510(k) Report

510(k) · Cleared

Cleared

RX Takeru Balloon Dilatation Catheter

K163372510(k) TraditionalCleared Apr 7, 2017127-day reviewClass II21 CFR 870.5100

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

KANEKA Corporation · LOX Catheters, Transluminal Coronary Angioplasty, Percutaneous

Cleared on Apr 7, 2017 as substantially equivalent to P860019/S208 (Scimed Life Systems, Inc., 2008) and 1 other predicate after a 127-day review. Its family tree reaches back 2 generations to 1980. 1 later device descends from it.

Decided
Apr 7, 2017
Review
127 days
Class
Class II
Product code
LOX
Recalls
None
Predicates
2

Its longest line of descent runs back 2 generations to P790017 (1980). 3 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2018: 1 descendant clearedP790017 · Usci Dilatation Catheter · Medtronic, Ireland · 1980K143480 · Euphora Rapid Exchange Balloon Dilatation Catheter · Medtronic, Inc. · 2015 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant K143480 · Euphora Rapid Exchange Balloon Dilatation Catheter · Medtronic, Inc. · 2015 · recalledP790017 · Usci Dilatation Catheter · Medtronic, Ireland · 19800
  1. Root · 2 generations backP790017 Usci Dilatation CatheterMedtronic, Ireland · 1980
  2. 1 generation backK143480 Euphora Rapid Exchange Balloon Dilatation CatheterMedtronic, Inc. · 2015 · Recalled
  3. This deviceK163372 RX Takeru Balloon Dilatation CatheterKANEKA Corporation · 2017
  4. Descendant · cites it directly · 0 of its ownK172264 OTW Takeru PTCA Balloon Dilatation CatheterKANEKA Corporation · 2018

Most influential descendants

Showing 1 of 1 · See all
  • K172264OTW Takeru PTCA Balloon Dilatation CatheterKANEKA Corporation · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2P790017Usci Dilatation CatheterMedtronic, Ireland1980—
−1K143480Euphora Rapid Exchange Balloon Dilatation CatheterMedtronic, Inc.20153
0 (this device)K163372RX Takeru Balloon Dilatation CatheterKANEKA Corporation2017—
+1K172264OTW Takeru PTCA Balloon Dilatation CatheterKANEKA Corporation2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
FIG. 1 — Lineage of K163372, 1980–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LOX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LOX device

  1. #1 K254022 FLASH Flex™ Aorto-Ostial Angioplasty Systemscore 12.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*P860019/S208Apex Monorail and Over-the-Wire Ptca Dilatation Catheters
Summary page 5
…II Classification Panel Cardiovascular (74) Predicate Devices:  Primary predicate device: ApexTM Monorail PTCA Dilatation Catheter or “Apex” [P860019/S208 (Boston Scientific Corporation)]  EuphoraTM Rapid Exchange Balloon Dilatation Catheter or “Euphora” [K143480 (Medtronic…
Scimed Life Systems, Inc.2008—
*K143480Euphora Rapid Exchange Balloon Dilatation Catheter
Summary page 5
…M Monorail PTCA Dilatation Catheter or “Apex” [P860019/S208 (Boston Scientific Corporation)]  EuphoraTM Rapid Exchange Balloon Dilatation Catheter or “Euphora” [K143480 (Medtronic Inc.)] Device Description: Rx Takeru PTCA Balloon Dilatation Catheter (Rx Takeru) is a rapid exchange type of b…
Medtronic, Inc.20153

Also cleared against P860019/S208: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172264OTW Takeru PTCA Balloon Dilatation CatheterKANEKA Corporation20180

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 37 years and passes through 3 companies; a new generation was cleared every 35.0 years on median.
  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K172264 (KANEKA Corporation, 2018) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate P860019/S208.

How does its category compare?

All LOX devices

LOX Catheters, Transluminal Coronary Angioplasty, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · LOX

4in 2024

2007: 192008: 222009: 282010: 322011: 92012: 102013: 62014: 62015: 42016: 92017: 62018: 72019: 12020: 12021: 62022: 52023: 42024: 42026 (partial year): 6
2007high 32 (2010) · low 0 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LOX

241in 2024

2011: 103 days median2012: 120 days median2013: 123 days median2014: 166 days median2015: 119 days median2016: 116 days median2017: 113 days median2018: 80 days median2021: 167 days median2022: 121 days median2023: 138 days median2024: 241 days median2026 (partial year): 156 days median
2007high 241 (2024) · low 80 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LOX

4in 2019

2007: 12008: 802019: 4
2007high 80 (2008) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LOX

1,890in 2025

2007: 1,0532008: 1,0632009: 1,3532010: 1,3922011: 2,0902012: 2,0922013: 2,4362014: 2,3162015: 2,0602016: 2,5782017: 2,7532018: 2,2252019: 2,1022020: 1,9742021: 1,7382022: 1,8162023: 1,7202024: 1,8042025: 1,890
2007high 2,753 (2017) · low 1,053 (2007)2025

MAUDE reports naming product code LOX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200719—11,053
200822—801,063
200928—01,353
201032—01,392
2011910302,090
20121012002,092
2013612302,436
2014616602,316
2015411902,060
2016911602,578
2017611302,753
201878002,225
20191—42,102
20201—01,974
2021616701,738
2022512101,816
2023413801,720
2024424101,804
20250—01,890
202661560—

What is it indicated for?

The RX Takeru PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion in the

coronary artery or bypass graft stenosis for the purpose of myocardial perfusion. This product (balloon models 2.0-5.0

mm) is also indicated for the post-delivery expansion of balloon expandable stents.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K163372: RX Takeru Balloon Dilatation Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K163372

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026