510(k) · Cleared
Recalled · Class II · 2022Etest Ceftolozane/ Tazobactam (0.016-256 ug/ml)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
bioMerieux, Inc. · JWY Manual Antimicrobial Susceptibility Test Systems
Cleared on May 31, 2017 as substantially equivalent to K151873 (Biomerieux S.A., 2015) after an 86-day review. Its family tree reaches back 3 generations to 2010. 3 later devices descend from it. It has been recalled once.
- Decided
- May 31, 2017
- Review
- 86 days
- Class
- Class II
- Product code
- JWY
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 3 generations to K102668 (2010). 3 later devices descend from it, 1 citing it directly.
- Root · 3 generations backK102668 Etest Tobramycin 0.016-256 UG/ML and 0.064-1024 UG/MLBiomerieux · 2010 · Recalled
- 2 generations backK121002 Etest CeftarolinebioMerieux, Inc. · 2012 · Recalled
- 1 generation backK151873 Etest Ceftaroline (0.002 - 32ug/ml)Biomerieux S.A. · 2015
- This deviceK170670 Etest Ceftolozane/ Tazobactam (0.016-256 ug/ml)bioMerieux, Inc. · 2017
- Descendant · cites it directly · 2 of its ownK172150 Etest Ceftazidime/Avibactam (CZA) (0.016 - 256 ug/mL)bioMerieux, Inc. · 2017 · Recalled
- Descendant · 2 generations on · 0 of its ownK250856 ETEST Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL)Biomérieux · 2025
- Descendant · 2 generations on · 0 of its ownK190154 ETEST Piperacillin/Tazobactam (P/T) (0.016/4-256/4 ug/mL)Biomerieux S.A. · 2019
Most influential descendants
Showing 3 of 3 · See all- K172150Etest Ceftazidime/Avibactam (CZA) (0.016 - 256 ug/mL)bioMerieux, Inc. · 2017 · cites it directly2descendants
- K250856ETEST Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL)Biomérieux · 2025 · 2 generations on0descendants
- K190154ETEST Piperacillin/Tazobactam (P/T) (0.016/4-256/4 ug/mL)Biomerieux S.A. · 2019 · 2 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K102668 | Etest Tobramycin 0.016-256 UG/ML and 0.064-1024 UG/ML | Biomerieux | 2010 | 2 |
| −2 | K121002 | Etest Ceftaroline | bioMerieux, Inc. | 2012 | 5 |
| −1 | K151873 | Etest Ceftaroline (0.002 - 32ug/ml) | Biomerieux S.A. | 2015 | — |
| 0 (this device) | K170670 | Etest Ceftolozane/ Tazobactam (0.016-256 ug/ml) | bioMerieux, Inc. | 2017 | — |
| +1 | K172150 | Etest Ceftazidime/Avibactam (CZA) (0.016 - 256 ug/mL) | bioMerieux, Inc. | 2017 | 1 |
| +2 | K250856 | ETEST Aztreonam/Avibactam (AZA) (0.016/4-256/4 µg/mL) | Biomérieux | 2025 | — |
| +2 | K190154 | ETEST Piperacillin/Tazobactam (P/T) (0.016/4-256/4 ug/mL) | Biomerieux S.A. | 2019 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2017 | 1 | 0 |
| 2019 | 1 | 0 |
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JWY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JWY device
- #1 K260447 ETEST Gepotidacin (GEP) (0.016-256 µg/mL)score 12.7 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K151873 | Etest Ceftaroline (0.002 - 32ug/ml)Summary page 5…Product Code: JWY Common Name: Etest® Ceftolozane / Tazobactam C. Predicate Device: Etest® Ceftaroline (0.002 - 32ug/ml) (K151873) D. Device Description: Etest® is a thin, inert and non-porous plastic strip carrying on one side (A) the MIC reading scale in… | 2015 |
What descends from it? (3)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K172150 | Etest Ceftazidime/Avibactam (CZA) (0.016 - 256 ug/mL) | 2017 | 2 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 9, 2022 | Class II | Z-0841-2023 | Storage | As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 1.9 years on median.
- 2 of the 3 ancestors have a recall of any class.
- Of its descendants, K172150 (bioMerieux, Inc., 2017) has gone furthest, with 2 descendants of its own.
- 1 other device also cites its predicate K151873.
How does its category compare?
All JWY devicesJWY Manual Antimicrobial Susceptibility Test Systems: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.
FDA decisions per year · JWY
14in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JWY
106in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JWY
31in 2022
Class I (13 since 2007) Class II/III not classified
Adverse event reports per year · JWY
1in 2024
MAUDE reports naming product code JWY, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 6 | 85 | 1 | 1 |
| 2009 | 2 | — | 0 | 1 |
| 2010 | 4 | 73 | 0 | 0 |
| 2011 | 4 | 125 | 3 | 0 |
| 2012 | 1 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 6 | 106 | 5 | 4 |
| 2015 | 9 | 90 | 6 (4) | 11 |
| 2016 | 4 | 88 | 0 | 20 |
| 2017 | 14 | 61 | 2 | 9 |
| 2018 | 12 | 51 | 0 | 13 |
| 2019 | 18 | 60 | 0 | 10 |
| 2020 | 5 | 66 | 0 | 20 |
| 2021 | 5 | 266 | 0 | 41 |
| 2022 | 0 | — | 31 (9) | 11 |
| 2023 | 7 | 85 | 0 | 9 |
| 2024 | 6 | 90 | 0 | 1 |
| 2025 | 14 | 106 | 0 | 0 |
| 2026 | 2 | — | 0 | — |
What is it indicated for?
Etest® is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram negative and Gram positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory…
Read the full statement (148 words)
Etest® is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram
negative and Gram positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and
Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and
Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the
Minimum Inhibitory Concentration (MIC), in μg/mL, of different antimicrobial agents against microorganisms as
tested on agar media using overnight incubation.
Etest® Ceftolozane / Tazobactam has been shown to be active against the Gram negative aerobic microorganisms
listed below, according to the FDA label for this antimicrobial agent:
Active both in vitro and in clinical infections:
Enterobacter cloacae
Escherichia coli
Klebsiella oxytoca
Klebsiella pneumoniae
Proteus mirabilis
Pseudomonas aeruginosa
The following in vitro data are available, but clinical significance is unknown:
Citrobacter koseri
Morganella morganii
Proteus vulgaris
Providencia rettgeri
Providencia stuartii
Serratia liquefaciens
Serratia marcescens
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K170670 · printed Nov 6, 2026
Cite this page
510(k) Report. “K170670: Etest Ceftolozane/ Tazobactam (0.016-256 ug/ml).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170670
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026