510(k) · Cleared
ClearedSterile Nitrile Surgical Powder Free Gloves Tested for Use with Chemotherapy Drugs and Sterile Latex Surgical Powder Free Gloves
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Top Glove Sdn. Bhd. · KGO Surgeon'S Gloves
Cleared on Oct 12, 2017 after a 164-day review. FDA’s file for it is a 510(k) statement, which names no predicate.
- Decided
- Oct 12, 2017
- Review
- 164 days
- Class
- Class I
- Product code
- KGO
- Recalls
- None
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KGO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KGO device
- #1 K192328 JR.MEDIK Latex Surgeon’s Gloves Powder Freescore 15.2 · 5 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
How does its category compare?
All KGO devicesKGO Surgeon'S Gloves: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.
FDA decisions per year · KGO
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KGO
185in 2023
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KGO
2in 2023
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · KGO
21in 2025
MAUDE reports naming product code KGO, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 10 | 113 | 0 | 13 |
| 2008 | 2 | — | 0 | 8 |
| 2009 | 3 | 48 | 15 | 14 |
| 2010 | 6 | 87 | 1 | 16 |
| 2011 | 11 | 90 | 0 | 14 |
| 2012 | 7 | 133 | 0 | 14 |
| 2013 | 6 | 184 | 0 | 41 |
| 2014 | 7 | 257 | 0 | 28 |
| 2015 | 12 | 225 | 0 | 11 |
| 2016 | 9 | 218 | 2 | 15 |
| 2017 | 12 | 111 | 1 | 20 |
| 2018 | 7 | 194 | 8 | 15 |
| 2019 | 11 | 204 | 0 | 51 |
| 2020 | 15 | 117 | 0 | 15 |
| 2021 | 14 | 199 | 0 | 14 |
| 2022 | 6 | 127 | 2 | 10 |
| 2023 | 14 | 185 | 2 | 28 |
| 2024 | 2 | — | 0 | 18 |
| 2025 | 1 | — | 0 | 21 |
| 2026 | 2 | — | 0 | — |
What is it indicated for?
Sterile Nitrile Surgical Powder Free Gloves Tested for Use with Chemotherapy Drugs is to be wotn on the hands of healthcare professionals during surgery to prevent cross contamination between healthcare personnel and the patient. These gloves are tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Chemotherapy Drug Permeation The following chemicals have been…
Read the full statement (162 words)
Sterile Nitrile Surgical Powder Free Gloves Tested for Use with Chemotherapy Drugs is to be wotn on the hands of healthcare professionals during surgery to prevent cross contamination between healthcare personnel and the patient.
These gloves are tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Chemotherapy Drug Permeation
The following chemicals have been tested with these gloves.
Chemotherapy Drug Concentration Breakthrough Detection Time in Minutes
Carmustin (BCNU) 3.3mg/ml 8.0
Cisplatin 1.0mg/ml >240
Cyclophosphamide (Cytoxan) 20.0mg/ml >240
Dacarbazine (DTIC) 10.0mg/ml >240
Doxorubicin Hydrochloride 2.0mg/ml >240
Etoposide (Toposar) 20.0mg/ml >240
Fluorouracil 50.0mg/ml >240
Paclitaxel (Taxol) 6.0mg/ml >240
Thiotepa 10.0mg/ml 16.2
* Please note that the following drugs have extremely low permeation times:
Carmustin (BCNU) : 8.0 minutes and Thiotepa : 16.2 minutes
Sterile Latex Surgical Powder Free Gloves is to be worn on the hands of healthcare professionals during surgery to prevent cross contamination between healthcare personnel and the patient.
Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K171279 · printed Nov 6, 2026
Cite this page
510(k) Report. “K171279: Sterile Nitrile Surgical Powder Free Gloves Tested for Use with Chemotherapy Drugs and Sterile Latex Surgical Powder Free Gloves.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K171279
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026