510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

CT 5300

K232491510(k) TraditionalCleared May 3, 2024260-day reviewClass II21 CFR 892.1750Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Healthcare (Suzhou) Co., Ltd. · JAK System, X-Ray, Tomography, Computed

Cleared on May 3, 2024 as substantially equivalent to K212441 (Philips Healthcare (Suzhou) Co., Ltd., 2022) after a 260-day review. Its family tree reaches back 8 generations to 1993. 1 later device descends from it. It has been recalled 5 times.

Decided
May 3, 2024
Review
260 days
Class
Class II
Product code
JAK
Recalls
5
Predicates
1

Its longest line of descent runs back 8 generations to K930090 (1993). 13 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 8 generations backK930090 HelicatElscint, Inc. · 1993
  2. 7 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 6 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 5 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 4 generations backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. 3 generations backK160743 Philips Ingenuity CTPhilips Medical Systems (Cleveland), Inc. · 2016 · Recalled
  7. 2 generations backK180015 Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · 2018 · Recalled
  8. 1 generation backK212441 Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · 2022 · Recalled
  9. This deviceK232491 CT 5300Philips Healthcare (Suzhou) Co., Ltd. · 2024
  10. Descendant · cites it directly · 0 of its ownK252992 CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd. · 2026

Most influential descendants

Showing 1 of 1 · See all
  • K252992CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd. · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K930090HelicatElscint, Inc.1993—
−7K961464Select/SPElscint, Inc.19963
−6K982060Volumax CT Imaging SystemElscint, Inc.1998—
−5K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−4K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
−3K160743Philips Ingenuity CTPhilips Medical Systems (Cleveland), Inc.201624
−2K180015Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd.201823
−1K212441Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd.202217
0 (this device)K232491CT 5300Philips Healthcare (Suzhou) Co., Ltd.2024—
+1K252992CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K232491, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K212441Philips Incisive CT
Summary page 4
…r: Philips Healthcare (Suzhou) Co., Ltd. 510(k) Clearance: K212441 Classification Regulation: 21CFR 892.1750 Classification name:…
Philips Healthcare (Suzhou) Co., Ltd.202217

Also cleared against K212441: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K252992CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd.20260

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 29, 2025Class IIZ-0882-2026Under Investigation by firmIf the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other…
Oct 29, 2025Class IIZ-0881-2026Under Investigation by firmIf the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other…
Sep 25, 2025Class IIZ-0375-2026Under Investigation by firmIssue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation)…
Sep 25, 2025Class IIZ-0376-2026Under Investigation by firmIssue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation)…
Sep 25, 2025Class IIZ-0377-2026Under Investigation by firmIssue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation)…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 5 companies; a new generation was cleared every 3.0 years on median.
  • 7 of the 13 ancestors have a recall of any class.
  • Of its descendants, K252992 (Philips Healthcare (Suzhou) Co., Ltd., 2026) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K212441.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The CT 5300 is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The CT 5300 is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. These…

Read the full statement (152 words)

The CT 5300 is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The CT 5300 is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K232491: CT 5300.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K232491

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.