510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

RADspeed Pro

K233722510(k) TraditionalCleared Feb 23, 202494-day reviewClass II21 CFR 892.1680Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shimadzu Corporation · KPR System, X-Ray, Stationary

Cleared on Feb 23, 2024 as substantially equivalent to K152244 (Shimadzu Corporation, 2016) after a 94-day review. Its family tree reaches back 7 generations to 1986. It has been recalled once.

Decided
Feb 23, 2024
Review
94 days
Class
Class II
Product code
KPR
Recalls
1
Predicates
1

Its longest line of descent runs back 7 generations to K862120 (1986). 14 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 7 generations backK862120 SCX SystemGeneral Electric Co. · 1986 · Recalled
  2. 6 generations backK982196 Digital Radiographic Imaging SystemGE Medical Systems · 1998
  3. 5 generations backK992066 Revolution TX/I Digital Radiographic Table SystemGE Medical Systems · 1999
  4. 4 generations backK012389 Revolution XR/D Digital Radiographic Imaging SystemGe Medical Systems, Inc. · 2001 · Recalled
  5. 3 generations backK051967 Revolution XR/D Digital Radiographic Imaging System with TomosynthesisGE Healthcare · 2005 · Recalled
  6. 2 generations backK121499 Fujifilm Tomosynthesis Option for FDR Acselerate Stationary X-Ray…Fujifilm Medical System U.S.A., Inc. · 2012
  7. 1 generation backK152244 RADspeed ProShimadzu Corporation · 2016 · Recalled
  8. This deviceK233722 RADspeed ProShimadzu Corporation · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K862120SCX SystemGeneral Electric Co.19863
−6K982196Digital Radiographic Imaging SystemGE Medical Systems1998—
−5K992066Revolution TX/I Digital Radiographic Table SystemGE Medical Systems1999—
−4K012389Revolution XR/D Digital Radiographic Imaging SystemGe Medical Systems, Inc.20014
−3K051967Revolution XR/D Digital Radiographic Imaging System with TomosynthesisGE Healthcare20059
−2K121499Fujifilm Tomosynthesis Option for FDR Acselerate Stationary X-Ray SystemFujifilm Medical System U.S.A., Inc.2012—
−1K152244RADspeed ProShimadzu Corporation20162
0 (this device)K233722RADspeed ProShimadzu Corporation2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199410
199510
200110
201020
201210
FIG. 1 — Lineage of K233722, 1986–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K152244RADspeed Pro
Summary page 4
…Regulation Number: 21CFR892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: II Product Code: KPR, MQB 2. Equivalent legally marketed device: K152244 Trade/Device Name: RADspeed PRO Manufacturer: Shimadzu. Regulation Number: 21CFR892.1680 Regulation Name: Stationary…
Shimadzu Corporation20162

Also cleared against K152244: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 20, 2026Class IIZ-2294-2026Component design/selectionElectromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 38 years and passes through 6 companies; a new generation was cleared every 4.0 years on median.
  • 5 of the 14 ancestors have a recall of any class.
  • 1 other device also cites its predicate K152244.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

The RADspeed Pro is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The RADspeed Pro enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position. The…

Read the full statement (160 words)

The RADspeed Pro is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The RADspeed Pro enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position. The RADspeed Pro uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes. The Tomosynthesis option is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition. The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals. The device is not intended for mammographic applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K233722: RADspeed Pro.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K233722

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026