510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

ACCU-CHEK Guide Blood Glucose Monitoring System

K160944510(k) TraditionalCleared Aug 31, 2016148-day reviewClass II21 CFR 862.1345Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diabetes Care, Inc. · NBW System, Test, Blood Glucose, Over The Counter

Cleared on Aug 31, 2016 as substantially equivalent to K141867 (Roche Diagnostics, 2015) and 1 other predicate after a 148-day review. Its family tree reaches back 6 generations to 2001. 3 later devices descend from it. It has been recalled 2 times.

Decided
Aug 31, 2016
Review
148 days
Class
Class II
Product code
NBW
Recalls
2
Predicates
2

Its longest line of descent runs back 6 generations to K010362 (2001). 7 ancestors in all. 3 later devices descend from it, 3 citing it directly.

  1. Root · 6 generations backK010362 Accu-Chek Comfort Curve Test StripRoche Diagnostics Corp. · 2001 · Recalled
  2. 5 generations backK043474 Accu-Chek Aviva SystemRoche Diagnostics Corp. · 2005 · Recalled
  3. 4 generations backK060620 Accu-Chek Aviva Test StripsRoche Diagnostics Corp. · 2006 · Recalled
  4. 3 generations backK101299 Accu-Chek Aviva Plus Blood Glucose Monitoring SystemRoche Diagnostics · 2011
  5. 2 generations backK133862 Accu-Chek Aviva Plus SystemRoche Diagnostics Corporation · 2014 · Recalled
  6. 1 generation backK141867 Accu-Chek Aviva Connect Blood Glucose Monitoring SystemRoche Diagnostics · 2015
  7. This deviceK160944 ACCU-CHEK Guide Blood Glucose Monitoring SystemRoche Diabetes Care, Inc. · 2016
  8. Descendant · cites it directly · 0 of its ownK213131 Accu-Chek Guide Solo diabetes manager blood glucose monitoring systemRoche Diabetes Care GmbH · 2023
  9. Descendant · cites it directly · 0 of its ownK203711 IWL2020 Blood Glucose Monitoring SystemRoche Diabetes Care, Inc. · 2022
  10. Descendant · cites it directly · 0 of its ownK181131 Accu-Chek Guide Me Blood Glucose Monitoring SystemRoche Diabetes Care · 2018

Most influential descendants

Showing 3 of 3 · See all
  • K213131Accu-Chek Guide Solo diabetes manager blood glucose monitoring systemRoche Diabetes Care GmbH · 2023 · cites it directly0descendants
  • K203711IWL2020 Blood Glucose Monitoring SystemRoche Diabetes Care, Inc. · 2022 · cites it directly0descendants
  • K181131Accu-Chek Guide Me Blood Glucose Monitoring SystemRoche Diabetes Care · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K010362Accu-Chek Comfort Curve Test StripRoche Diagnostics Corp.20016
−5K043474Accu-Chek Aviva SystemRoche Diagnostics Corp.200510
−4K060620Accu-Chek Aviva Test StripsRoche Diagnostics Corp.20061
−3K101299Accu-Chek Aviva Plus Blood Glucose Monitoring SystemRoche Diagnostics2011—
−2K133862Accu-Chek Aviva Plus SystemRoche Diagnostics Corporation20142
−1K141867Accu-Chek Aviva Connect Blood Glucose Monitoring SystemRoche Diagnostics2015—
0 (this device)K160944ACCU-CHEK Guide Blood Glucose Monitoring SystemRoche Diabetes Care, Inc.2016—
+1K213131Accu-Chek Guide Solo diabetes manager blood glucose monitoring systemRoche Diabetes Care GmbH2023—
+1K203711IWL2020 Blood Glucose Monitoring SystemRoche Diabetes Care, Inc.2022—
+1K181131Accu-Chek Guide Me Blood Glucose Monitoring SystemRoche Diabetes Care2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202210
202310
FIG. 1 — Lineage of K160944, 2001–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NBW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NBW device

  1. #1 K260545 CONTOUR® PLUS BLUE Blood Glucose Monitoring…score 14.6 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141867Accu-Chek Aviva Connect Blood Glucose Monitoring System
Summary page 5
…nassayed) 1 510(k) Summary 3. Predicate Device Accu-Chek Aviva Connect Blood Glucose Monitoring System (k141867), concurrence received on March 3, 2015. The Accu-Chek Aviva Controls (k043474), concurrence received on April 27, 2005. The Acc…
Roche Diagnostics2015—
*K043474Accu-Chek Aviva System
Summary page 5
…= Page 5 === 3. Predicate Device Accu-Chek Aviva Connect Blood Glucose Monitoring System (k141867), concurrence received on March 3, 2015. The Accu-Chek Aviva Controls (k043474), concurrence received on April 27, 2005. The Accu-Chek Aviva Connect Blood Glucose Monitoring System is intended for the quanti…
Roche Diagnostics Corp.200510

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K213131Accu-Chek Guide Solo diabetes manager blood glucose monitoring systemRoche Diabetes Care GmbH20230
K203711IWL2020 Blood Glucose Monitoring SystemRoche Diabetes Care, Inc.20220
K181131Accu-Chek Guide Me Blood Glucose Monitoring SystemRoche Diabetes Care20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 13, 2024Class IIZ-2174-2024Reprocessing ControlsThe reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times…
Sep 30, 2019Class IIZ-0747-2020Device DesignPotential power issues with the Accu-Chek Guide Blood Glucose Monitoring System.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 2.6 years on median.
  • 4 of the 7 ancestors have a recall of any class.
  • Of its descendants, K213131 (Roche Diabetes Care GmbH, 2023) has gone furthest, with 0 descendants of its own.

How does its category compare?

All NBW devices

NBW System, Test, Blood Glucose, Over The Counter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · NBW

2in 2025

2007: 292008: 372009: 382010: 312011: 302012: 302013: 472014: 242015: 322016: 202017: 252018: 192019: 102020: 72021: 72022: 122023: 52024: 72025: 22026 (partial year): 5
2007high 47 (2013) · low 2 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NBW

243in 2024

2007: 85 days median2008: 128 days median2009: 140 days median2010: 132 days median2011: 328 days median2012: 230 days median2013: 167 days median2014: 202 days median2015: 237 days median2016: 233 days median2017: 244 days median2018: 168 days median2019: 127 days median2020: 121 days median2021: 465 days median2022: 410 days median2023: 193 days median2024: 243 days median2026 (partial year): 190 days median
2007high 465 (2021) · low 85 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NBW

1in 2025

2007: 92008: 22009: 52010: 112011: 12012: 12013: 5, 2 Class I2014: 12016: 12019: 142025: 1
2007high 14 (2019) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · NBW

4,745in 2025

2007: 7,9912008: 7,7042009: 10,9632010: 11,6312011: 22,9092012: 23,4492013: 35,4082014: 38,5242015: 62,1112016: 76,6212017: 37,5382018: 16,9112019: 14,4722020: 10,7032021: 9,1982022: 8,1032023: 7,2562024: 6,2202025: 4,745
2007high 76,621 (2016) · low 4,745 (2025)2025

MAUDE reports naming product code NBW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007298597,991
20083712827,704
200938140510,963
2010311321111,631
201130328122,909
201230230123,449
2013471675 (2)35,408
201424202138,524
201532237062,111
201620233176,621
201725244037,538
201819168016,911
2019101271414,472
20207121010,703
2021746509,198
20221241008,103
2023519307,256
2024724306,220
20252—14,745
202651900—

What is it indicated for?

The Accu-Chek Guide Blood Glucose Monitoring System is comprised of the Accu-Chek Guide meter and the Accu-Chek Guide test strips. The Accu-Chek Guide Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. The Accu-Chek Guide Blood Glucose Monitoring System is intended for in vitro diagnostic…

Read the full statement (162 words)

The Accu-Chek Guide Blood Glucose Monitoring System is comprised of the Accu-Chek Guide meter and the Accu-Chek Guide test strips.

The Accu-Chek Guide Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control.

The Accu-Chek Guide Blood Glucose Monitoring System is intended for in vitro diagnostic single-patient use by people with diabetes.

The Accu-Chek Guide Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

This system is not for use in diagnosis or screening of diabetes mellitus, nor for neonatal use.

Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

Accu-Chek Guide Control Solutions are for use with the Accu-Chek Guide Blood Glucose Monitoring System to check that the meters and test strips are working together properly and that the test is performing correctly.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K160944: ACCU-CHEK Guide Blood Glucose Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K160944

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026