510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Mako Partial Knee Application

K170891510(k) TraditionalCleared May 24, 201758-day reviewClass II21 CFR 882.4560Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Mako Surgical Corp. · OLO Orthopedic Stereotaxic Instrument

Cleared on May 24, 2017 as substantially equivalent to K142530 (Mako Surgical Corporation, 2015) after a 58-day review. Its family tree reaches back 12 generations to 1991. 2 later devices descend from it. It has been recalled 3 times.

Decided
May 24, 2017
Review
58 days
Class
Class II
Product code
OLO
Recalls
3
Predicates
1

Its longest line of descent runs back 12 generations to K905070 (1991). 41 ancestors in all, the oldest K861692 (1986). 2 later devices descend from it, 1 citing it directly.

  1. Root · 12 generations backK905070 VistarPhilips Medical Systems (Cleveland), Inc. · 1991
  2. 8 generations between
  3. 3 generations backK112507 Restoris Partial Knee Application (Pka), Rio Standard SystemMako Surgical Corporation · 2012 · Recalled
  4. 2 generations backK143752 Total Knee Application (TKA)Mako Surgical Corp. · 2015 · Recalled
  5. 1 generation backK142530 Partial Knee Application (PKA)Mako Surgical Corporation · 2015
  6. This deviceK170891 Mako Partial Knee ApplicationMako Surgical Corp. · 2017
  7. Descendant · cites it directly · 1 of its ownK172301 Mako Partial Knee ApplicationMako Surgical Corp. · 2017 · Recalled
  8. Descendant · 2 generations on · 0 of its ownK242373 Mako Shoulder Application (1.0)Mako Surgical Corp. · 2024
Show all 12 generations
  1. Root · 12 generations backK905070 VistarPhilips Medical Systems (Cleveland), Inc. · 1991
  2. 11 generations backK961168 ViewpointPhilips Medical Systems (Cleveland), Inc. · 1996
  3. 10 generations backK963221 Optical Digitizer Option for View PointPhilips Medical Systems (Cleveland), Inc. · 1997
  4. 9 generations backK970604 Viewpoint - 3.0 SoftwarePhilips Medical Systems (Cleveland), Inc. · 1997
  5. 8 generations backK984298 Z Kat Flurotactic Guidance System MkiZ-Kat, Inc. · 1999
  6. 7 generations backK013025 Modification to Flourolab PlusZ-Kat, Inc. · 2001
  7. 6 generations backK031337 Acumen Surgical Navigation SystemBiomet, Inc. · 2003
  8. 5 generations backK072806 Mako Surgical Tactile Guidance System (TGS)Mako Surgical Corp. · 2008 · Recalled
  9. 4 generations backK081867 Mako Surgical Tactile Guidance System Version 2.0Mako Surgical Corp. · 2008 · Recalled
  10. 3 generations backK112507 Restoris Partial Knee Application (Pka), Rio Standard SystemMako Surgical Corporation · 2012 · Recalled
  11. 2 generations backK143752 Total Knee Application (TKA)Mako Surgical Corp. · 2015 · Recalled
  12. 1 generation backK142530 Partial Knee Application (PKA)Mako Surgical Corporation · 2015
  13. This deviceK170891 Mako Partial Knee ApplicationMako Surgical Corp. · 2017

Most influential descendants

Showing 2 of 2 · See all
  • K172301Mako Partial Knee ApplicationMako Surgical Corp. · 2017 · cites it directly1descendant
  • K242373Mako Shoulder Application (1.0)Mako Surgical Corp. · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K905070VistarPhilips Medical Systems (Cleveland), Inc.1991—
−11K961168ViewpointPhilips Medical Systems (Cleveland), Inc.1996—
−10K963221Optical Digitizer Option for View PointPhilips Medical Systems (Cleveland), Inc.1997—
−9K970604Viewpoint - 3.0 SoftwarePhilips Medical Systems (Cleveland), Inc.1997—
−8K984298Z Kat Flurotactic Guidance System MkiZ-Kat, Inc.1999—
−7K013025Modification to Flourolab PlusZ-Kat, Inc.2001—
−6K031337Acumen Surgical Navigation SystemBiomet, Inc.2003—
−5K072806Mako Surgical Tactile Guidance System (TGS)Mako Surgical Corp.20081
−4K081867Mako Surgical Tactile Guidance System Version 2.0Mako Surgical Corp.20082
−3K112507Restoris Partial Knee Application (Pka), Rio Standard SystemMako Surgical Corporation20126
−2K143752Total Knee Application (TKA)Mako Surgical Corp.20152
−1K142530Partial Knee Application (PKA)Mako Surgical Corporation2015—
0 (this device)K170891Mako Partial Knee ApplicationMako Surgical Corp.2017—
+1K172301Mako Partial Knee ApplicationMako Surgical Corp.20176
+2K242373Mako Shoulder Application (1.0)Mako Surgical Corp.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199410
199510
199620
199710
199920
200030
200130
200240
200310
200420
200540
200630
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202410
FIG. 1 — Lineage of K170891, 1991–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OLO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OLO device

  1. #1 K240662 Spine Guidance Software (version 5.0); Xia 3/…score 23.7 · 15 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K142530Partial Knee Application (PKA)
Summary page 4
…OLO Substantial Equivalence Claimed To: The Mako Partial Knee Application is substantially equivalent to Mako Surgical’s Partial Knee Application cleared via K142530. Device Modifications: The following changes have been made to the Mako Partial Knee Application: • Labeling - Implant comp…
Mako Surgical Corporation2015—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172301Mako Partial Knee ApplicationMako Surgical Corp.20171

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 10, 2018Class IIZ-2794-2018Nonconforming Material/ComponentOut of tolerance.
Dec 27, 2017Class IIZ-3182-2018Nonconforming Material/ComponentCommunication-connection error.
Dec 1, 2017Class IIZ-1644-2018Error in labelingIncorrect product and/or label. Sizing is labeled incorrectly.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 5 companies; a new generation was cleared every 2.1 years on median.
  • 10 of the 41 ancestors have a recall of any class.
  • Of its descendants, K172301 (Mako Surgical Corp., 2017) has gone furthest, with 1 descendant of its own.

How does its category compare?

All OLO devices

OLO Orthopedic Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · OLO

40in 2025

2009: 112010: 92011: 112012: 72013: 72014: 102015: 102016: 142017: 332018: 242019: 352020: 342021: 282022: 402023: 452024: 442025: 402026 (partial year): 42
2007high 45 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OLO

119in 2025

2009: 126 days median2010: 113 days median2011: 138 days median2012: 129 days median2013: 137 days median2014: 119 days median2015: 176 days median2016: 120 days median2017: 91 days median2018: 97 days median2019: 95 days median2020: 89 days median2021: 91 days median2022: 89 days median2023: 90 days median2024: 90 days median2025: 119 days median2026 (partial year): 109 days median
2007high 176 (2015) · low 89 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OLO

1in 2025

2010: 12011: 12012: 12014: 32015: 22016: 32017: 22018: 12019: 12020: 62022: 52023: 7, 1 Class I2024: 42025: 12026 (partial year): 6
2007high 7 (2023) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · OLO

3,306in 2025

2008: 12009: 42010: 42011: 102012: 602013: 362014: 2452015: 4662016: 7502017: 1,0762018: 1,2302019: 1,5852020: 2,7462021: 1,6962022: 1,5682023: 1,4932024: 2,3532025: 3,306
2007high 3,306 (2025) · low 0 (2007)2025

MAUDE reports naming product code OLO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—01
20091112604
2010911314
201111138110
20127129160
20137137036
2014101193245
2015101762466
2016141203750
2017339121,076
2018249711,230
2019359511,585
2020348962,746
2021289101,696
2022408951,568
202345907 (1)1,493
2024449042,353
20254011913,306
2026421096—

What is it indicated for?

The Partial Knee Application (PKA), for use with the Mako System, is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. The Partial Knee Application (PKA), for use with the Mako System, is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid…

Read the full statement (107 words)

The Partial Knee Application (PKA), for use with the Mako System, is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

The Partial Knee Application (PKA), for use with the Mako System, is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include unicondylar knee replacement and/or patellofemoral knee replacement.

The Implant systems with which the system is compatible:

• Restoris Multicompartmental Knee System

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170891: Mako Partial Knee Application.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K170891

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026