Hydroven 12, Hydroven 12 Garments

K152228510(k) TraditionalCleared Apr 4, 2016241-day reviewClass II21 CFR 870.5800

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Huntleigh Healthcare, Ltd. · JOW Sleeve, Limb, Compressible

Cleared on Apr 4, 2016 as substantially equivalent to K013331 (Mego Afek, 2001) after a 241-day review. Its family tree reaches back 2 generations to 1981.

Decided
Apr 4, 2016
Review
241 days
Class
Class II
Product code
JOW
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K810338 (1981). 3 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1988: 1 other ancestorK810338 · Lympha Press · Mego Afek · 1981K013331 · Lympha Press Plus Device · Mego Afek · 20011980199020002010
Other ancestors, by generation 2 generations back: 1 other ancestor K013331 · Lympha Press Plus Device · Mego Afek · 2001K810338 · Lympha Press · Mego Afek · 19810
  1. Root · 2 generations backK810338 Lympha PressMego Afek · 1981
  2. 1 generation backK013331 Lympha Press Plus DeviceMego Afek · 2001
  3. This deviceK152228 Hydroven 12, Hydroven 12 GarmentsHuntleigh Healthcare, Ltd. · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K810338Lympha PressMego Afek1981—
−1K013331Lympha Press Plus DeviceMego Afek2001—
0 (this device)K152228Hydroven 12, Hydroven 12 GarmentsHuntleigh Healthcare, Ltd.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
FIG. 1 — Lineage of K152228, 1981–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOW device

  1. #1 K253555 Flexitouch® Plus Advanced Pneumatic…score 14.8 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K013331Lympha Press Plus Device
Summary page 4
…JOW 870.5800 Classification Name: Sleeve, Limb, Compressible Predicate Device: Lympha Press Plus and compression Garments (K013331) manufactured by Mego Afek AC Ltd. Kibbutz Afek, 30042, Israel Indications for Use: In…
Mego Afek2001—

Also cleared against K013331: 5 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 35 years and passes through 2 companies; a new generation was cleared every 20.1 years on median.
  • None of the 3 ancestors has a recall.
  • 5 other devices also cite its predicate K013331.

How does its category compare?

All JOW devices

JOW Sleeve, Limb, Compressible: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · JOW

7in 2025

2007: 62008: 42009: 32010: 182011: 112012: 132013: 172014: 162015: 102016: 92017: 92018: 112019: 112020: 142021: 162022: 72023: 62024: 82025: 72026 (partial year): 6
2007high 18 (2010) · low 3 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOW

168in 2025

2007: 66 days median2008: 112 days median2009: 84 days median2010: 71 days median2011: 104 days median2012: 63 days median2013: 94 days median2014: 91 days median2015: 160 days median2016: 121 days median2017: 116 days median2018: 136 days median2019: 88 days median2020: 158 days median2021: 132 days median2022: 80 days median2023: 160 days median2024: 197 days median2025: 168 days median2026 (partial year): 252 days median
2007high 252 (2026) · low 63 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JOW

1in 2019

2008: 72010: 22011: 32012: 22017: 12019: 1
2007high 7 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOW

24in 2025

2007: 122008: 362009: 452010: 662011: 502012: 1002013: 1692014: 932015: 1372016: 1502017: 1492018: 1142019: 1022020: 502021: 342022: 272023: 572024: 302025: 24
2007high 169 (2013) · low 12 (2007)2025

MAUDE reports naming product code JOW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007666012
20084112736
2009384045
20101871266
201111104350
201213632100
201317940169
20141691093
2015101600137
201691210150
201791161149
2018111360114
201911881102
202014158050
202116132034
2022780027
20236160057
20248197030
20257168024
202662520—

What is it indicated for?

To enhance fluid flow into and along the vascular and lymphatic vessels. As a treatment for venous and lymphatic disorders.

As an effective in the treatment of the following clinical conditions, when combined with an individualized monitoring programme:

Lymphedema.

Primary and secondary (including post-surgery, radio or chemotherapy).

Chronic Edema of Venous Origin

Chronic venous insufficiency.

Acute and chronic wounds including venous leg ulcers and post-surgical wounds.

IPC may also be beneficial in the management of:

Lipoedema.

Varicose veins

Postoperative Venous Ligation or Stripping

Sports injuries

Post Traumatic Edema

Lymphatic Filariasis

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K152228: Hydroven 12, Hydroven 12 Garments.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152228

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026