510(k) Report

510(k) · Cleared

Recalled · Class I · 2020

AC3 Series IABP System

K162820510(k) TraditionalCleared Mar 31, 2017175-day reviewClass II21 CFR 870.3535Recalls 2 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Arrow International, Inc. · DSP System, Balloon, Intra-Aortic And Control

Cleared on Mar 31, 2017 as substantially equivalent to K060309 (Arrow Intl., Inc., 2006) after a 175-day review. Its family tree reaches back 4 generations to 1991. 7 of the 20 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 2 times, including 2 Class I recalls.

Decided
Mar 31, 2017
Review
175 days
Class
Class II
Product code
DSP
Recalls
2 (2 Class I)
Predicates
1

Its longest line of descent runs back 4 generations to K905313 (1991). 15 ancestors in all, the oldest K892799 (1989). 4 later devices descend from it, 1 citing it directly. 7 of the 20 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor, 1 with a Class I recall12000: 3 other ancestors, 2 with a Class I recall21998: 1 other ancestor, 1 with a Class I recall11997: 1 other ancestor1996: 2 other ancestors1995: 1 other ancestor1992: 1 other ancestor1989: 1 other ancestor2019: 1 descendant cleared2020: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared2025: 1 descendant clearedK905313 · Kontron Intra-Aortic Balloon Pump, Model Kaat II · Kontron Instruments, Inc. · 1991K965209 · Acat 1 (Arrow Cardiac Assist Technology 1) · Arrow Intl., Inc. · 1998 · recalledK002256 · Arrow Acat 2 Intra-Aortic Balloon Pump · Arrow Intl., Inc. · 2001K060309 · Autocat Intra-Aortic Balloon Pump Series · Arrow Intl., Inc. · 2006 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors, 1 with a Class I recall13 generations back: 7 other ancestors, 3 with a Class I recall34 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant, 1 with a Class I recall13 generations forward: 1 descendant4 generations forward: 1 descendant K060309 · Autocat Intra-Aortic Balloon Pump Series · Arrow Intl., Inc. · 2006 · Class I recallK002256 · Arrow Acat 2 Intra-Aortic Balloon Pump · Arrow Intl., Inc. · 2001K965209 · Acat 1 (Arrow Cardiac Assist Technology 1) · Arrow Intl., Inc. · 1998 · recalledK905313 · Kontron Intra-Aortic Balloon Pump, Model Kaat II · Kontron Instruments, Inc. · 19910
  1. Root · 4 generations backK905313 Kontron Intra-Aortic Balloon Pump, Model Kaat IIKontron Instruments, Inc. · 1991
  2. 3 generations backK965209 Acat 1 (Arrow Cardiac Assist Technology 1)Arrow Intl., Inc. · 1998 · Recalled
  3. 2 generations backK002256 Arrow Acat 2 Intra-Aortic Balloon PumpArrow Intl., Inc. · 2001
  4. 1 generation backK060309 Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc. · 2006 · Recalled (Class I)
  5. This deviceK162820 AC3 Series IABP SystemArrow International, Inc. · 2017
  6. Descendant · cites it directly · 3 of its ownK192238 AC3 Series Intra-Aortic Balloon Pump (IABP)Arrow International, Teleflex · 2019
  7. Descendant · 2 generations on · 2 of its ownK201112 AC3 IABP, AC3 Optimus IABP, AutoCAT2 Wave IABP, AutoCAT2 IABPArrow International, Inc. · 2020 · Recalled (Class I)
  8. Descendant · 3 generations on · 1 of its ownK232343 AC3™ Series IABPArrow International, LLC · 2023
  9. Descendant · 4 generations on · 0 of its ownK250542 AC3™ Range™ Intra-Aortic Balloon PumpArrow International, LLC · 2025

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K192238AC3 Series Intra-Aortic Balloon Pump (IABP)Arrow International, Teleflex · 2019 · cites it directly3descendants
  • K201112AC3 IABP, AC3 Optimus IABP, AutoCAT2 Wave IABP, AutoCAT2 IABPArrow International, Inc. · 2020 · 2 generations on2descendants
  • K232343AC3™ Series IABPArrow International, LLC · 2023 · 3 generations on1descendant
  • K250542AC3™ Range™ Intra-Aortic Balloon PumpArrow International, LLC · 2025 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K905313Kontron Intra-Aortic Balloon Pump, Model Kaat IIKontron Instruments, Inc.1991—
−3K965209Acat 1 (Arrow Cardiac Assist Technology 1)Arrow Intl., Inc.19981
−2K002256Arrow Acat 2 Intra-Aortic Balloon PumpArrow Intl., Inc.2001—
−1K060309Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc.2006Class I
0 (this device)K162820AC3 Series IABP SystemArrow International, Inc.2017—
+1K192238AC3 Series Intra-Aortic Balloon Pump (IABP)Arrow International, Teleflex2019—
+2K201112AC3 IABP, AC3 Optimus IABP, AutoCAT2 Wave IABP, AutoCAT2 IABPArrow International, Inc.2020Class I
+3K232343AC3™ Series IABPArrow International, LLC2023—
+4K250542AC3™ Range™ Intra-Aortic Balloon PumpArrow International, LLC2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199210
199510
199620
199710
199811
200032
200211

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202011
202310
202510
FIG. 1 — Lineage of K162820, 1991–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSP device

  1. #1 K250542 AC3™ Range™ Intra-Aortic Balloon Pumpscore 11.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060309Autocat Intra-Aortic Balloon Pump Series
Summary page 4
…ubject of this submission, is substantially equivalent to the Legally previously cleared Arrow AutoCAT®2 IABP System which was Marketed previously cleared via K060309. Devices Page 1 of 11…
Arrow Intl., Inc.2006Class I

Also cleared against K060309: 2 other devices →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K192238AC3 Series Intra-Aortic Balloon Pump (IABP)Arrow International, Teleflex20193

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 20, 2020Class IZ-2416-2020Component design/selectionA potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of…
May 20, 2020Class IZ-2417-2020Component design/selectionA potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K060309Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc. · 2006 · Direct predicate · ◆ on the line of descent in FIG. 128 Class I recalls
  • K021462Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement System, Models…Arrow Intl., Inc. · 2002 · 2 generations back · ◆ in the 2002 ancestors bar of FIG. 12 Class I recalls
  • K000729Arrow Ultra 8 Intra-Aortic Balloon Catheters (Iabs) 8FR 30CC and 40CC UniversalArrow Intl., Inc. · 2000 · 3 generations back · ◆ in the 2000 ancestors bar of FIG. 17 Class I recalls
  • K9939668FR. Narrowflex Universal Intra-Aortic Balloon CatheterArrow Intl., Inc. · 2000 · 3 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K981660Bard 8 Fr., 40cc. & 7Fr., 30cc Intra-Aortic Balloon CathetersC.R. Bard, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 12 Class I recalls
  • K201112AC3 IABP, AC3 Optimus IABP, AutoCAT2 Wave IABP, AutoCAT2 IABPArrow International, Inc. · 2020 · 2 generations on · ◆ in the 2020 descendants bar of FIG. 111 Class I recalls

What does this lineage say?

  • The line of descent spans 26 years and passes through 3 companies; a new generation was cleared every 6.6 years on median.
  • 6 of the 15 ancestors have a recall of any class; 5 have a Class I recall.
  • Of its descendants, K192238 (Arrow International, Teleflex, 2019) has gone furthest, with 3 descendants of its own.
  • 2 other devices also cite its predicate K060309.

How does its category compare?

All DSP devices

DSP System, Balloon, Intra-Aortic And Control: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · DSP

1in 2025

2007: 22009: 22011: 22012: 22013: 12015: 12017: 32018: 12019: 32020: 22023: 12025: 1
2007high 3 (2017) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSP

140in 2019

2017: 73 days median2019: 140 days median
2007high 140 (2019) · low 73 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSP

4in 2025

2007: 32009: 142010: 82012: 32014: 12016: 9, 9 Class I2018: 2, 2 Class I2019: 32021: 5, 3 Class I2022: 33, 33 Class I2024: 12025: 4
2007high 33 (2022) · low 0 (2008)2026

Class I (47 since 2007) Class II/III not classified

Adverse event reports per year · DSP

7,422in 2025

2007: 4612008: 6172009: 6562010: 1,0192011: 5902012: 4902013: 4952014: 1,7382015: 4042016: 4192017: 2,1642018: 3,3312019: 3,5522020: 3,3932021: 4,3452022: 5,1132023: 6,9802024: 7,3172025: 7,422
2007high 7,422 (2025) · low 404 (2015)2025

MAUDE reports naming product code DSP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—3461
20080—0617
20092—14656
20100—81,019
20112—0590
20122—3490
20131—0495
20140—11,738
20151—0404
20160—9 (9)419
201737302,164
20181—2 (2)3,331
2019314033,552
20202—03,393
20210—5 (3)4,345
20220—33 (33)5,113
20231—06,980
20240—17,317
20251—47,422
20260—0—

What is it indicated for?

The AC3™ Series IABP (Intra-Aortic Balloon Pump) System is clinically indicated for the following conditions:

a. Acute Coronary Syndrome

b. Cardiac and Non-Cardiac Surgery

c. Complications of Heart Failure

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K162820: AC3 Series IABP System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162820

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.