510(k) Report

510(k) · Cleared

Cleared

Anthogyr Surgical Cassettes

K230680510(k) TraditionalCleared Jul 11, 2023120-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Anthogyr · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Jul 11, 2023 as substantially equivalent to K160730 (Anthogyr, 2017) after a 120-day review. Its family tree reaches back 8 generations to 1984.

Decided
Jul 11, 2023
Review
120 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
1

Its longest line of descent runs back 8 generations to K844002 (1984). 11 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 8 generations backK844002 HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc. · 1984
  2. 7 generations backK980065 C/T Med-Systems Cassette SystemC/T Med System, Ltd. · 1998
  3. 6 generations backK990761 Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd. · 1999
  4. 5 generations backK012105 Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc. · 2002
  5. 4 generations backK023658 Sterilization Cases and TraysCarr Metal Products, Inc. · 2003
  6. 3 generations backK052992 Conmed Linvatec Instrument Sterilization TrayConmed Linvatec · 2006
  7. 2 generations backK142519 InterActive Complete Surgical TrayImplant Direct Sybron Manufacturing, LLC · 2015
  8. 1 generation backK160730 Instrument kitsAnthogyr · 2017
  9. This deviceK230680 Anthogyr Surgical CassettesAnthogyr · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K844002HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc.1984—
−7K980065C/T Med-Systems Cassette SystemC/T Med System, Ltd.1998—
−6K990761Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd.1999—
−5K012105Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc.2002—
−4K023658Sterilization Cases and TraysCarr Metal Products, Inc.2003—
−3K052992Conmed Linvatec Instrument Sterilization TrayConmed Linvatec2006—
−2K142519InterActive Complete Surgical TrayImplant Direct Sybron Manufacturing, LLC2015—
−1K160730Instrument kitsAnthogyr2017—
0 (this device)K230680Anthogyr Surgical CassettesAnthogyr2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199710
200110
FIG. 1 — Lineage of K230680, 1984–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K160730Instrument kits
Summary page 5
…Anthogyr Surgical Cassettes 510(k) Summary Predicate Device(s) Primary Predicate: • K160730 – Instrument Kits (Anthogyr) Classification Name: Sterilization Wrap Containers, Trays, Cassettes & Other Acc…
Anthogyr2017—
†K203618Neodent EasyGuide Kit CasesJjgc Industria E Comercio DE Materiais Dentarios S.A.2021—

Also cleared against K160730: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 39 years and passes through 8 companies; a new generation was cleared every 3.6 years on median.
  • None of the 11 ancestors has a recall.
  • 1 other device also cites its predicate K160730.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The Anthogyr Surgical Cassettes are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and protect the instruments that are sterilized by healthcare provider. The cassette is to be enclosed in an FDA cleared steam sterilizable pouch. The cassettes are not intended on their own to maintain sterility. The cycle of sterilization is: INMODOPS3, INMODOPS3L, INMODOPP3, INMODIGM, INMODOPS3V, INMODOPS3LV, INMODOPP3V…

Read the full statement (138 words)

The Anthogyr Surgical Cassettes are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and protect the instruments that are sterilized by healthcare provider. The cassette is to be enclosed in an FDA cleared steam sterilizable pouch. The cassettes are not intended on their own to maintain sterility.

The cycle of sterilization is:

INMODOPS3, INMODOPS3L, INMODOPP3, INMODIGM, INMODOPS3V, INMODOPS3LV, INMODOPP3V, INMODIGMV: Pre-vacuum steam: 132 °C (270º F) during 4 minutes with 20 minutes drying time

The Anthogyr Surgical Cassettes have been validated for a maximum load of with the associated instruments:

The worst-case recommended load is 412g.

The device dimensions are listed below for Anthogyr Surgical Cassettes:

- INMODOPS3, INMODOPS3V, INMODOPS3L, INMODOPS3LV, INMODOPP3, INMODOPP3V: 130x155x47 mm

- INMODIGM, INMODIGMV: 76x155x47 mm

The cassettes are not intended to be stacked during sterilization process.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K230680: Anthogyr Surgical Cassettes.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K230680

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026