510(k) · Cleared
Recalled · 2003Modified Concentric Retriever, Model 90037
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Concentric Medical, Inc. · DQY Catheter, Percutaneous
Cleared on Mar 11, 2003 as substantially equivalent to K003410 (Concentric Medical, Inc., 2001) after a 26-day review. Its family tree reaches back 1 generation to 2001. 8 of the 166 devices in its family, counting itself, have a Class I recall. 164 later devices descend from it. It has been recalled 2 times.
- Decided
- Mar 11, 2003
- Review
- 26 days
- Class
- Class II
- Product code
- DQY
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 1 generation to K003410 (2001). 164 later devices descend from it, 4 citing it directly. 8 of the 166 devices in this family, counting this one, have a Class I recall.
- Root · 1 generation backK003410 Concentric Retriever, Model 90030Concentric Medical, Inc. · 2001 · Recalled
- This deviceK030476 Modified Concentric Retriever, Model 90037Concentric Medical, Inc. · 2003
- Descendant · cites it directly · 159 of its ownK033736 Merci Retriever, Models 90065, 90066Concentric Medical, Inc. · 2004
- Descendant · 2 generations on · 120 of its ownK072718 Penumbra SystemPenumbra, Inc. · 2007 · Recalled
- Descendant · 3 generations on · 118 of its ownK100769 Penumbra System Separator Flex (026,032,041,054), Models PSF026…Penumbra, Inc. · 2010
- Descendant · 3 generations on · 116 of its ownK113163 Penumbra System MaxPenumbra, Inc. · 2011
- Descendant · 3 generations on · 61 of its ownK133317 Penumbra System / Penumbra System MaxPenumbra, Inc. · 2014
Showing 5 of 164 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 164 · See all- K033736Merci Retriever, Models 90065, 90066Concentric Medical, Inc. · 2004 · cites it directly159descendants
- K072718Penumbra SystemPenumbra, Inc. · 2007 · 2 generations on120descendants
- K100769Penumbra System Separator Flex (026,032,041,054), Models PSF026, PSF032, PSF041, PSF054Penumbra, Inc. · 2010 · 3 generations on118descendants
- K113163Penumbra System MaxPenumbra, Inc. · 2011 · 3 generations on116descendants
- K133317Penumbra System / Penumbra System MaxPenumbra, Inc. · 2014 · 3 generations on61descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K003410 | Concentric Retriever, Model 90030 | Concentric Medical, Inc. | 2001 | 1 |
| 0 (this device) | K030476 | Modified Concentric Retriever, Model 90037 | Concentric Medical, Inc. | 2003 | — |
| +1 | K033736 | Merci Retriever, Models 90065, 90066 | Concentric Medical, Inc. | 2004 | — |
| +2 | K072718 | Penumbra System | Penumbra, Inc. | 2007 | 1 |
| +3 | K100769 | Penumbra System Separator Flex (026,032,041,054), Models PSF026, PSF032… | Penumbra, Inc. | 2010 | — |
| +3 | K113163 | Penumbra System Max | Penumbra, Inc. | 2011 | — |
| +3 | K133317 | Penumbra System / Penumbra System Max | Penumbra, Inc. | 2014 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 3 | 0 |
| 2005 | 1 | 0 |
| 2007 | 2 | 0 |
| 2009 | 1 | 0 |
| 2010 | 2 | 0 |
| 2011 | 2 | 0 |
| 2012 | 6 | 0 |
| 2014 | 4 | 0 |
| 2015 | 8 | 1 |
| 2016 | 8 | 0 |
| 2017 | 4 | 1 |
| 2018 | 16 | 0 |
| 2019 | 10 | 2 |
| 2020 | 14 | 2 |
| 2021 | 11 | 0 |
| 2022 | 7 | 0 |
| 2023 | 13 | 2 |
| 2024 | 12 | 0 |
| 2025 | 20 | 0 |
| 2026 | 20 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
8 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DQY device
- #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K003410 | Concentric Retriever, Model 90030 | 2001 |
What descends from it? (164)
See all 1644 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (8 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 24, 2003 | Not classified | Z-0085-04 | Other | The device has the potential for tip breakage during use. |
| Sep 24, 2003 | Not classified | Z-0086-04 | Other | The device has the potential for tip breakage during use. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K141516Mindframe Capture LP Revascularization DeviceMicro Therapeutics, Inc. d/b/a ev3 Neurovascular · 2015 · 4 generations on · ◆ in the 2015 descendants bar of FIG. 11 Class I recall
- K190779Trevo XP ProVue RetrieverStryker · 2019 · 4 generations on · ◆ in the 2019 descendants bar of FIG. 11 Class I recall
- K162901Penumbra 3D Revascularization DevicePenumbra, Inc. · 2017 · 6 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
- K223530Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing SetRoute 92 Medical, Inc. · 2023 · 6 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
- K190010Penumbra System Reperfusion Catheter JET 7Penumbra, Inc. · 2019 · 8 generations on · ◆ in the 2019 descendants bar of FIG. 14 Class I recalls
- K191946Penumbra System Penumbra JET 7XPenumbra, Inc. · 2020 · 9 generations on · ◆ in the 2020 descendants bar of FIG. 14 Class I recalls
- K202251Penumbra System (Penumbra JET 7 Reperfusion Catheter with Xtra Flex Technology; Penumbra JET 7MAX)Penumbra, Inc. · 2020 · 9 generations on · ◆ in the 2020 descendants bar of FIG. 14 Class I recalls
- K222786072 Aspiration SystemQ'Apel Medical, Inc. · 2023 · 10 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
Nearest 5 of 8; all are on the web page.
What does this lineage say?
How does its category compare?
All DQY devicesDQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DQY
21in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DQY
115in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DQY
1in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · DQY
3,056in 2025
MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 16 | 85 | 6 | 313 |
| 2008 | 31 | 66 | 12 | 496 |
| 2009 | 37 | 73 | 17 | 819 |
| 2010 | 22 | 74 | 32 | 1,293 |
| 2011 | 36 | 71 | 6 | 1,211 |
| 2012 | 30 | 87 | 2 | 1,191 |
| 2013 | 32 | 92 | 36 (1) | 1,352 |
| 2014 | 19 | 127 | 3 | 1,497 |
| 2015 | 27 | 118 | 3 | 2,337 |
| 2016 | 22 | 117 | 22 (7) | 2,252 |
| 2017 | 26 | 113 | 8 | 2,489 |
| 2018 | 25 | 106 | 2 | 2,855 |
| 2019 | 30 | 111 | 1 | 2,902 |
| 2020 | 20 | 120 | 2 | 2,889 |
| 2021 | 18 | 81 | 2 | 3,006 |
| 2022 | 18 | 74 | 0 | 2,643 |
| 2023 | 18 | 114 | 17 | 3,049 |
| 2024 | 16 | 87 | 1 | 3,190 |
| 2025 | 21 | 115 | 1 (1) | 3,056 |
| 2026 | 11 | 30 | 2 (1) | — |
What is it indicated for?
The Concentric Retriever is indicated for use in the retrieval of foreign bodies misplaced during interventional radiological procedures in the neuro, peripheral and coronary vascular systems.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K030476 · printed Nov 6, 2026
Cite this page
510(k) Report. “K030476: Modified Concentric Retriever, Model 90037.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K030476
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026