510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Diana Medication Transfer Set

K170110510(k) TraditionalCleared Aug 10, 2018575-day reviewClass II21 CFR 880.5440Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icu Medical, Inc. · ONB Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System

Cleared on Aug 10, 2018 as substantially equivalent to K050339 (Laboratorios Grifols, S.A., 2005) after a 575-day review. Its family tree reaches back 5 generations to 1991. 1 later device descends from it. It has been recalled once.

Decided
Aug 10, 2018
Review
575 days
Class
Class II
Product code
ONB
Recalls
1
Predicates
1

Its longest line of descent runs back 5 generations to K903493 (1991). 7 ancestors in all, the oldest K894827 (1989). 1 later device descends from it, 1 citing it directly.

  1. Root · 5 generations backK903493 Namic Contrast Savings Delivery SystemNorth American Instrument Corp. · 1991
  2. 4 generations backK961794 Fluid Administration Set - Contrast ManagerMerit Medical Systems, Inc. · 1996
  3. 3 generations backK022431 Medrad Transfer SetMedrad, Inc. · 2002
  4. 2 generations backK040456 Fleboset MultipleLaboratorios Grifols, S.A. · 2004
  5. 1 generation backK050339 Set Gri-Fill 3.0-1 Way, 2 Ways, MultipleLaboratorios Grifols, S.A. · 2005
  6. This deviceK170110 Diana Medication Transfer SetIcu Medical, Inc. · 2018
  7. Descendant · cites it directly · 0 of its ownK190157 Diana ChemoClave Transfer SetIcu Medical · 2019

Most influential descendants

Showing 1 of 1 · See all
  • K190157Diana ChemoClave Transfer SetIcu Medical · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K903493Namic Contrast Savings Delivery SystemNorth American Instrument Corp.1991—
−4K961794Fluid Administration Set - Contrast ManagerMerit Medical Systems, Inc.1996—
−3K022431Medrad Transfer SetMedrad, Inc.2002—
−2K040456Fleboset MultipleLaboratorios Grifols, S.A.2004—
−1K050339Set Gri-Fill 3.0-1 Way, 2 Ways, MultipleLaboratorios Grifols, S.A.2005—
0 (this device)K170110Diana Medication Transfer SetIcu Medical, Inc.2018—
+1K190157Diana ChemoClave Transfer SetIcu Medical2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
FIG. 1 — Lineage of K170110, 1991–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in ONB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new ONB device

  1. #1 K240171 ProSeal™ Injector Plus (Model no. 421050)score 18.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K050339Set Gri-Fill 3.0-1 Way, 2 Ways, Multiple
Summary page 4
…us Drug Reconstitution And Transfer System Regulatory Class II Product Code: ONB III. PREDICATE DEVICE Predicate – K050339 SETS GRI-FILL Reference – K131549 ChemoLock (utilized for vapor containment characteristics/methods) IV. DEVICE DESCRIPTION Diana…
Laboratorios Grifols, S.A.2005—
†K131549Chemolock, Chemolock PortIcu Medical, Inc.2013Class I

Also cleared against K050339: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K190157Diana ChemoClave Transfer SetIcu Medical20190

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 12, 2026Class IIZ-3161-2026Process controlIntravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 5 companies; a new generation was cleared every 5.4 years on median.
  • 1 of the 7 ancestors has a recall of any class.
  • Of its descendants, K190157 (Icu Medical, 2019) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K050339.

How does its category compare?

All ONB devices

ONB Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · ONB

6in 2025

2013: 32014: 22015: 32017: 22018: 62019: 12020: 32021: 32022: 12023: 32024: 82025: 6
2007high 8 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ONB

60in 2025

2013: 58 days median2015: 235 days median2018: 249 days median2020: 213 days median2021: 277 days median2023: 175 days median2024: 102 days median2025: 60 days median
2007high 277 (2021) · low 58 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · ONB

38in 2025

2016: 12019: 1, 1 Class I2020: 42022: 12023: 12024: 12025: 382026 (partial year): 3
2007high 38 (2025) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · ONB

934in 2025

2012: 12013: 22014: 32015: 42016: 132017: 662018: 2282019: 4642020: 4422021: 5482022: 4482023: 4562024: 6862025: 934
2007high 934 (2025) · low 0 (2007)2025

MAUDE reports naming product code ONB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—01
201335802
20142—03
2015323504
20160—113
20172—066
201862490228
20191—1 (1)464
202032134442
202132770548
20221—1448
202331751456
202481021686
202566038934
20260—3—

What is it indicated for?

The Diana ChemoLock Transfer Set is a sterile, single-use closed system transfer device used for drug preparation to transfer drug from a drug vial to an IV bag for intravenous drug administration. It prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170110: Diana Medication Transfer Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170110

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026