510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Astral 100/150

K172875510(k) TraditionalCleared Apr 26, 2018217-day reviewClass II21 CFR 868.5895Recalls 2 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Resmed, Ltd. · CBK Ventilator, Continuous, Facility Use

Cleared on Apr 26, 2018 as substantially equivalent to K152068 (Resmed, Ltd., 2016) after a 217-day review. Its family tree reaches back 9 generations to 1991. 8 of the 58 devices in its family, counting itself, have a Class I recall. It has been recalled 2 times, including 2 Class I recalls.

Decided
Apr 26, 2018
Review
217 days
Class
Class II
Product code
CBK
Recalls
2 (2 Class I)
Predicates
1

Its longest line of descent runs back 9 generations to K902859 (1991). 57 ancestors in all, the oldest K841279 (1984). No later device cites it as a predicate yet. 8 of the 58 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2010: 1 other ancestor, 1 with a Class I recall12009: 1 other ancestor, 1 with a Class I recall12005: 1 other ancestor2001: 4 other ancestors2007: 1 other ancestor2004: 1 other ancestor2003: 3 other ancestors, 1 with a Class I recall12000: 4 other ancestors1999: 4 other ancestors2002: 1 other ancestor1998: 7 other ancestors, 2 with a Class I recall21997: 2 other ancestors1996: 3 other ancestors1995: 3 other ancestors1984: 2 other ancestors1994: 3 other ancestors1992: 2 other ancestors1991: 1 other ancestor1990: 4 other ancestorsReference device: K161492 · Juno VPAP ST-A · Resmed, Ltd. · 2017K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK960010 · Servo Ventilator 300 Computer Interface · Siemens Elema AB · 1996K970839 · Servo Ventilator 300A · Siemens Elema AB · 1997 · recalledK992788 · Bear Cub 750 PSV · Bear Medical Systems, Inc. · 1999K013642 · AVEA · Bird Products Corp. · 2002 · Class I recallK022674 · Bird Avea Ventilator · Bird Products Corp. · 2002 · recalledK060647 · LTV 1200 Ventilator · Pulmonetic Systems, Inc. · 2006K133868 · Astral Ventilator · Resmed, Ltd. · 2014 · recalledK152068 · Astral 100/150 · Resmed, Ltd. · 2016 · Class I recall19801990200020102020
  1. Root · 9 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 5 generations between
  3. 3 generations backK060647 LTV 1200 VentilatorPulmonetic Systems, Inc. · 2006
  4. 2 generations backK133868 Astral VentilatorResmed, Ltd. · 2014 · Recalled
  5. 1 generation backK152068 Astral 100/150Resmed, Ltd. · 2016 · Recalled (Class I)
  6. This deviceK172875 Astral 100/150Resmed, Ltd. · 2018
Show all 9 generations
  1. Root · 9 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 8 generations backK960010 Servo Ventilator 300 Computer InterfaceSiemens Elema AB · 1996
  3. 7 generations backK970839 Servo Ventilator 300ASiemens Elema AB · 1997 · Recalled
  4. 6 generations backK992788 Bear Cub 750 PSVBear Medical Systems, Inc. · 1999
  5. 5 generations backK013642 AVEABird Products Corp. · 2002 · Recalled (Class I)
  6. 4 generations backK022674 Bird Avea VentilatorBird Products Corp. · 2002 · Recalled
  7. 3 generations backK060647 LTV 1200 VentilatorPulmonetic Systems, Inc. · 2006
  8. 2 generations backK133868 Astral VentilatorResmed, Ltd. · 2014 · Recalled
  9. 1 generation backK152068 Astral 100/150Resmed, Ltd. · 2016 · Recalled (Class I)
  10. This deviceK172875 Astral 100/150Resmed, Ltd. · 2018

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K902859Servo Ventilator 300Siemens Life Support Systems19913
−8K960010Servo Ventilator 300 Computer InterfaceSiemens Elema AB1996—
−7K970839Servo Ventilator 300ASiemens Elema AB19974
−6K992788Bear Cub 750 PSVBear Medical Systems, Inc.1999—
−5K013642AVEABird Products Corp.2002Class I
−4K022674Bird Avea VentilatorBird Products Corp.20025
−3K060647LTV 1200 VentilatorPulmonetic Systems, Inc.2006—
−2K133868Astral VentilatorResmed, Ltd.20142
−1K152068Astral 100/150Resmed, Ltd.2016Class I
0 (this device)K172875Astral 100/150Resmed, Ltd.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198420
199040
199110
199220
199430
199530
199630
199720
199872
199940
200040
200140
200210
200331
200410
200510
200710
200911
201011
FIG. 1 — Lineage of K172875, 1991–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

8 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K152068Astral 100/150
Summary page 4
…Product Code Primary product code CBK Secondary product code NOU Predicate Device ResMed Astral 100/150 (K152068) Reference Device ResMed Juno VPAP ST-A (K161492) 4. Indications For Use The Astral 100/150…
Resmed, Ltd.2016Class I
†K161492Juno VPAP ST-AResmed, Ltd.2017—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (8 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 25, 2026Class IZ-2735-2026Component design/selectionDue to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
Sep 13, 2023Class IZ-0111-2024Device DesignIf ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K152068Astral 100/150Resmed, Ltd. · 2016 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K093416Trilogy 200 VentilatorRespironics, Inc. · 2010 · 3 generations back · ◆ in the 2010 ancestors bar of FIG. 14 Class I recalls
  • K083526Trilogy 100 VentilatorRespironics Inc., Sleep & Home Respiratory Group · 2009 · 4 generations back · ◆ in the 2009 ancestors bar of FIG. 15 Class I recalls
  • K013642AVEABird Products Corp. · 2002 · 5 generations back · ◆ on the line of descent in FIG. 113 Class I recalls
  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 6 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K981072Esprit VentilatorRespironics, Inc. · 1998 · 6 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 6 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 27 years and passes through 6 companies; a new generation was cleared every 2.0 years on median.
  • 25 of the 57 ancestors have a recall of any class; 7 have a Class I recall.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Astral 100/150 provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.

The iVAPS mode with optional AutoEPAP is intended for patients weighing more than 66lb (30kg).

The Astral device is intended to be used in home, institution/hospital and portable applications for both invasive and non-invasive ventilation.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172875: Astral 100/150.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172875

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026