510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

SureFlo EVD Catheter

K170599510(k) TraditionalCleared Aug 31, 2017183-day reviewClass II21 CFR 882.5550Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Arkis Biosciences, Inc. · JXG Shunt, Central Nervous System And Components

Cleared on Aug 31, 2017 as substantially equivalent to K972994 (Integra Neurocare, LLC, 1997) and 1 other predicate after a 183-day review. Its family tree reaches back 2 generations to 1986. It has been recalled once.

Decided
Aug 31, 2017
Review
183 days
Class
Class II
Product code
JXG
Recalls
1
Predicates
2

Its longest line of descent runs back 2 generations to K864676 (1986). 6 ancestors in all, the oldest K800168 (1980). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2008: 1 other ancestor1995: 1 other ancestor1985: 1 other ancestor1980: 1 other ancestorReference device: K121089 · Nmi Picc Iii · Navilyst Medical, Inc. · 2012K864676 · Heyer-Schulte Hermetic External CSF Drain. System · American V. Mueller · 1986K972994 · External CSF Drainage Management Systems · Integra Neurocare, LLC · 19971980199020002010
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors K972994 · External CSF Drainage Management Systems · Integra Neurocare, LLC · 1997K864676 · Heyer-Schulte Hermetic External CSF Drain. System · American V. Mueller · 19860
  1. Root · 2 generations backK864676 Heyer-Schulte Hermetic External CSF Drain. SystemAmerican V. Mueller · 1986
  2. 1 generation backK972994 External CSF Drainage Management SystemsIntegra Neurocare, LLC · 1997
  3. This deviceK170599 SureFlo EVD CatheterArkis Biosciences, Inc. · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K864676Heyer-Schulte Hermetic External CSF Drain. SystemAmerican V. Mueller1986—
−1K972994External CSF Drainage Management SystemsIntegra Neurocare, LLC1997—
0 (this device)K170599SureFlo EVD CatheterArkis Biosciences, Inc.2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198510
199510
200810
FIG. 1 — Lineage of K170599, 1986–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JXG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K972994External CSF Drainage Management Systems
Summary page 4
…duct Code JXG Classification Name Shunt, Central Nervous System and Components Predicate Device(s): Primary Predicate: TraumaCath, Integra NeuroSiciences (K972994) Predicate: Neurocath Ag, Vygon Neuro, (K081942) Reference Device: NMI…
Integra Neurocare, LLC1997—
*K081942Neurocath AG
Summary page 4
…ponents Predicate Device(s): Primary Predicate: TraumaCath, Integra NeuroSiciences (K972994) Predicate: Neurocath Ag, Vygon Neuro, (K081942) Reference Device: NMI PICC III, Navilyst Medical (K121089) Performance Performance standar…
Vygon Neuro2008—
†K121089Nmi Picc IiiNavilyst Medical, Inc.2012—

Also cleared against K972994: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 10, 2026Class IIZ-2173-2026Process controlOut-of-specification endotoxin result that did not meet the acceptance criteria.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 3 companies; a new generation was cleared every 19.8 years on median.
  • None of the 6 ancestors has a recall.
  • 2 other devices also cite its predicate K972994.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—

What is it indicated for?

The SureFlo EVD Catheter is indicated for temporary insertion into a ventricular cavity of the brain for external drainage of cerebrospinal fluid (CSF) in those patients with elevated intracranial pressure (ICP), intraventricular hemorrhage, or hydrocephalic shunt infections.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170599: SureFlo EVD Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170599

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026