510(k) Report

510(k) · Cleared

Cleared

Enhance Allograft Wedge Instrument Tray

K152969510(k) TraditionalCleared Mar 31, 2016175-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Conmed Corporation · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Mar 31, 2016 as substantially equivalent to K090560 (Conmed Linvatec, 2009) after a 175-day review. Its family tree reaches back 2 generations to 2008.

Decided
Mar 31, 2016
Review
175 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K080531 (2008). No later device cites it as a predicate yet.

  1. Root · 2 generations backK080531 Conmed Linvatec Microfracture Instrument Sterilization TrayConmed Linvatec · 2008
  2. 1 generation backK090560 Conmed Linvatec Anodized Aluminum Sterilization TraysConmed Linvatec · 2009
  3. This deviceK152969 Enhance Allograft Wedge Instrument TrayConmed Corporation · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K080531Conmed Linvatec Microfracture Instrument Sterilization TrayConmed Linvatec2008—
−1K090560Conmed Linvatec Anodized Aluminum Sterilization TraysConmed Linvatec2009—
0 (this device)K152969Enhance Allograft Wedge Instrument TrayConmed Corporation2016—
FIG. 1 — Lineage of K152969, 2008–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090560Conmed Linvatec Anodized Aluminum Sterilization Trays
Summary page 4
…Device Name: ConMed Linvatec Anodized Aluminum Sterilization Trays Company Name: ConMed Linvatec 510(k) #: K090560 IV. DEVICE DESCRIPTION The Enhance® Allograft Wedge Instrumentation Tray (T903) is constructed primarily of al…
Conmed Linvatec2009—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 2 companies; a new generation was cleared every 6.8 years on median.
  • None of the 2 ancestors has a recall.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The Enhance Allograft Wedge Instrumentation Tray (T903) is a containment device intended for medical device sterilization. The Enhance Allograft Wedge Instrumentation Tray is used for loading surgical instruments in order to conveniently organize, sterilize, transport, and store the instruments between uses. This tray is intended for use with the following instruments: REF Description REF Description 903106 Evans Trial 6mm 903206 Cotton Trial 6mm 903108…

Read the full statement (224 words)

The Enhance Allograft Wedge Instrumentation Tray (T903) is a containment device intended for medical device sterilization. The Enhance Allograft Wedge Instrumentation Tray is used for loading surgical instruments in order to conveniently organize, sterilize, transport, and store the instruments between uses. This tray is intended for use with the following instruments:

REF Description REF Description

903106 Evans Trial 6mm 903206 Cotton Trial 6mm

903108 Evans Trial 8mm 903207 Cotton Trial 7mm

903110 Evans Trial 10mm 903208 Cotton Trial 8mm

903112 Evans Trial 12mm 903006 Mallet

903010 Tamp Evans 903008 Distractor

903012 Tamp Cotton 903002 Osteotome 16mm

903205 Cotton Trial 5mm 903004 Osteotome 10mm

The Enhance Allograft Wedge Instrumentation Tray (T903) is not intended to maintain sterility; it is intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to maintain sterility of the enclosed medical instruments. Sterilize the instrument tray using the following sterilization parameters:

Method Cycle Temperature Exposure Dry Cycle Cool Down Duration

Steam Pre-vacuum 270°F (132°C) 4 minutes 20 minutes 118°F (48°C) 25 minutes (wrapped)

Steam Gravity 250°F (121°C) 30 minutes 15 minutes 118°F (48°C) 37 minutes (wrapped)

- Validation was conducted using an FDA cleared sterilization wrap with a maximum load 14 lbs. (6.4 kg).

- Only single unstacked trays may be used during the sterilization process.

- Double wrap tray using an FDA cleared wrap.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K152969: Enhance Allograft Wedge Instrument Tray.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K152969

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026