510(k) Report

510(k) · Cleared

Cleared

NAPA LP-15 Airway Pressure Monitor

K172284510(k) TraditionalCleared Feb 9, 2018196-day reviewClass II21 CFR 868.2600

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Drw Medical, LLC · CAP Monitor, Airway Pressure (Includes Gauge And/Or Alarm)

Cleared on Feb 9, 2018 as substantially equivalent to K992101 (Caradyne, Ltd., 2000) after a 196-day review. Its family tree reaches back 2 generations to 1987.

Decided
Feb 9, 2018
Review
196 days
Class
Class II
Product code
CAP
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K871083 (1987). 5 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1993: 2 other ancestors1987: 1 other ancestorK871083 · CM 5000 Monitor Airway Pressure · Canadian Monaghan, Ltd. · 1987K992101 · Criterion 40 or PM 40 · Caradyne, Ltd. · 200019851990199520002005201020152020
Other ancestors, by generation 2 generations back: 3 other ancestors K992101 · Criterion 40 or PM 40 · Caradyne, Ltd. · 2000K871083 · CM 5000 Monitor Airway Pressure · Canadian Monaghan, Ltd. · 19870
  1. Root · 2 generations backK871083 CM 5000 Monitor Airway PressureCanadian Monaghan, Ltd. · 1987
  2. 1 generation backK992101 Criterion 40 or PM 40Caradyne, Ltd. · 2000
  3. This deviceK172284 NAPA LP-15 Airway Pressure MonitorDrw Medical, LLC · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K871083CM 5000 Monitor Airway PressureCanadian Monaghan, Ltd.1987—
−1K992101Criterion 40 or PM 40Caradyne, Ltd.2000—
0 (this device)K172284NAPA LP-15 Airway Pressure MonitorDrw Medical, LLC2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199320
FIG. 1 — Lineage of K172284, 1987–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CAP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CAP device

  1. #1 K201666 Single Use Manometerscore 0.3 · 1 generation · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K992101Criterion 40 or PM 40
Summary page 4
…(3) Predicate Device(s): Substantial Equivalence to: K Number Model Manufacturer K992101 Criterion 40 Airway Pressure Caradyne Ltd. (now Monitor R…
Caradyne, Ltd.2000—

Also cleared against K992101: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 3 companies; a new generation was cleared every 18.1 years on median.
  • None of the 5 ancestors has a recall.
  • 3 other devices also cite its predicate K992101.

How does its category compare?

All CAP devices

CAP Monitor, Airway Pressure (Includes Gauge And/Or Alarm): FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · CAP

1in 2020

2008: 12010: 12011: 22012: 12013: 22014: 12015: 12016: 12018: 12019: 22020: 1
2007high 2 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CAP

Too few CAP 510(k)s in any year since 2007 to give a median.

Recalls per year · CAP

No recall names CAP since 2007.

Adverse event reports per year · CAP

2in 2025

2009: 12012: 32019: 42020: 12021: 52022: 62023: 112024: 112025: 2
2007high 11 (2023) · low 0 (2007)2025

MAUDE reports naming product code CAP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—00
20090—01
20101—00
20112—00
20121—03
20132—00
20141—00
20151—00
20161—00
20170—00
20181—00
20192—04
20201—01
20210—05
20220—06
20230—011
20240—011
20250—02
20260—0—

What is it indicated for?

The NAPA LP-15 Airway Pressure Monitor is intended to measure and monitor mean airway pressure for neonatal patients being treated with positive pressure therapy, including neonatal Continuous Positive Airway Pressure (CPAP) devices (e.g. Bubble CPAP). The device provides audible and visual alarms when the airway pressure falls outside of the user selected high and low alarm limits. The device is intended to be pole mounted for stationary use in hospitals only. For professional use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172284: NAPA LP-15 Airway Pressure Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172284

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026