnostics Corp. · 2008
  3. This deviceK193313 Elecsys Anti-TSHRRoche Diagnostics · 2020
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K033454Brahms Diagnostica Lumitest Trak HumanBrahms Diagnostica, LLC2004—
−1K080092Elecsys Anti-TSHR Immunoassay, Elecsys Precicontrol ThyroabRoche Diagnostics Corp.2008—
0 (this device)K193313Elecsys Anti-TSHRRoche Diagnostics2020—
FIG. 1 — Lineage of K193313, 2004–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JZO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JZO device

  1. #1 K250250 ADVIA Centaur Anti-Thyroid Peroxidase IIscore 10.5 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080092Elecsys Anti-TSHR Immunoassay, Elecsys Precicontrol Thyroab
Summary page 5
…JIT, 21CFR862.1150, EXEMPT Regulation Numbers JJX, 21CFR862.1660, EXEMPT Predicate Devices Elecsys Anti-TSHR Immunoassay (K080092) For the Elecsys Anti-TSHR, the establishment registration number for Roche…
Roche Diagnostics Corp.2008—

Also cleared against K080092: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 21, 2025Class IIZ-1009-2026Process controlIssues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 11.6 years on median.
  • None of the 2 ancestors has a recall.
  • 2 other devices also cite its predicate K080092.

How does its category compare?

All JZO devices

JZO System, Test, Thyroid Autoantibody: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · JZO

1in 2025

2008: 22009: 22010: 12016: 22020: 12023: 22024: 12025: 1
2007high 2 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JZO

Too few JZO 510(k)s in any year since 2007 to give a median.

Recalls per year · JZO

1in 2025

2025: 1
2007high 1 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JZO

44in 2025

2010: 12011: 52012: 42013: 42014: 22015: 42016: 82017: 122018: 352019: 392020: 172021: 192022: 252023: 332024: 292025: 44
2007high 44 (2025) · low 0 (2007)2025

MAUDE reports naming product code JZO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—00
20092—00
20101—01
20110—05
20120—04
20130—04
20140—02
20150—04
20162—08
20170—012
20180—035
20190—039
20201—017
20210—019
20220—025
20232—033
20241—029
20251—144
20260—0—

What is it indicated for?

Immunoassay for the in vitro quantitative determination of autoantibodies to thyroid stimulating hormone (TSH) receptor in human serum using a human thyroid stimulating monoclonal antibody. The anti‑TSH receptor determination is used in the assessment of patients suspected of Graves' disease (autoimmune hyperthyroidism).

The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e 601 immunoassay analyzers.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K193313: Elecsys Anti-TSHR.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K193313

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026