510(k) Report

510(k) · Cleared

Cleared

Instrument kits

K160730510(k) TraditionalCleared Apr 27, 2017407-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Anthogyr · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Apr 27, 2017 as substantially equivalent to K142519 (Implant Direct Sybron Manufacturing, LLC, 2015) after a 407-day review. Its family tree reaches back 7 generations to 1984. 20 later devices descend from it.

Decided
Apr 27, 2017
Review
407 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K844002 (1984). 10 ancestors in all. 20 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1997: 1 other ancestor1994: 1 other ancestor2018: 1 descendant cleared2023: 3 descendants cleared2019: 3 descendants cleared2021: 4 descendants cleared2020: 2 descendants cleared2025: 2 descendants cleared2022: 3 descendants cleared2024: 2 descendants clearedK844002 · HU-Friedy Ims Cassette System · Hu-Friedy Mfg. Co., Inc. · 1984K980065 · C/T Med-Systems Cassette System · C/T Med System, Ltd. · 1998K990761 · Tuttnauer Mini Container Cassette, Tuttnauer Standard Case Cassette · Tuttnauer U.S.A Co, Ltd. · 1999K012105 · Polyvac Surgical Instrument Delivery System · Symmetry Medical, Inc. · 2002K023658 · Sterilization Cases and Trays · Carr Metal Products, Inc. · 2003K052992 · Conmed Linvatec Instrument Sterilization Tray · Conmed Linvatec · 2006K142519 · InterActive Complete Surgical Tray · Implant Direct Sybron Manufacturing, LLC · 201519801990200020102020
  1. Root · 7 generations backK844002 HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc. · 1984
  2. 6 generations backK980065 C/T Med-Systems Cassette SystemC/T Med System, Ltd. · 1998
  3. 5 generations backK990761 Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd. · 1999
  4. 4 generations backK012105 Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc. · 2002
  5. 3 generations backK023658 Sterilization Cases and TraysCarr Metal Products, Inc. · 2003
  6. 2 generations backK052992 Conmed Linvatec Instrument Sterilization TrayConmed Linvatec · 2006
  7. 1 generation backK142519 InterActive Complete Surgical TrayImplant Direct Sybron Manufacturing, LLC · 2015
  8. This deviceK160730 Instrument kitsAnthogyr · 2017
  9. Descendant · cites it directly · 18 of its ownK171713 Neodent Instrument KitsJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2018
  10. Descendant · 2 generations on · 13 of its ownK182865 Neodent Instruments KitsJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2019
  11. Descendant · 3 generations on · 5 of its ownK192670 Neodent Instrument Kit CasesJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2019
  12. Descendant · 3 generations on · 4 of its ownK201688 S.I.N. Instrument KitsS.I.N. - Sistema DE Implante Nacional S.A. · 2020
  13. Descendant · 2 generations on · 2 of its ownK201878 ST-Z5Z-Systems AG · 2021

Showing 5 of 20 descendants. See all descendants

Descendants cleared per year

2018: 12019: 32020: 22021: 42022: 32023: 32024: 22025: 2
2018high 4 (2021) · low 0 (2026)2026

Most influential descendants

Showing 5 of 20 · See all
  • K171713Neodent Instrument KitsJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2018 · cites it directly18descendants
  • K182865Neodent Instruments KitsJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2019 · 2 generations on13descendants
  • K192670Neodent Instrument Kit CasesJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2019 · 3 generations on5descendants
  • K201688S.I.N. Instrument KitsS.I.N. - Sistema DE Implante Nacional S.A. · 2020 · 3 generations on4descendants
  • K201878ST-Z5Z-Systems AG · 2021 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K844002HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc.1984—
−6K980065C/T Med-Systems Cassette SystemC/T Med System, Ltd.1998—
−5K990761Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd.1999—
−4K012105Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc.2002—
−3K023658Sterilization Cases and TraysCarr Metal Products, Inc.2003—
−2K052992Conmed Linvatec Instrument Sterilization TrayConmed Linvatec2006—
−1K142519InterActive Complete Surgical TrayImplant Direct Sybron Manufacturing, LLC2015—
0 (this device)K160730Instrument kitsAnthogyr2017—
+1K171713Neodent Instrument KitsJjgc Industria E Comercio DE Materiais Dentarios S.A.2018—
+2K182865Neodent Instruments KitsJjgc Industria E Comercio DE Materiais Dentarios S.A.2019—
+3K192670Neodent Instrument Kit CasesJjgc Industria E Comercio DE Materiais Dentarios S.A.2019—
+3K201688S.I.N. Instrument KitsS.I.N. - Sistema DE Implante Nacional S.A.2020—
+2K201878ST-Z5Z-Systems AG2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
201930
202020
202140
202230
202330
202420
202520
FIG. 1 — Lineage of K160730, 1984–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K142519InterActive Complete Surgical Tray
Summary page 4
…RILIZATION CASSETTE Class II Product Code: KCT CFR Section: 21CFR 880.6850 Device panel: GENERAL HOSPITAL 5.3 PREDICATE DEVICE InterActive Complete Surgical Tray (K142519), manufactured by Implant Direct 5.4 DEVICE DESCRIPTION Instrument kits are cassettes used to enclose and hold surgical in…
Implant Direct Sybron Manufacturing, LLC2015—

Also cleared against K142519: 4 other devices →

What descends from it? (20)

See all 20

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171713Neodent Instrument KitsJjgc Industria E Comercio DE Materiais Dentarios S.A.201818
K230680Anthogyr Surgical CassettesAnthogyr20230

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 8 companies; a new generation was cleared every 3.1 years on median.
  • None of the 10 ancestors has a recall.
  • Of its descendants, K171713 (Jjgc Industria E Comercio DE Materiais Dentarios S.A., 2018) has gone furthest, with 18 descendants of its own.
  • 4 other devices also cite its predicate K142519.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The instrument kits are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and protect the instruments that are sterilized by healthcare provider. The cassette is to be enclosed in an FDA cleared steam sterilizable pouch. The cassettes are not intended on their own to maintain sterility.

The cycle of sterilization is:

Pre-vacuum steam: 134 °C during 3 minutes with 16 minutes drying time

Anthogyr does not make any lumen claims for the instrument kits.

The Axiom range surgery kit represents the worst case validated load due to the number of components (34 instruments).

The cassettes are not intended to be stacked during sterilization process.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160730: Instrument kits.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K160730

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026