510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Turbett Surgical Container

K153025510(k) TraditionalCleared Jan 14, 201690-day reviewClass II21 CFR 880.6850Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Turbett Surgical, LLC · FRG Wrap, Sterilization

Cleared on Jan 14, 2016 as substantially equivalent to K110898 (Ammed Surgical Equipment, LLC, 2012) after a 90-day review. Its family tree reaches back 1 generation to 2012. 3 later devices descend from it. It has been recalled 4 times.

Decided
Jan 14, 2016
Review
90 days
Class
Class II
Product code
FRG
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K110898 (2012). 3 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2020: 1 descendant cleared2021: 1 descendant cleared2025: 1 descendant clearedK110898 · S.C.O.R.E.S. Unit · Ammed Surgical Equipment, LLC · 2012 · recalled2010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K110898 · S.C.O.R.E.S. Unit · Ammed Surgical Equipment, LLC · 2012 · recalled0
  1. Root · 1 generation backK110898 S.C.O.R.E.S. UnitAmmed Surgical Equipment, LLC · 2012 · Recalled
  2. This deviceK153025 Turbett Surgical ContainerTurbett Surgical, LLC · 2016
  3. Descendant · cites it directly · 2 of its ownK200240 Turbett Instrument PodTurbett Surgical, Inc. · 2020
  4. Descendant · 2 generations on · 1 of its ownK202593 Turbett Instrument PodTurbett Surgical, Inc. · 2021
  5. Descendant · 3 generations on · 0 of its ownK250011 Turbett Surgical Instrument Pod (TS1500); Turbett Surgical…Turbett Surgical · 2025

Most influential descendants

Showing 3 of 3 · See all
  • K200240Turbett Instrument PodTurbett Surgical, Inc. · 2020 · cites it directly2descendants
  • K202593Turbett Instrument PodTurbett Surgical, Inc. · 2021 · 2 generations on1descendant
  • K250011Turbett Surgical Instrument Pod (TS1500); Turbett Surgical Instrument Pod (TS1200)…Turbett Surgical · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K110898S.C.O.R.E.S. UnitAmmed Surgical Equipment, LLC20121
0 (this device)K153025Turbett Surgical ContainerTurbett Surgical, LLC2016—
+1K200240Turbett Instrument PodTurbett Surgical, Inc.2020—
+2K202593Turbett Instrument PodTurbett Surgical, Inc.2021—
+3K250011Turbett Surgical Instrument Pod (TS1500); Turbett Surgical Instrument Pod…Turbett Surgical2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202110
202510
FIG. 1 — Lineage of K153025, 2012–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K110898S.C.O.R.E.S. Unit
Summary page 6
…n factors usability study Substantial Equivalence: The Turbett Surgical Container is substantially equivalent to the SCORES Unit Sterilization Container (K110898). Both products have the same indications for use and perform in a similar manner.…
Ammed Surgical Equipment, LLC20121

Also cleared against K110898: 1 other device →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K200240Turbett Instrument PodTurbett Surgical, Inc.20202

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 5, 2022Class IIZ-1038-2023Device DesignPod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.
Dec 5, 2022Class IIZ-1037-2023Device DesignPod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.
Apr 14, 2022Class IIZ-1501-2022Process controlDevices were subject to unapproved rework processes.
Apr 1, 2022Class IIZ-1111-2022Device DesignEight units manufactured before June 2019 have exhibited weld failures breaching the sterile barrier after prolonged use.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K200240 (Turbett Surgical, Inc., 2020) has gone furthest, with 2 descendants of its own.
  • 1 other device also cites its predicate K110898.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

The Turbett Surgical Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical filters. The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of…

Read the full statement (160 words)

The Turbett Surgical Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical filters.

The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270°F (132°C) and exposure time of 4 minutes. The container may be loaded with up to 15 trays and up to 25 lbs. per tray.

Validation was done using three trays per level and a maximum instrument load of 25 Ibs. per tray. The validation load included six Imm x 500mm lumens and six 3mm x 400mm lumens. The total weight of instruments and trays validated is 375 lbs.

Use only uncovered, perforated or wire mesh general delivery trays within the Turbett Surgical Container.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K153025: Turbett Surgical Container.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K153025

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026