510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

BD Pen Needle

K162516510(k) TraditionalCleared Dec 9, 201691-day reviewClass II21 CFR 880.5570Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Becton, Dickinson and Company · FMI Needle, Hypodermic, Single Lumen

Cleared on Dec 9, 2016 as substantially equivalent to K051899 (Becton, Dickinson & CO, 2005) and 1 other predicate after a 91-day review. Its family tree reaches back 4 generations to 2005. 10 later devices descend from it. It has been recalled 2 times.

Decided
Dec 9, 2016
Review
91 days
Class
Class II
Product code
FMI
Recalls
2
Predicates
2

Its longest line of descent runs back 4 generations to K051899 (2005). 10 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2021: 3 descendants cleared2018: 2 descendants cleared2022: 2 descendants cleared2020: 1 descendant cleared2023: 1 descendant cleared2026: 1 descendant clearedReference device: K110703 · BD Autoshield Duo Pen Needle · Becton, Dickinson and Company · 2011K051899 · BD Pen Needle · Becton, Dickinson & CO · 2005K100005 · BD Pen Needle · Becton, Dickinson and Co. · 2010K123300 · BD Pen Needle · Becton, Dickinson & CO · 2013K131358 · BD Pen Needle · Becton, Dickinson & CO · 201320052010201520202025
  1. Root · 4 generations backK051899 BD Pen NeedleBecton, Dickinson & CO · 2005
  2. 3 generations backK100005 BD Pen NeedleBecton, Dickinson and Co. · 2010
  3. 2 generations backK123300 BD Pen NeedleBecton, Dickinson & CO · 2013
  4. 1 generation backK131358 BD Pen NeedleBecton, Dickinson & CO · 2013
  5. This deviceK162516 BD Pen NeedleBecton, Dickinson and Company · 2016
  6. Descendant · cites it directly · 3 of its ownK171982 Droplet Pen NeedleHtl-Strefa S.A · 2018
  7. Descendant · cites it directly · 2 of its ownK182320 BD Contoured Base Pen NeedleBecton, Dickinson and Company · 2018
  8. Descendant · cites it directly · 1 of its ownK213407 UltiCare Disposable Pen NeedlesUltimed Incorporated · 2021
  9. Descendant · 2 generations on · 0 of its ownK260739 Profoject™ Pen Needles (Model A; Model B)CMT Health PTE., Ltd. · 2026
  10. Descendant · 2 generations on · 0 of its ownK230635 Pen NeedleJiangsu Caina Medical Co.,Ltd · 2023

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2018: 22020: 12021: 32022: 22023: 12026: 1
2018high 3 (2021) · low 0 (2019)2026

Most influential descendants

Showing 5 of 10 · See all
  • K171982Droplet Pen NeedleHtl-Strefa S.A · 2018 · cites it directly3descendants
  • K182320BD Contoured Base Pen NeedleBecton, Dickinson and Company · 2018 · cites it directly2descendants
  • K213407UltiCare Disposable Pen NeedlesUltimed Incorporated · 2021 · cites it directly1descendant
  • K260739Profoject™ Pen Needles (Model A; Model B)CMT Health PTE., Ltd. · 2026 · 2 generations on0descendants
  • K230635Pen NeedleJiangsu Caina Medical Co.,Ltd · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K051899BD Pen NeedleBecton, Dickinson & CO2005—
−3K100005BD Pen NeedleBecton, Dickinson and Co.2010—
−2K123300BD Pen NeedleBecton, Dickinson & CO2013—
−1K131358BD Pen NeedleBecton, Dickinson & CO2013—
0 (this device)K162516BD Pen NeedleBecton, Dickinson and Company2016—
+1K171982Droplet Pen NeedleHtl-Strefa S.A2018—
+1K182320BD Contoured Base Pen NeedleBecton, Dickinson and Company2018—
+1K213407UltiCare Disposable Pen NeedlesUltimed Incorporated2021—
+2K260739Profoject™ Pen Needles (Model A; Model B)CMT Health PTE., Ltd.2026—
+2K230635Pen NeedleJiangsu Caina Medical Co.,Ltd2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201820
202010
202130
202220
202310
202610
FIG. 1 — Lineage of K162516, 2005–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FMI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMI device

  1. #1 K252886 Pen Needlescore 14.5 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051899BD Pen Needle
Summary page 4
…Product Code: FMI (hypodermic single lumen needle) Legally marketed predicate devices to which substantial equivalence is being claimed: K051899 and K131358: BD Pen Needle Reference Device: K110703: BD AutoShield™ Duo Pen Needle Device Description: The BD Pen Needle is des…
Becton, Dickinson & CO2005—
*K131358BD Pen Needle
Summary page 4
…Product Code: FMI (hypodermic single lumen needle) Legally marketed predicate devices to which substantial equivalence is being claimed: K051899 and K131358: BD Pen Needle Reference Device: K110703: BD AutoShield™ Duo Pen Needle Device Description: The BD Pen Needle is designed for us…
Becton, Dickinson & CO2013—
†K110703BD Autoshield Duo Pen NeedleBecton, Dickinson and Company2011—

Also cleared against K051899: 13 other devices →

What descends from it? (10)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171982Droplet Pen NeedleHtl-Strefa S.A20183
K182320BD Contoured Base Pen NeedleBecton, Dickinson and Company20182
K213407UltiCare Disposable Pen NeedlesUltimed Incorporated20211
K213478BD Pen NeedleBecton, Dickinson and Company20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 16, 2021Class IIZ-1933-2021Environmental controlDue to shipping delays the devices were exposed to temperature excursions of 31.9¿F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.
Oct 23, 2020Class IIZ-0508-2021Under Investigation by firmProduct shelf cartons incorrectly labeled as products intended for the Latin American market. Although the product description on the label 32G x 4mm pen needles is correct, information pertinent to users in the US regarding compatibility…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 3.4 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K171982 (Htl-Strefa S.A, 2018) has gone furthest, with 3 descendants of its own.
  • 13 other devices also cite its predicate K051899.

How does its category compare?

All FMI devices

FMI Needle, Hypodermic, Single Lumen: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · FMI

11in 2025

2007: 152008: 102009: 72010: 152011: 122012: 172013: 112014: 142015: 102016: 202017: 92018: 202019: 122020: 112021: 182022: 232023: 182024: 32025: 112026 (partial year): 12
2007high 23 (2022) · low 3 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMI

141in 2025

2007: 71 days median2008: 85 days median2009: 177 days median2010: 119 days median2011: 113 days median2012: 107 days median2013: 87 days median2014: 155 days median2015: 89 days median2016: 217 days median2017: 253 days median2018: 176 days median2019: 186 days median2020: 263 days median2021: 143 days median2022: 189 days median2023: 112 days median2024: 90 days median2025: 141 days median2026 (partial year): 119 days median
2007high 263 (2020) · low 71 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMI

1in 2024

2007: 22008: 42009: 32010: 792011: 52012: 12013: 12014: 32015: 12017: 162019: 52020: 32021: 72023: 22024: 12026 (partial year): 1, 1 Class I
2007high 79 (2010) · low 0 (2016)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · FMI

2,424in 2025

2007: 2082008: 1772009: 2632010: 2492011: 2402012: 2152013: 2552014: 2872015: 5672016: 9412017: 1,5472018: 2,3982019: 4,8962020: 4,2452021: 4,5322022: 4,8852023: 4,9862024: 4,0532025: 2,424
2007high 4,986 (2023) · low 177 (2008)2025

MAUDE reports naming product code FMI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715712208
200810854177
200971773263
20101511979249
2011121135240
2012171071215
201311871255
2014141553287
201510891567
2016202170941
20179253161,547
20182017602,398
20191218654,896
20201126334,245
20211814374,532
20222318904,885
20231811224,986
202439014,053
20251114102,424
2026121191 (1)—

What is it indicated for?

BD Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K162516: BD Pen Needle.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162516

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026