510(k) Report

510(k) · Cleared

Recalled · Class I · 2023

AirFit N20

K171212510(k) TraditionalCleared Sep 28, 2017156-day reviewClass II21 CFR 868.5905Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Resmed, Ltd. · BZD Ventilator, Non-Continuous (Respirator)

Cleared on Sep 28, 2017 as substantially equivalent to K161978 (Resmed, Ltd., 2016) after a 156-day review. Its family tree reaches back 11 generations to 1998. 3 of the 41 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
Sep 28, 2017
Review
156 days
Class
Class II
Product code
BZD
Recalls
1 (1 Class I)
Predicates
1

Its longest line of descent runs back 11 generations to K982530 (1998). 37 ancestors in all, the oldest K900164 (1990). 3 later devices descend from it, 2 citing it directly. 3 of the 41 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 2 other ancestors2005: 3 other ancestors2009: 2 other ancestors1996: 3 other ancestors2008: 1 other ancestor2004: 1 other ancestor2003: 2 other ancestors2002: 3 other ancestors2000: 1 other ancestor2007: 1 other ancestor2001: 1 other ancestor1999: 2 other ancestors1998: 2 other ancestors1995: 1 other ancestor1990: 1 other ancestor2026: 1 descendant cleared2018: 1 descendant cleared2023: 1 descendant clearedK982530 · Sullivan Mirage Full Face Mask · Resmed, Ltd. · 1998K023244 · Mirage Full Face Mask Series 2 · Resmed, Ltd. · 2002K032916 · Mirage Activa Mask · Resmed, Ltd. · 2003K042403 · Mirage Swift · Resmed, Ltd. · 2004K063122 · Mirage Quattro Full Face Mask · Resmed, Ltd. · 2007K072940 · Mirage Micro · Resmed, Ltd. · 2007K102746 · Mirage FX · Resmed, Ltd. · 2010K123789 · Swift FX Nano · Resmed, Ltd. · 2013K132887 · Airfit N10 · Resmed, Ltd. · 2013 · Class I recallK153673 · AirFit N20 · Resmed, Ltd. · 2016K161978 · AirFit N20 · Resmed, Ltd. · 2016 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 4 generations back: 2 other ancestors5 generations back: 3 other ancestors6 generations back: 8 other ancestors7 generations back: 7 other ancestors8 generations back: 4 other ancestors9 generations back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 1 descendant K161978 · AirFit N20 · Resmed, Ltd. · 2016 · Class I recallK153673 · AirFit N20 · Resmed, Ltd. · 2016K132887 · Airfit N10 · Resmed, Ltd. · 2013 · Class I recallK123789 · Swift FX Nano · Resmed, Ltd. · 2013K102746 · Mirage FX · Resmed, Ltd. · 2010K072940 · Mirage Micro · Resmed, Ltd. · 2007K063122 · Mirage Quattro Full Face Mask · Resmed, Ltd. · 2007K042403 · Mirage Swift · Resmed, Ltd. · 2004K032916 · Mirage Activa Mask · Resmed, Ltd. · 2003K023244 · Mirage Full Face Mask Series 2 · Resmed, Ltd. · 2002K982530 · Sullivan Mirage Full Face Mask · Resmed, Ltd. · 1998−10−50
  1. Root · 11 generations backK982530 Sullivan Mirage Full Face MaskResmed, Ltd. · 1998
  2. 7 generations between
  3. 3 generations backK132887 Airfit N10Resmed, Ltd. · 2013 · Recalled (Class I)
  4. 2 generations backK153673 AirFit N20Resmed, Ltd. · 2016
  5. 1 generation backK161978 AirFit N20Resmed, Ltd. · 2016 · Recalled (Class I)
  6. This deviceK171212 AirFit N20Resmed, Ltd. · 2017
  7. Descendant · cites it directly · 1 of its ownK180497 Scone MaskResmed, Ltd. · 2018
  8. Descendant · cites it directly · 0 of its ownK251847 Sleepnet Arie Full Face Vented Mask; Sleepnet Arie Nasal Vented MaskSleepnet Corporation · 2026
  9. Descendant · 2 generations on · 0 of its ownK223747 Whitsundays Mask SystemResmed Pty Ltd (Registration Number: 3004604967) · 2023
Show all 11 generations
  1. Root · 11 generations backK982530 Sullivan Mirage Full Face MaskResmed, Ltd. · 1998
  2. 10 generations backK023244 Mirage Full Face Mask Series 2Resmed, Ltd. · 2002
  3. 9 generations backK032916 Mirage Activa MaskResmed, Ltd. · 2003
  4. 8 generations backK042403 Mirage SwiftResmed, Ltd. · 2004
  5. 7 generations backK063122 Mirage Quattro Full Face MaskResmed, Ltd. · 2007
  6. 6 generations backK072940 Mirage MicroResmed, Ltd. · 2007
  7. 5 generations backK102746 Mirage FXResmed, Ltd. · 2010
  8. 4 generations backK123789 Swift FX NanoResmed, Ltd. · 2013
  9. 3 generations backK132887 Airfit N10Resmed, Ltd. · 2013 · Recalled (Class I)
  10. 2 generations backK153673 AirFit N20Resmed, Ltd. · 2016
  11. 1 generation backK161978 AirFit N20Resmed, Ltd. · 2016 · Recalled (Class I)
  12. This deviceK171212 AirFit N20Resmed, Ltd. · 2017

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K180497Scone MaskResmed, Ltd. · 2018 · cites it directly1descendant
  • K251847Sleepnet Arie Full Face Vented Mask; Sleepnet Arie Nasal Vented MaskSleepnet Corporation · 2026 · cites it directly0descendants
  • K223747Whitsundays Mask SystemResmed Pty Ltd (Registration Number: 3004604967) · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K982530Sullivan Mirage Full Face MaskResmed, Ltd.1998—
−10K023244Mirage Full Face Mask Series 2Resmed, Ltd.2002—
−9K032916Mirage Activa MaskResmed, Ltd.2003—
−8K042403Mirage SwiftResmed, Ltd.2004—
−7K063122Mirage Quattro Full Face MaskResmed, Ltd.2007—
−6K072940Mirage MicroResmed, Ltd.2007—
−5K102746Mirage FXResmed, Ltd.2010—
−4K123789Swift FX NanoResmed, Ltd.2013—
−3K132887Airfit N10Resmed, Ltd.2013Class I
−2K153673AirFit N20Resmed, Ltd.2016—
−1K161978AirFit N20Resmed, Ltd.2016Class I
0 (this device)K171212AirFit N20Resmed, Ltd.2017—
+1K180497Scone MaskResmed, Ltd.2018—
+1K251847Sleepnet Arie Full Face Vented Mask; Sleepnet Arie Nasal Vented MaskSleepnet Corporation2026—
+2K223747Whitsundays Mask SystemResmed Pty Ltd (Registration Number: 3004604967)2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199510
199630
199820
199920
200010
200110
200230
200320
200410
200530
200620
200710
200810
200920

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202310
202610
FIG. 1 — Lineage of K171212, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZD device

  1. #1 K250624 myAirscore 15.7 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K161978AirFit N20
Summary page 4
…ation & 21 CFR 868.5905, 73 BZD (Class II) Classification Name Accessory to Noncontinuous Ventilator (IPPB) Legally Marketed AirFit N20 (K161978) Predicate Devices Device Description The AirFit N20 is an externally placed vented respiratory mask…
Resmed, Ltd.2016Class I

Also cleared against K161978: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180497Scone MaskResmed, Ltd.20181
K251847Sleepnet Arie Full Face Vented Mask; Sleepnet Arie Nasal Vented MaskSleepnet Corporation20260

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 20, 2023Class IZ-0539-2024Device DesignMasks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K161978AirFit N20Resmed, Ltd. · 2016 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K132887Airfit N10Resmed, Ltd. · 2013 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 19 years and passes through 1 company; a new generation was cleared every 1.0 years on median.
  • 2 of the 37 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K180497 (Resmed, Ltd., 2018) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K161978.

How does its category compare?

All BZD devices

BZD Ventilator, Non-Continuous (Respirator): FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · BZD

7in 2025

2007: 212008: 172009: 232010: 122011: 232012: 212013: 232014: 142015: 112016: 122017: 52018: 52019: 62020: 62021: 62022: 42023: 52024: 102025: 72026 (partial year): 10
2007high 23 (2009) · low 4 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZD

189in 2025

2007: 83 days median2008: 108 days median2009: 90 days median2010: 122 days median2011: 125 days median2012: 133 days median2013: 127 days median2014: 172 days median2015: 224 days median2016: 217 days median2017: 221 days median2018: 280 days median2019: 231 days median2020: 137 days median2021: 192 days median2022: 253 days median2023: 204 days median2024: 132 days median2025: 189 days median2026 (partial year): 214 days median
2007high 280 (2018) · low 83 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZD

8in 2023

2007: 292008: 22009: 12011: 12019: 4, 4 Class I2023: 8, 7 Class I2026 (partial year): 2, 2 Class I
2007high 29 (2007) · low 0 (2010)2026

Class I (13 since 2007) Class II/III not classified

Adverse event reports per year · BZD

60,279in 2025

2007: 652008: 672009: 1472010: 1432011: 1892012: 1392013: 822014: 862015: 1142016: 1182017: 1482018: 1012019: 1882020: 1262021: 6,0662022: 90,7682023: 27,9872024: 68,2622025: 60,279
2007high 90,768 (2022) · low 65 (2007)2025

MAUDE reports naming product code BZD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200721832965
200817108267
200923901147
2010121220143
2011231251189
2012211330139
201323127082
201414172086
2015112240114
2016122170118
201752210148
201852800101
201962314 (4)188
202061370126
2021619206,066
20224253090,768
202352048 (7)27,987
202410132068,262
20257189060,279
2026102142 (2)—

What is it indicated for?

The AirFit N20 channels airflow non-invasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel device.

The AirFit N20 is:

• to be used by patients weighing more than 66 lb (30 kg) for whom positive airway pressure has been prescribed

• intended for single patient re-use in the home environment and multi-patient re-use in the hospital/institutional environment.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K171212: AirFit N20.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K171212

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026