510(k) Report

510(k) · Cleared

Cleared

Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs; Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs

K170397510(k) TraditionalCleared Jul 10, 2017151-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Kossan International Sdn Bhd · LZA Polymer Patient Examination Glove

Cleared on Jul 10, 2017 after a 151-day review. FDA’s file for it is a 510(k) statement, which names no predicate.

Decided
Jul 10, 2017
Review
151 days
Class
Class I
Product code
LZA
Recalls
None
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K170397Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZA device

  1. #1 K192333 Blue Nitrile Examination Gloves Powder Freescore 71.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All LZA devices

LZA Polymer Patient Examination Glove: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · LZA

24in 2025

2007: 212008: 182009: 282010: 312011: 262012: 332013: 202014: 272015: 202016: 302017: 442018: 222019: 252020: 302021: 1392022: 1462023: 622024: 302025: 242026 (partial year): 15
2007high 146 (2022) · low 18 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZA

110in 2025

2007: 95 days median2008: 97 days median2009: 112 days median2010: 97 days median2011: 122 days median2012: 110 days median2013: 130 days median2014: 197 days median2015: 112 days median2016: 171 days median2017: 124 days median2018: 178 days median2019: 212 days median2020: 101 days median2021: 111 days median2022: 111 days median2023: 92 days median2024: 89 days median2025: 110 days median2026 (partial year): 103 days median
2007high 212 (2019) · low 89 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZA

3so far in 2026, the most since 2007

2015: 12016: 12026 (partial year): 3
2007high 3 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZA

73in 2025

2007: 72008: 22009: 122010: 122011: 92012: 112013: 102014: 82015: 122016: 252017: 42018: 32019: 212020: 42021: 202022: 192023: 242024: 762025: 73
2007high 76 (2024) · low 2 (2008)2025

MAUDE reports naming product code LZA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007219507
2008189702
200928112012
20103197012
20112612209
201233110011
201320130010
20142719708
201520112112
201630171125
20174412404
20182217803
201925212021
20203010104
2021139111020
2022146111019
20236292024
20243089076
202524110073
2026151033—

What is it indicated for?

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Chemotherapy Drug and Concentration Minimum Breakthrough Detection Time in Minutes Carmustine (BCNU) (3.3…

Read the full statement (292 words)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and Concentration Minimum Breakthrough Detection Time in Minutes

Carmustine (BCNU) (3.3 mg/ml) 10.2

Cisplatin (1.0 mg/ml) >240

Cyclophosphamide (Cytoxan) (20.0 mg/ml) >240

Cytarabine(100 mg/ml) >240

Dacarbazine (DTIC) (10.0 mg/ml) >240

Doxorubicin Hydrochloride (2.0 mg/ml) >240

Etoposide (20.0 mg/ml) >240

Fluorouracil (50.0 mg/ml) >240

Ifosfamide (50.0 mg/mi) >240

Methotrexate (25.0 mg/mi) >240

Mitomycin C (0.5 mg/ml) >240

Mitoxantrone (2.0 mg/mt) >240

Paclitaxe] (Taxol) (6.0 mg/ml) >240

Thiotepa (10.0 mg/ml) 30.4

Vincristine Sulfate (1.0 mg/ml) >240

Please note that Carmustine (BCNU) and Thiotepa has extremely low permeation times of 10.2 minutes and 30.4 minutes.

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and Concentration Minimum Breakthrough Detection Time in Minutes

Carmustine (BCNU) (3.3 mg/ml) 10.1

Cisplatin (1.0 mg/ml) >240

Cyclophosphamide (Cytoxan) (20.0 mg/ml) >240

Cytarabine(100 mg/ml) >240

Dacarbazine (DTIC) (10.0 mg/ml) >240

Doxorubicin Hydrochloride (2.0 mg/ml) >240

Etoposide (20.0 mg/ml) >240

Fluorouracil (50.0 mg/ml) >240

Ifosfamide (50.0 mg/ml) >240

Methotrexate (25.0 mg/ml) >240

Mitomycin C (0.5 mg/ml) >240

Mitoxantrone (2.0 mg/ml) >240

Paclitaxel (Taxol) (6.0 mg/m!) >240

Thiotepa (10.0 mg/ml) 304

Vincristine Sulfate (1.0 mg/mi) >240

Please note that Carmustine (BCNU) and Thiotepa has extremely low permeation times of 10,1 minutes and 30.4 minutes,

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K170397: Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs; Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K170397

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026