Penumbra Embolectomy Aspiration System (INDIGOTM Aspiration System)

K160533510(k) TraditionalCleared May 24, 201696-day reviewClass II21 CFR 870.5150

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Penumbra, Inc. · QEW Peripheral Mechanical Thrombectomy With Aspiration

Cleared on May 24, 2016 as substantially equivalent to K142870 (Penumbra, Inc., 2015) after a 96-day review. Its family tree reaches back 8 generations to 2001. 27 later devices descend from it.

Decided
May 24, 2016
Review
96 days
Class
Class II
Product code
QEW
Recalls
None
Predicates
1

Its longest line of descent runs back 8 generations to K003410 (2001). 10 ancestors in all, the oldest K003086 (2001). 27 later devices descend from it, 1 citing it directly.

Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor4 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 3 descendants5 generations forward: 6 descendants6 generations forward: 3 descendants7 generations forward: 5 descendants8 generations forward: 4 descendants K142870 · Penumbra Embolectomy Aspiration System [INDIGO Aspiration System] · Penumbra, Inc. · 2015K121917 · Penumbra Embolectomy Aspiration System · Penumbra, Inc. · 2012K113163 · Penumbra System Max · Penumbra, Inc. · 2011K100769 · Penumbra System Separator Flex (026,032,041,054), Models PSF026, PSF032, PSF041, PSF054 · Penumbra, Inc. · 2010K072718 · Penumbra System · Penumbra, Inc. · 2007 · recalledK033736 · Merci Retriever, Models 90065, 90066 · Concentric Medical, Inc. · 2004K030476 · Modified Concentric Retriever, Model 90037 · Concentric Medical, Inc. · 2003 · recalledK003410 · Concentric Retriever, Model 90030 · Concentric Medical, Inc. · 2001 · recalled−50+5
  1. Root · 8 generations backK003410 Concentric Retriever, Model 90030Concentric Medical, Inc. · 2001 · Recalled
  2. 7 generations backK030476 Modified Concentric Retriever, Model 90037Concentric Medical, Inc. · 2003 · Recalled
  3. 6 generations backK033736 Merci Retriever, Models 90065, 90066Concentric Medical, Inc. · 2004
  4. 5 generations backK072718 Penumbra SystemPenumbra, Inc. · 2007 · Recalled
  5. 4 generations backK100769 Penumbra System Separator Flex (026,032,041,054), Models PSF026…Penumbra, Inc. · 2010
  6. 3 generations backK113163 Penumbra System MaxPenumbra, Inc. · 2011
  7. 2 generations backK121917 Penumbra Embolectomy Aspiration SystemPenumbra, Inc. · 2012
  8. 1 generation backK142870 Penumbra Embolectomy Aspiration System [INDIGO Aspiration System]Penumbra, Inc. · 2015
  9. This deviceK160533 Penumbra Embolectomy Aspiration System (INDIGOTM Aspiration System)Penumbra, Inc. · 2016
  10. Descendant · cites it directly · 26 of its ownK161523 INDIGO Aspiration SystemPenumbra, Inc. · 2016
  11. Descendant · 2 generations on · 22 of its ownK180105 Indigo Aspiration System - Penumbra Engine Pump, Indigo Aspiration…Penumbra, Inc. · 2018
  12. Descendant · 3 generations on · 21 of its ownK180939 Indigo Aspiration System - Modified 110 Aspiration TubingPenumbra, Inc. · 2018
  13. Descendant · 4 generations on · 14 of its ownK193244 Indigo Aspiration System – Lightning Aspiration TubingPenumbra, Inc. · 2020
  14. Descendant · 5 generations on · 11 of its ownK210323 Indigo Aspiration System - Lightning Aspiration TubingPenumbra, Inc. · 2021

Showing 5 of 27 descendants. See all descendants

Descendants cleared per year

2016: 12018: 22020: 42021: 22022: 22023: 12024: 52025: 52026: 5
2016high 5 (2024) · low 0 (2017)2026

Most influential descendants

Showing 5 of 27 · See all
  • K161523INDIGO Aspiration SystemPenumbra, Inc. · 2016 · cites it directly26descendants
  • K180105Indigo Aspiration System - Penumbra Engine Pump, Indigo Aspiration System - Penumbra…Penumbra, Inc. · 2018 · 2 generations on22descendants
  • K180939Indigo Aspiration System - Modified 110 Aspiration TubingPenumbra, Inc. · 2018 · 3 generations on21descendants
  • K193244Indigo Aspiration System – Lightning Aspiration TubingPenumbra, Inc. · 2020 · 4 generations on14descendants
  • K210323Indigo Aspiration System - Lightning Aspiration TubingPenumbra, Inc. · 2021 · 5 generations on11descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K003410Concentric Retriever, Model 90030Concentric Medical, Inc.20011
−7K030476Modified Concentric Retriever, Model 90037Concentric Medical, Inc.20032
−6K033736Merci Retriever, Models 90065, 90066Concentric Medical, Inc.2004—
−5K072718Penumbra SystemPenumbra, Inc.20071
−4K100769Penumbra System Separator Flex (026,032,041,054), Models PSF026, PSF032…Penumbra, Inc.2010—
−3K113163Penumbra System MaxPenumbra, Inc.2011—
−2K121917Penumbra Embolectomy Aspiration SystemPenumbra, Inc.2012—
−1K142870Penumbra Embolectomy Aspiration System [INDIGO Aspiration System]Penumbra, Inc.2015—
0 (this device)K160533Penumbra Embolectomy Aspiration System (INDIGOTM Aspiration System)Penumbra, Inc.2016—
+1K161523INDIGO Aspiration SystemPenumbra, Inc.2016—
+2K180105Indigo Aspiration System - Penumbra Engine Pump, Indigo Aspiration System -…Penumbra, Inc.2018—
+3K180939Indigo Aspiration System - Modified 110 Aspiration TubingPenumbra, Inc.2018—
+4K193244Indigo Aspiration System – Lightning Aspiration TubingPenumbra, Inc.2020—
+5K210323Indigo Aspiration System - Lightning Aspiration TubingPenumbra, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201820
202040
202120
202220
202310
202450
202550
202650
FIG. 1 — Lineage of K160533, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in QEW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QEW device

  1. #1 K213402 Triever24, Triever20score 28.6 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K142870Penumbra Embolectomy Aspiration System [INDIGO Aspiration System]
Summary page 4
…Name of Name of Device Clearance Date Manufacturer Predicate Device K142870[26May2015] Penumbra Embolectomy Aspiration System Penu…
Penumbra, Inc.2015—
†K121917Penumbra Embolectomy Aspiration SystemPenumbra, Inc.2012—
†K122756Penumbra Pump Max (110VAC), Penumbra Pump Max (230VAC), Penumbar Pump Max Canister/TubingPenumbra, Inc.2012—

Also cleared against K142870: 8 other devices →

What descends from it? (27)

See all 27

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K161523INDIGO Aspiration SystemPenumbra, Inc.201626

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 2 companies; a new generation was cleared every 1.8 years on median.
  • 3 of the 10 ancestors have a recall of any class.
  • Of its descendants, K161523 (Penumbra, Inc., 2016) has gone furthest, with 26 descendants of its own.
  • 8 other devices also cite its predicate K142870.

How does its category compare?

All QEW devices

QEW Peripheral Mechanical Thrombectomy With Aspiration: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · QEW

14in 2025

2008: 22009: 22012: 12013: 22014: 22015: 42016: 42017: 22018: 142019: 82020: 162021: 112022: 122023: 232024: 142025: 142026 (partial year): 10
2007high 23 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QEW

58in 2025

2015: 88 days median2016: 63 days median2018: 61 days median2019: 51 days median2020: 105 days median2021: 60 days median2022: 140 days median2023: 102 days median2024: 65 days median2025: 58 days median2026 (partial year): 74 days median
2007high 140 (2022) · low 51 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · QEW

7in 2022

2022: 7, 7 Class I
2007high 7 (2022) · low 0 (2007)2026

Class I (7 since 2007) Class II/III not classified

Adverse event reports per year · QEW

207in 2025

2019: 332020: 1972021: 3222022: 1252023: 1652024: 1622025: 207
2007high 322 (2021) · low 0 (2007)2025

MAUDE reports naming product code QEW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—00
20092—00
20100—00
20110—00
20121—00
20132—00
20142—00
201548800
201646300
20172—00
2018146100
2019851033
2020161050197
202111600322
2022121407 (7)125
2023231020165
202414650162
202514580207
202610740—

What is it indicated for?

As part of the Penumbra Embolectomy Aspiration System (INDIGO™ Aspiration System), the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems. Not for use in the coronaries or the neurovasculature.

As part of the Penumbra Embolectomy Aspiration System (INDIGO™ Aspiration System), the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Pump MAX.

The Penumbra Pump MAX is indicated as a vacuum source for Penumbra Aspiration Systems.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160533: Penumbra Embolectomy Aspiration System (INDIGOTM Aspiration System).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K160533

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026