K132847510(k) TraditionalCleared Nov 20, 201370-day reviewClass II21 CFR 892.5050Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ion Beam Applications S.A. · LHN System, Radiation Therapy, Charged-Particle, Medical

Cleared on Nov 20, 2013 as substantially equivalent to K080742 (Medcom GmbH, 2008) and 3 other predicates after a 70-day review. Its family tree reaches back 6 generations to 1981. It has been recalled 3 times.

Decided
Nov 20, 2013
Review
70 days
Class
Class II
Product code
LHN
Recalls
3
Predicates
4

Its longest line of descent runs back 6 generations to K803130 (1981). 31 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2007: 2 other ancestors2004: 3 other ancestors2003: 2 other ancestors2001: 2 other ancestors1995: 3 other ancestors2000: 1 other ancestor1998: 3 other ancestors1996: 1 other ancestor1999: 1 other ancestor1997: 1 other ancestor1994: 2 other ancestors1992: 1 other ancestor1987: 1 other ancestor1986: 1 other ancestor1988: 1 other ancestorK803130 · Multix U. Bucky Table X-Ray Unit · Siemens Corp. · 1981 · recalledK971452 · Multix Top/Pro Radiographic X-Ray Table · Siemens Medical Solutions USA, Inc. · 1997 · recalledK983732 · Thorax FD and Multix FD · Siemens Medical Solutions USA, Inc. · 1999 · recalledK992794 · Stingray DR, Model 1 · Infimed, Inc. · 1999K024147 · Paxscan 4030 Medical Digital Imaging System · Varian Medical Systems, Inc. · 2003K080742 · Verisuite 1.6, Verisuite-Particle 1.6 · Medcom GmbH · 200819801985199019952000200520102015
  1. Root · 6 generations backK803130 Multix U. Bucky Table X-Ray UnitSiemens Corp. · 1981 · Recalled
  2. 5 generations backK971452 Multix Top/Pro Radiographic X-Ray TableSiemens Medical Solutions USA, Inc. · 1997 · Recalled
  3. 4 generations backK983732 Thorax FD and Multix FDSiemens Medical Solutions USA, Inc. · 1999 · Recalled
  4. 3 generations backK992794 Stingray DR, Model 1Infimed, Inc. · 1999
  5. 2 generations backK024147 Paxscan 4030 Medical Digital Imaging SystemVarian Medical Systems, Inc. · 2003
  6. 1 generation backK080742 Verisuite 1.6, Verisuite-Particle 1.6Medcom GmbH · 2008
  7. This deviceK132847 I2CIon Beam Applications S.A. · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K803130Multix U. Bucky Table X-Ray UnitSiemens Corp.19811
−5K971452Multix Top/Pro Radiographic X-Ray TableSiemens Medical Solutions USA, Inc.19978
−4K983732Thorax FD and Multix FDSiemens Medical Solutions USA, Inc.19991
−3K992794Stingray DR, Model 1Infimed, Inc.1999—
−2K024147Paxscan 4030 Medical Digital Imaging SystemVarian Medical Systems, Inc.2003—
−1K080742Verisuite 1.6, Verisuite-Particle 1.6Medcom GmbH2008—
0 (this device)K132847I2CIon Beam Applications S.A.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198710
198810
199210
199420
199530
199610
199710
199830
199910
200010
200120
200320
200430
200720
FIG. 1 — Lineage of K132847, 1981–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LHN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LHN device

  1. #1 K260597 Virtue QAscore 15.4 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080742Verisuite 1.6, Verisuite-Particle 1.6
Summary page 2
…verification and correction of the patient's position during a Verisuite radiotherapy treatment with external beams or charged particles. It is (K080742, Medcom) based on stereoscopic X-ray images and DRRs calculated from a CT image series of the treatme…
Medcom GmbH2008—
*K072046Exactrac 3RD Party, Exactrac X-Ray 6D 3RD Party, Exactrac Oem
Summary page 2
…linear accelerator for analyzing the current patient position and ExacTrac 3rd Party calculating - when applicable - a necessary correction shift. The (K072046, Brainlab) correction shift is then exported to the MHI-TM2000 linear accelerator. The…
Brainlab AG2007—
*K040192ON-Board Imager Device
Summary page 2
…ect the patient position before Iand during treatment. 0BI The On-Board Imager device is used for verification of correct patient (K040192 & position in relation to isocenter and verification of the treatment fields K042720, Varian) in relation to an…
Varian Medical Systems, Inc.20042
*K042720ON-Board Imager Device
Summary page 2
…The On-Board Imager device is used for verification of correct patient (K040192 & position in relation to isocenter and verification of the treatment fields K042720, Varian) in relation to anatomical and/or fidlucial landmarks Indication for Use of the device 120 is used with a charg…
Varian Medical Systems20045

Also cleared against K080742: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 28, 2016Class IIZ-2109-2018Software change controlIBA is initiating this recall to address an issue identified with AdaPTinsight software and to reduce risk related to this problem.
Sep 16, 2016Class IIZ-2138-2018Finished device change controlIBA is initiating this recall to notify its customers that the use of non-IBA accessories and/or third party equipment or unauthorized modifications to an IBA medical device may pose a safety risk.
Nov 7, 2014Class IIZ-0836-2015Software design (manufacturing process)Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 32 years and passes through 6 companies; a new generation was cleared every 5.2 years on median.
  • 15 of the 31 ancestors have a recall of any class.
  • 1 other device also cites its predicate K080742.

How does its category compare?

All LHN devices

LHN System, Radiation Therapy, Charged-Particle, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · LHN

3in 2025

2007: 22008: 32009: 52010: 42011: 12012: 32013: 32014: 42015: 32016: 72017: 42018: 32019: 22020: 42022: 22023: 22024: 12025: 32026 (partial year): 4
2007high 7 (2016) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LHN

190in 2025

2008: 113 days median2009: 69 days median2010: 96 days median2012: 111 days median2013: 70 days median2014: 93 days median2015: 53 days median2016: 120 days median2017: 173 days median2018: 78 days median2020: 74 days median2025: 190 days median2026 (partial year): 194 days median
2007high 194 (2026) · low 53 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LHN

1in 2023

2007: 22008: 22014: 22015: 12017: 52018: 22020: 12021: 12022: 22023: 1
2007high 5 (2017) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LHN

11in 2025

2008: 12009: 52010: 32011: 12012: 12013: 12014: 52015: 22016: 12017: 72018: 82019: 32020: 202021: 242022: 102023: 72024: 22025: 11
2007high 24 (2021) · low 0 (2007)2025

MAUDE reports naming product code LHN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—20
2008311321
200956905
201049603
20111—01
2012311101
201337001
201449325
201535312
2016712001
2017417357
201837828
20192—03
2020474120
20210—124
20222—210
20232—17
20241—02
20253190011
202641940—

What is it indicated for?

12C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K132847: I2C.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K132847

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.