510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Infinity Delta, Infinity Delta XL, Infinity Kappa

K152407510(k) SpecialCleared Sep 23, 201529-day reviewClass II21 CFR 870.1025Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Draeger Medical Sytems, Inc. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Sep 23, 2015 as substantially equivalent to K070566 (Draeger Medical Systems, Inc., 2007) after a 29-day review. Its family tree reaches back 10 generations to 1992. It has been recalled 4 times.

Decided
Sep 23, 2015
Review
29 days
Class
Class II
Product code
MHX
Recalls
4
Predicates
1

Its longest line of descent runs back 10 generations to K923682 (1992). 58 ancestors in all, the oldest K844187 (1985). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2005: 3 other ancestors2006: 1 other ancestor2004: 7 other ancestors2003: 4 other ancestors2002: 4 other ancestors2001: 2 other ancestors2000: 4 other ancestors1999: 1 other ancestor1998: 4 other ancestors1997: 3 other ancestors1993: 2 other ancestors1996: 4 other ancestors1995: 2 other ancestors1992: 1 other ancestor1991: 2 other ancestors1990: 3 other ancestors1985: 1 other ancestorReference device: K100941 · Infinity Mcable - Mainstream CO2 · Draeger Medical AG & Co. KG · 2010K923682 · Model M1032A Vuelink Interface Plug-in Module · Hewlett-Packard Co. · 1992K970368 · SC9000/SC9015 Medical Information Bus (Mis) Protocol Converter · Siemens Medical Solutions USA, Inc. · 1997K991661 · Siemens Medical Information Bus (Mib) Protocol Converter · Siemens Medical Solutions USA, Inc. · 1999K012461 · Siemens Infinity Mib II Duo Protocol Converter · Siemens Medical Solutions USA, Inc. · 2001K020277 · Siemens Medical Infromation Bus (Mib II) Protocol Converter · Siemens Medical Solutions USA, Inc. · 2002K022766 · Siemens Medical Information Bus (Mib/Mib II/Mib Duo) Protocol Coverter · Siemens Medical Solutions USA, Inc. · 2002K033807 · Infinityedical Information Bus (Mib/Mib II and Mib Duo) Protocol Converter · Draeger Medical Systems, Inc. · 2004K043439 · Infinity Modular Monitors with VF5 Modifications (Delta/Kappa/Delta XL/Vista XL/Gamma X XL and SC 7000/8000/9000XL · Draeger Medical Systems, Inc. · 2005K060039 · Infinity Delta and Kappa Series Monitors with VF6 Modifications · Draeger Medical Systems, Inc. · 2006 · recalledK070566 · Infinity Delta/Delta XL/ Kappa/ Gamma X XL/ Vista XL/ SC7000/SC8000/ SC9000XL · Draeger Medical Systems, Inc. · 2007 · recalled1985199019952000200520102015
Other ancestors, by generation 2 generations back: 1 other ancestor3 generations back: 8 other ancestors4 generations back: 20 other ancestors5 generations back: 9 other ancestors6 generations back: 4 other ancestors7 generations back: 6 other ancestors K070566 · Infinity Delta/Delta XL/ Kappa/ Gamma X XL/ Vista XL/ SC7000/SC8000/ SC9000XL · Draeger Medical Systems, Inc. · 2007 · recalledK060039 · Infinity Delta and Kappa Series Monitors with VF6 Modifications · Draeger Medical Systems, Inc. · 2006 · recalledK043439 · Infinity Modular Monitors with VF5 Modifications (Delta/Kappa/Delta XL/Vista XL/Gamma X XL and SC 7000/8000/9000XL · Draeger Medical Systems, Inc. · 2005K033807 · Infinityedical Information Bus (Mib/Mib II and Mib Duo) Protocol Converter · Draeger Medical Systems, Inc. · 2004K022766 · Siemens Medical Information Bus (Mib/Mib II/Mib Duo) Protocol Coverter · Siemens Medical Solutions USA, Inc. · 2002K020277 · Siemens Medical Infromation Bus (Mib II) Protocol Converter · Siemens Medical Solutions USA, Inc. · 2002K012461 · Siemens Infinity Mib II Duo Protocol Converter · Siemens Medical Solutions USA, Inc. · 2001K991661 · Siemens Medical Information Bus (Mib) Protocol Converter · Siemens Medical Solutions USA, Inc. · 1999K970368 · SC9000/SC9015 Medical Information Bus (Mis) Protocol Converter · Siemens Medical Solutions USA, Inc. · 1997K923682 · Model M1032A Vuelink Interface Plug-in Module · Hewlett-Packard Co. · 1992−10−50
  1. Root · 10 generations backK923682 Model M1032A Vuelink Interface Plug-in ModuleHewlett-Packard Co. · 1992
  2. 6 generations between
  3. 3 generations backK043439 Infinity Modular Monitors with VF5 Modifications (Delta/Kappa/Delta…Draeger Medical Systems, Inc. · 2005
  4. 2 generations backK060039 Infinity Delta and Kappa Series Monitors with VF6 ModificationsDraeger Medical Systems, Inc. · 2006 · Recalled
  5. 1 generation backK070566 Infinity Delta/Delta XL/ Kappa/ Gamma X XL/ Vista XL/ SC7000/SC8000/…Draeger Medical Systems, Inc. · 2007 · Recalled
  6. This deviceK152407 Infinity Delta, Infinity Delta XL, Infinity KappaDraeger Medical Sytems, Inc. · 2015
Show all 10 generations
  1. Root · 10 generations backK923682 Model M1032A Vuelink Interface Plug-in ModuleHewlett-Packard Co. · 1992
  2. 9 generations backK970368 SC9000/SC9015 Medical Information Bus (Mis) Protocol ConverterSiemens Medical Solutions USA, Inc. · 1997
  3. 8 generations backK991661 Siemens Medical Information Bus (Mib) Protocol ConverterSiemens Medical Solutions USA, Inc. · 1999
  4. 7 generations backK012461 Siemens Infinity Mib II Duo Protocol ConverterSiemens Medical Solutions USA, Inc. · 2001
  5. 6 generations backK020277 Siemens Medical Infromation Bus (Mib II) Protocol ConverterSiemens Medical Solutions USA, Inc. · 2002
  6. 5 generations backK022766 Siemens Medical Information Bus (Mib/Mib II/Mib Duo) Protocol CoverterSiemens Medical Solutions USA, Inc. · 2002
  7. 4 generations backK033807 Infinityedical Information Bus (Mib/Mib II and Mib Duo) Protocol…Draeger Medical Systems, Inc. · 2004
  8. 3 generations backK043439 Infinity Modular Monitors with VF5 Modifications (Delta/Kappa/Delta…Draeger Medical Systems, Inc. · 2005
  9. 2 generations backK060039 Infinity Delta and Kappa Series Monitors with VF6 ModificationsDraeger Medical Systems, Inc. · 2006 · Recalled
  10. 1 generation backK070566 Infinity Delta/Delta XL/ Kappa/ Gamma X XL/ Vista XL/ SC7000/SC8000/…Draeger Medical Systems, Inc. · 2007 · Recalled
  11. This deviceK152407 Infinity Delta, Infinity Delta XL, Infinity KappaDraeger Medical Sytems, Inc. · 2015
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K923682Model M1032A Vuelink Interface Plug-in ModuleHewlett-Packard Co.1992—
−9K970368SC9000/SC9015 Medical Information Bus (Mis) Protocol ConverterSiemens Medical Solutions USA, Inc.1997—
−8K991661Siemens Medical Information Bus (Mib) Protocol ConverterSiemens Medical Solutions USA, Inc.1999—
−7K012461Siemens Infinity Mib II Duo Protocol ConverterSiemens Medical Solutions USA, Inc.2001—
−6K020277Siemens Medical Infromation Bus (Mib II) Protocol ConverterSiemens Medical Solutions USA, Inc.2002—
−5K022766Siemens Medical Information Bus (Mib/Mib II/Mib Duo) Protocol CoverterSiemens Medical Solutions USA, Inc.2002—
−4K033807Infinityedical Information Bus (Mib/Mib II and Mib Duo) Protocol ConverterDraeger Medical Systems, Inc.2004—
−3K043439Infinity Modular Monitors with VF5 Modifications (Delta/Kappa/Delta XL/Vista…Draeger Medical Systems, Inc.2005—
−2K060039Infinity Delta and Kappa Series Monitors with VF6 ModificationsDraeger Medical Systems, Inc.20061
−1K070566Infinity Delta/Delta XL/ Kappa/ Gamma X XL/ Vista XL/ SC7000/SC8000/ SC9000XLDraeger Medical Systems, Inc.20076
0 (this device)K152407Infinity Delta, Infinity Delta XL, Infinity KappaDraeger Medical Sytems, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
199030
199120
199210
199320
199520
199640
199730
199840
199910
200040
200120
200240
200340
200470
200530
200610
FIG. 1 — Lineage of K152407, 1992–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K070566Infinity Delta/Delta XL/ Kappa/ Gamma X XL/ Vista XL/ SC7000/SC8000/ SC9000XL
Summary page 7
…II 870.2775 Identification of Legally Marketed Devices to which the equivalence is claimed: 510(k) number Trade name INFINITY DELTA/DELTA XL/ KAPPA/ K070566 GAMMA X XL/ VISTA XL/ Predicate Device SC7000/SC8000/ SC9000XL Device Description: Infi…
Draeger Medical Systems, Inc.20076
†K100941Infinity Mcable - Mainstream CO2Draeger Medical AG & Co. KG2010—

Also cleared against K070566: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 8, 2019Class IIZ-0974-2019Software designThe firm is recalling their Delta family of patient monitors software due to cybersecurity vulnerabilities, which may cause the device to reboot, lose functionality, and/or lose communication.
Mar 28, 2017Class IIZ-1773-2017Under Investigation by firmIt was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.
Mar 28, 2017Class IIZ-1774-2017Under Investigation by firmIt was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.
Mar 28, 2017Class IIZ-1772-2017Under Investigation by firmIt was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 1.5 years on median.
  • 4 of the 58 ancestors have a recall of any class.
  • 1 other device also cites its predicate K070566.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The Infinity Delta Series Monitor (Delta/Delta XL/Kappa) are capable of monitoring: • Heart rate • Respiration rate • Invasive pressure • Non-invasive pressure • Arrhythmia • Temperature • Cardiac output • Arterial oxygen saturation • Pulse rate • Apnea • ST Segment Analysis • 12-Lead ST Segment Analysis • tcpO2/tcpCO2 • EEG signals • FiO2 • etCO2 • Respiratory mechanics • Anesthetic agents • Neuromuscular transmission The devices are intended to be used…

Read the full statement (226 words)

The Infinity Delta Series Monitor (Delta/Delta XL/Kappa) are capable of

monitoring:

• Heart rate

• Respiration rate

• Invasive pressure

• Non-invasive pressure

• Arrhythmia

• Temperature

• Cardiac output

• Arterial oxygen saturation

• Pulse rate

• Apnea

• ST Segment Analysis

• 12-Lead ST Segment Analysis

• tcpO2/tcpCO2

• EEG signals

• FiO2

• etCO2

• Respiratory mechanics

• Anesthetic agents

• Neuromuscular transmission

The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e.

Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their

professional assessment of the patient's medical condition.

The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal

populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are

intended for use in the adult and pediatric populations only; and tcpO2, which for the neonatal population, is to only be

used when the patient is not under gas anesthesia.

For combination with Scio gas module:

Scio gas module samples breathing gases from adults and pediatrics. The gas module continuously measure the content of

CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any

mixture and communicates real time and derived gas information to the Infinity monitors.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K152407: Infinity Delta, Infinity Delta XL, Infinity Kappa.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152407

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026