510(k) · Cleared
Recalled · Class II · 2026RADspeed Pro
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Shimadzu Corporation · KPR System, X-Ray, Stationary
Cleared on Jun 20, 2016 as substantially equivalent to K093427 (Fujifilm Medical System U.S.A., Inc., 2010) and 3 other predicates after a 315-day review. Its family tree reaches back 6 generations to 1986. 3 later devices descend from it. It has been recalled 2 times.
- Decided
- Jun 20, 2016
- Review
- 315 days
- Class
- Class II
- Product code
- KPR
- Recalls
- 2
- Predicates
- 4
Its longest line of descent runs back 6 generations to K862120 (1986). 13 ancestors in all. 3 later devices descend from it, 2 citing it directly.
- Root · 6 generations backK862120 SCX SystemGeneral Electric Co. · 1986 · Recalled
- 5 generations backK982196 Digital Radiographic Imaging SystemGE Medical Systems · 1998
- 4 generations backK992066 Revolution TX/I Digital Radiographic Table SystemGE Medical Systems · 1999
- 3 generations backK012389 Revolution XR/D Digital Radiographic Imaging SystemGe Medical Systems, Inc. · 2001 · Recalled
- 2 generations backK051967 Revolution XR/D Digital Radiographic Imaging System with TomosynthesisGE Healthcare · 2005 · Recalled
- 1 generation backK121499 Fujifilm Tomosynthesis Option for FDR Acselerate Stationary X-Ray…Fujifilm Medical System U.S.A., Inc. · 2012
- This deviceK152244 RADspeed ProShimadzu Corporation · 2016
- Descendant · cites it directly · 1 of its ownK173517 RADspeed fitShimadzu Corporation · 2018
- Descendant · cites it directly · 0 of its ownK233722 RADspeed ProShimadzu Corporation · 2024 · Recalled
- Descendant · 2 generations on · 0 of its ownK220518 Skanrad 400Skanray Technologies Limited · 2022
Most influential descendants
Showing 3 of 3 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K862120 | SCX System | General Electric Co. | 1986 | 3 |
| −5 | K982196 | Digital Radiographic Imaging System | GE Medical Systems | 1998 | — |
| −4 | K992066 | Revolution TX/I Digital Radiographic Table System | GE Medical Systems | 1999 | — |
| −3 | K012389 | Revolution XR/D Digital Radiographic Imaging System | Ge Medical Systems, Inc. | 2001 | 4 |
| −2 | K051967 | Revolution XR/D Digital Radiographic Imaging System with Tomosynthesis | GE Healthcare | 2005 | 9 |
| −1 | K121499 | Fujifilm Tomosynthesis Option for FDR Acselerate Stationary X-Ray System | Fujifilm Medical System U.S.A., Inc. | 2012 | — |
| 0 (this device) | K152244 | RADspeed Pro | Shimadzu Corporation | 2016 | — |
| +1 | K173517 | RADspeed fit | Shimadzu Corporation | 2018 | — |
| +1 | K233722 | RADspeed Pro | Shimadzu Corporation | 2024 | 1 |
| +2 | K220518 | Skanrad 400 | Skanray Technologies Limited | 2022 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1988 | 1 | 0 |
| 1994 | 1 | 0 |
| 1995 | 1 | 0 |
| 2001 | 1 | 0 |
| 2010 | 2 | 0 |
| 2012 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2018 | 1 | 0 |
| 2022 | 1 | 0 |
| 2024 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KPR device
- #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K093427 | FDR Acselerate System, DR-ID 200 Flat Panel DetectorSummary page 4…ay system (21 CFR§892.1680) Classification Panel : Radiology Regulatory Class : Class II Product Code : KPR 6) Predicate Device Primary K093427 FDR ACSELERATE SYSTEM, DR-ID 200 FUJIFILM FLAT PANEL DETECTOR… | 2010 |
| * | K121499 | Fujifilm Tomosynthesis Option for FDR Acselerate Stationary X-Ray SystemSummary page 4…K093427 FDR ACSELERATE SYSTEM, DR-ID 200 FUJIFILM FLAT PANEL DETECTOR CORPORATION Secondary K121499 FUJIFILM TOMOSYNTHESIS OPTION FUJIFILM FOR FDR ACSELERATE STATIONARY… | 2012 |
| * | K101686 | FDR Image Stitching OptionSummary page 4…OPTION FUJIFILM FOR FDR ACSELERATE STATIONARY CORPORATION X-RAY SYSTEM Tertiary K101686 FDR IMAGE STITCHING OPTION FUJIFILM CORPORATION Quaternary… | 2010 |
| * | K122454 | Fujifilm Dual Energy Subtraction (Des) Software OptionSummary page 4…X-RAY SYSTEM Tertiary K101686 FDR IMAGE STITCHING OPTION FUJIFILM CORPORATION Quaternary K122454 FUJIFILM DUAL ENERGY SUBTRACTION FUJIFILM (DES) SOFTWARE OPTION CORPORATION No refere… | 2012 |
What descends from it? (3)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 20, 2026 | Class II | Z-2294-2026 | Component design/selection | Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off. |
| May 23, 2017 | Class II | Z-2654-2017 | Device Design | Cracks may occur over time on the tube holding shaft near the base of the tube mounting flange. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 30 years and passes through 6 companies; a new generation was cleared every 4.0 years on median.
- 4 of the 13 ancestors have a recall of any class.
- Of its descendants, K173517 (Shimadzu Corporation, 2018) has gone furthest, with 1 descendant of its own.
- 2 other devices also cite its predicate K093427.
How does its category compare?
All KPR devicesKPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.
FDA decisions per year · KPR
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KPR
155in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KPR
7in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · KPR
39in 2025
MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 31 | 10 | 120 |
| 2008 | 15 | 48 | 10 | 95 |
| 2009 | 9 | 35 | 5 | 61 |
| 2010 | 9 | 126 | 3 | 82 |
| 2011 | 9 | 152 | 5 | 103 |
| 2012 | 11 | 58 | 1 | 74 |
| 2013 | 5 | 163 | 5 | 24 |
| 2014 | 10 | 108 | 8 | 34 |
| 2015 | 18 | 81 | 1 | 59 |
| 2016 | 11 | 84 | 4 | 18 |
| 2017 | 6 | 75 | 3 | 31 |
| 2018 | 15 | 30 | 4 | 36 |
| 2019 | 12 | 33 | 1 | 77 |
| 2020 | 16 | 39 | 1 | 41 |
| 2021 | 6 | 55 | 0 | 37 |
| 2022 | 8 | 43 | 2 | 27 |
| 2023 | 9 | 76 | 4 | 39 |
| 2024 | 11 | 162 | 1 | 34 |
| 2025 | 8 | 155 | 7 | 39 |
| 2026 | 6 | 197 | 0 | — |
What is it indicated for?
The RADspeed Pro is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The RADspeed Pro enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position. The…
Read the full statement (160 words)
The RADspeed Pro is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The RADspeed Pro enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position.
The RADspeed Pro uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes. The Tomosynthesis option is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition.
The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals.
The device is not intended for mammographic applications.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K152244 · printed Nov 6, 2026
Cite this page
510(k) Report. “K152244: RADspeed Pro.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152244
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026