510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

IBA Proton Therapy System- Proteus 235 (Proteus One, Proteus Plus, Proteus TK2, Proteus Class)

K163500510(k) TraditionalCleared Aug 17, 2017247-day reviewClass II21 CFR 892.5050Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ion Beam Applications S.A. · LHN System, Radiation Therapy, Charged-Particle, Medical

Cleared on Aug 17, 2017 as substantially equivalent to K152224 (Ion Beam Application S.A., 2016) after a 247-day review. Its family tree reaches back 7 generations to 1988. It has been recalled 11 times.

Decided
Aug 17, 2017
Review
247 days
Class
Class II
Product code
LHN
Recalls
11
Predicates
1

Its longest line of descent runs back 7 generations to K872369 (1988). 15 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 7 generations backK872369 Loma Linda Univ. Proton Beam Therapy SystemLoma Linda University Medical Center · 1988
  2. 6 generations backK983024 Proton Therapy SystemIon Beam Applications S.A. · 2001
  3. 5 generations backK053641 Iba Proton Therapy System-Proteus 235Ion Beam Applications S.A. · 2006
  4. 4 generations backK060695 Iba Proton Therapy System - Proteus 235Ion Beam Applications S.A. · 2006 · Recalled
  5. 3 generations backK062891 Proton Therapy SystemIndiana University Cyclotron Facility · 2006
  6. 2 generations backK120676 S-250 Proton Beam Radiation Therapy SystemMevion Medical Systems · 2012 · Recalled
  7. 1 generation backK152224 Proton Therapy System - Proteus 235 (Proteus One, Proteus Plus…Ion Beam Application S.A. · 2016 · Recalled
  8. This deviceK163500 IBA Proton Therapy System- Proteus 235 (Proteus One, Proteus Plus…Ion Beam Applications S.A. · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K872369Loma Linda Univ. Proton Beam Therapy SystemLoma Linda University Medical Center1988—
−6K983024Proton Therapy SystemIon Beam Applications S.A.2001—
−5K053641Iba Proton Therapy System-Proteus 235Ion Beam Applications S.A.2006—
−4K060695Iba Proton Therapy System - Proteus 235Ion Beam Applications S.A.20062
−3K062891Proton Therapy SystemIndiana University Cyclotron Facility2006—
−2K120676S-250 Proton Beam Radiation Therapy SystemMevion Medical Systems201212
−1K152224Proton Therapy System - Proteus 235 (Proteus One, Proteus Plus, Proteus TK2)Ion Beam Application S.A.20167
0 (this device)K163500IBA Proton Therapy System- Proteus 235 (Proteus One, Proteus Plus, Proteus…Ion Beam Applications S.A.2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200110
200620
200810
201020
201410
FIG. 1 — Lineage of K163500, 1988–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LHN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LHN device

  1. #1 K260597 Virtue QAscore 15.4 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K152224Proton Therapy System - Proteus 235 (Proteus One, Proteus Plus, Proteus TK2)
Summary page 5
…– Proteus 235 (PTS) with the addition of the IBA wireless hand- pendant System is substantially equivalent to the previously cleared IBA Proton Therapy System (K152224). The current PTS and its Primary predicate device have the same intended use and principles of operation, and are substanti…
Ion Beam Application S.A.20167

Also cleared against K152224: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 3, 2026Class IIZ-1497-2026Software designIt was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external…
Jul 10, 2024Class IIZ-2693-2024Process designTemporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues
Jul 8, 2024Class IIZ-3089-2024Under Investigation by firmSafety Parameters Verification Mechanisms can be deactivated in clinical runtime
May 10, 2024Class IIZ-1945-2024Device DesignIrradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
Mar 4, 2024Class IIZ-1603-2024Device DesignThere is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance.
Sep 18, 2023Class IIZ-0295-2024Device DesignThe patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
Apr 20, 2022Class IIZ-1205-2022Under Investigation by firmFirst layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed
Feb 15, 2022Class IIZ-0747-2022Device DesignWhen resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range…
Nov 29, 2021Class IIZ-0466-2022Device DesignProton Therapy System (PTS) software can be started in the clinical environment (clinical user) while some processes are still running in the test environment (tcs user). Initiating the startup of the clinical processes in the environment…
Jul 17, 2020Class IIZ-2715-2020Software designIBA became aware of an issue to resume aborted irradiations because the plan to complete the treatment cannot be loaded.
Apr 21, 2020Class IIZ-2773-2020Software designSignature from the user is necessary to proceed with specific actions in the Proton Therapy System (PTS). IBA became aware that the PTS does not accept user names with more than ten characters. It is an issue when the user has no other…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 29 years and passes through 5 companies; a new generation was cleared every 4.2 years on median.
  • 7 of the 15 ancestors have a recall of any class.
  • 1 other device also cites its predicate K152224.

How does its category compare?

All LHN devices

LHN System, Radiation Therapy, Charged-Particle, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · LHN

3in 2025

2007: 22008: 32009: 52010: 42011: 12012: 32013: 32014: 42015: 32016: 72017: 42018: 32019: 22020: 42022: 22023: 22024: 12025: 32026 (partial year): 4
2007high 7 (2016) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LHN

190in 2025

2008: 113 days median2009: 69 days median2010: 96 days median2012: 111 days median2013: 70 days median2014: 93 days median2015: 53 days median2016: 120 days median2017: 173 days median2018: 78 days median2020: 74 days median2025: 190 days median2026 (partial year): 194 days median
2007high 194 (2026) · low 53 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LHN

1in 2023

2007: 22008: 22014: 22015: 12017: 52018: 22020: 12021: 12022: 22023: 1
2007high 5 (2017) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LHN

11in 2025

2008: 12009: 52010: 32011: 12012: 12013: 12014: 52015: 22016: 12017: 72018: 82019: 32020: 202021: 242022: 102023: 72024: 22025: 11
2007high 24 (2021) · low 0 (2007)2025

MAUDE reports naming product code LHN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—20
2008311321
200956905
201049603
20111—01
2012311101
201337001
201449325
201535312
2016712001
2017417357
201837828
20192—03
2020474120
20210—124
20222—210
20232—17
20241—02
20253190011
202641940—

What is it indicated for?

The IBA Proton Therapy System (PTS) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K163500: IBA Proton Therapy System- Proteus 235 (Proteus One, Proteus Plus, Proteus TK2, Proteus Class).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K163500

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026